Analytical-Scale LC Purification with High Load and Low Dispersion
Technical Reference Summary
Covered Models & Document Scope
Resumen técnico de un flujo de trabajo de purificación a escala analítica que utiliza un Agilent 1260 Infinity II Analytical LC ampliado con el 1260 Infinity II Analytical-Scale Fraction Collector (G1364F); compara las columnas ZORBAX Eclipse Plus C18, InfinityLab Poroshell 120 y Pursuit XRs. No es una ficha técnica de producto con códigos de pedido.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Maximum flow rate for fraction collection | 10 mL/min p.2 ↗ | 1260 Infinity II Analytical-Scale Fraction Collector (G1364F) |
| Injection volume for quantitation | 0.5 µL p.3 ↗ | |
| Injection volume for purification | 75 to 100 µL p.3 ↗ | |
| Needle wash | 10 s p.3 ↗ | acetonitrile/water 1:1 (v:v) |
| Total run time | 13 minutes p.3 ↗ | |
| Peak width (10%) — ZORBAX Eclipse Plus C18 | 0.085 p.3 ↗ | ZORBAX Eclipse Plus C18 |
| Resolution (EP) — ZORBAX Eclipse Plus C18 | 13.1 p.3 ↗ | ZORBAX Eclipse Plus C18 |
| Selectivity — ZORBAX Eclipse Plus C18 | 1.71 p.3 ↗ | ZORBAX Eclipse Plus C18 |
| Peak width (10%) — InfinityLab Poroshell 120 | 0.065 p.3 ↗ | InfinityLab Poroshell 120 |
| Resolution (EP) — InfinityLab Poroshell 120 | 16.3 p.3 ↗ | InfinityLab Poroshell 120 |
| Selectivity — InfinityLab Poroshell 120 | 1.73 p.3 ↗ | InfinityLab Poroshell 120 |
| Peak width (10%) — Pursuit XRs | 0.085 p.3 ↗ | Pursuit XRs |
| Resolution (EP) — Pursuit XRs | 13.8 p.3 ↗ | Pursuit XRs |
| Selectivity — Pursuit XRs | 1.60 p.3 ↗ | Pursuit XRs |
| Maximum injection volume without significant coelution — Pursuit XRs | 100 µL p.5 ↗ | Pursuit XRs, 4 mg target compound |
| Maximum injection volume without significant coelution — Pursuit XRs (repeated loading) | 80 µL p.7 ↗ | Pursuit XRs, high-concentration sample |
| Purity of collected fractions | >99 % p.6 ↗ | all three columns tested |
| Recovery | 96 to 97 % p.6 ↗ | average, all three columns tested |
| Recovery — Pursuit XRs (repeated loading) | 97 % p.7 ↗ | average |
| Maximum purified target compound | more than 50 mg p.7 ↗ | Pursuit XRs, four runs |
| Sample concentration — HBA | 8 mg/mL p.2 ↗ | acetonitrile/water |
| Sample concentration — target compound | 40 mg/mL p.2 ↗ | acetonitrile/water |
Operating Limits
- Maximum flow rate for fraction collection: 10 mL/min p.2 ↗
- Maximum injection volume without significant coelution (Pursuit XRs): 100 µL p.5 ↗
- Maximum injection volume without significant coelution (Pursuit XRs, high-concentration sample): 80 µL p.7 ↗