VS Series Vacuum Pump Controller Instruction Manual

Sommaire

  1. Overview pp. 1-6
    Introduces VS Series controllers and document conventions
  2. pp. 7-12
    List of figures and illustrations index
  3. Overview pp. 13-18
    Safety notes, conventions, and manual organization
  4. Procedure pp. 19-45
    Mechanical and electrical installation instructions
  5. Procedure pp. 46-120
    Operating modes, controls, and configuration procedures
  6. Procedure pp. 139-162
    Profinet setup and operation details
  7. Procedure pp. 163-246
    Component replacement and maintenance steps
  8. Spec pp. 247-252
    Technical specifications and product data
  9. pp. 250-252
    Vacuum Products Division contact details
  10. p. 247 table, 14 rows
  11. p. 131 table, 8 rows
  12. p. 134 table, 5 rows
  13. p. 129 table, 12 rows
  14. p. 144 table, 34 rows
  15. p. 41 table, 4 rows
  16. p. 59 table, 7 rows

Table E-1 Leak Detector Series Specifications

SpecificationsDescription
Minimal Detectable Leak5E-10 atm std cc/sec He, 5E-10 mbar l/s, 5E-11 Pa m3/sec
Selectable Sensitivity RangesFour working ranges covering to E-3 to E-9 atm cc/sec He
Maximum Test Port (foreline pressure5 Torr, 7 mbar, 667 Pa
Calibration RoutineAutomated or manual (internal or external)
Background SuppressionAuto Zero (turn on/off background) and Auto Zero < Zero function
Display Interface (optional)High clarity TFT color touch screen
Display - Selectable languagesEnglish, French, German, Spanish, Japanese, Korean & Mandarin
Set Points (process control)Five set points: four leak rate and /or pressure and one audio
Analog Output Voltage0 - 10 VDC: Log 1V/decade, log 2V/decade and linear within the decade (DB-9F)
Serial InterfaceRS232 isolated interface @ 9600 baud (DB-9M)
Discrete I/OOptically isolated, 5-24 VDC outputs and inputs (DB-25S)
Networking Interface (optionalProfinet® I/O option (RJ45)
Power Input24 VDC power input; 24 VDC, +/- 5%, max current: 5 A
Operating Temperature+5 to +48°C

Table 4-3 Non-Volatile Operating Parameters

CommandParameter
INIT-AZ<0Preceded by 0 or 1, sets the status of Auto-zero < 0. 0 = off, 1 = on.
INIT-CPLOCKOUTSets the status of the control panel lockout. Preceded by (0) or (1), 0 = OFF, 1 = ON.
INIT-DATESets the system date and time. Preceded by: dow, mm dd yyyy hh mm ss AmPm 24-12 mode, Where: ❑ 2 is the day of the week (dow) (0 = Sunday, 1 = Monday, etc), ❑ mm/dd/yyyy (date), ❑ hh:mm:ss (time) ❑ AmPm (0=AM, 1=PM) ❑ 24-12 mode (0 for 24 hr, 1 for 12 hr). Examples: ❑ 2 02 12 2005 04 16 33 1 0 INIT-DATE ❑ (Tuesday, February 12, 2005, 4:16:33 on a 24 hr clock) ❑ 2 02 12 2005 04 16 33 1 1 INIT-DATE(Tuesday, February 12, 2005, 4:16:33 PM) ❑ ?CURRDATE reports the stored value
INIT-DISPLAYSets the status of the bar graph display to either a log or linear display. Preceded by (1) Log or (0) Linear.
INIT-EMISSIONSets the emission of the ion source. Preceded by a three-digit number ( counts), in the range 0 to 255.
INIT-EXPONENTSets the range that the leak detector will operate in when Manual Range is ON. Preceded by -XX. This command is ignored if the system is in internal calibrated leak mode.
INIT-EXTLEAKSets the value of the external leak. Preceded by a two-digit leak rate mantissa with a decimal point after the first digit followed by E-, and then by a two-digit leak range exponent: the helium leak rate number in atm cc/sec of the calibration standard leak. Example: 3.2E-08 INIT-EXTLEAK
INIT-FILAMENTSets the operating filament in the ion source. Preceded by a single digit, either 1 or 2.

Table 4-4 Spectrometer Operating Parameters

CommandParameter
PUT-EXPONENTSets the exponent leak range for Manual mode. Preceded by a two-character numbers. Acceptable values are 0 through -10. Unacceptable values will not be stored. This command is ignored if the system is in internal calibrated leak mode.
PUT-IONA three digit number of volts, in the range 200 to 350, which sets the ion voltage of the ion source.
PUT-EMISSIONA four digit number of microamps, in the range 0300 to 2000, which sets the emission current of the ion source.
PUT-GAINIf the system is in the standard leak mode, it sets the value of the (internal) gain used for adjusting the helium signal to match an internal standard leak. If the system is in TEST mode, it sets the value of the (external) gain used for adjusting the helium signal to match an external calibrated leak. Preceded by a one or up to three-digit number and one decimal point. In the range 0.2 to 9.0 for the internal calibrated leak gain and 0.2 to 150.0 for the external calibrated leak gain.
PUT-RANGESTOPSets the Range Stop exponent. Preceded by two or three characters -XX or –X. This command is ignored if the system is in internal calibrated leak mode.

Table 4-2 Internal Operating Parameters (Continued)

CommandDescription
?PWDONOFFReports status of system password (ON or OFF).
?RANGEA two character (minus followed by one digit) number indicating the exponent of the least sensitive ranges of detectable leak. The reported leak rates are in this range and three lower (more negative exponent) ranges. See also INIT-RANGE.
?RANGESTOPReports the value of the Range Stop variable.
?xREJECTReports the reject leak rate setpoint (e.g., 7.0E-05) (where x is a number from 1 to 4 or the letter A). A six character number consisting of a two digit reject leak rate mantissa with a decimal point after the first digit followed by E– then a single digit which is the reject leak rate range exponent (eg. 7.0E-5). This inquiry is used for reporting the reject leak rate most recently input using INIT-xREJECT
?xREJECTPReports the reject test port pressure setpoint (e.g., 1.0E+01) (where x is a number from 1 to 4). A six character number consisting of a two digit reject test port pressure mantissa with a decimal point after the first digit followed by E– then a single digit which is the reject pressure range exponent (eg. 1.0E+01). This inquiry is used for reporting the reject test port pressure most recently input using INIT-xREJECTP.
?>xREJECTReports the reject set point is on or off (where x is a number from 1 to 4 or the letter A).
?RSONOFFReports the status of the Range stop function as Enabled or Disabled.
SERVICE1Displays the current status of the operating parameters.
SERVICE2Displays the current status of the operating parameters.
?SETUPReports seven lines, each begins with a <cr><lf>. The first line reports the turbo pump speed with the RPM value or OFF. The second line reports the selected manual leak rate range, and the ranging method as auto or manual. The third line reports the least sensitive leak rate range. The fourth line reports the reject set point leak rate. The fifth line reports the internal calibrated leak rate value. The sixth line reports the DAC output method as Linear, Log (2V) The seventh line reports the active filament status as One or Two, followed by Lit or Out.
?SOFTSTARTReturns the status of Soft Start (on or off)
?SPLITFLOWReturns the status of the split flow function, ON or OFF.

Table 5-1 Configuration and Status Variables

IndexVariableAccessDescriptionResponse
00R/WClear Variable Command Word.
01RESERVED
02CPLOCKOUTR/WThe control panel lockout feature, ON or OFF.Word 5: 0=OFF, 1= ON
03CURRDATER/WThe system date as kept by the I/O PCB and transferred to the CPU on a Write.Word 5: mm, Word 6: dd, Word 7: yyyy
04DISPLAYR/WThe display bar graph mode, log or lin.Word 5: 1 = LOG, 0 = LIN
05EXPONENTR/WThe manual-mode exponent.Word 5: xx i.e. 03 or 10
06EXTLEAKR/WThe external calibrated leak.value.Word 5: xxx (mantissa X 1000), Word 6: yy (exponent) Word 7: 0 = neg, 1 = pos exponent
07GAINR/WThe system gain value.Word 5: xx (gain X 10 i.e. 22 = 2.2)
08RESERVED
09TIMER/WThe system time as kept by the I/O PCB and transferred to the CPU on a Write.Word 5: hhmm,Word 6: ssap (ap = 1 for pm)
10RESERVED
11LKEXPIRER/WThe internal calibrated leak expiration date mm/dd/yyyy.Word 5: mm, Word 6: dd, Word 7: yyyy
12LEAKTEMPR/WThe internal calibrated leak temperature that the factory calibration was performed at (degrees C).Word 5: tt (temp X 10)
13TPPRESSURERThe current Test Port pressure in mTorr.Word 5: xxx (mantissa X 1000), Word 6: yy (exponent) Word 7: 0 = neg exponent
14SYPRESSURERThe current System Pressure in mTorr.Word 5: xxx (mantissa X 1000), Word 6: yy (exponent) Word 7: 0 = neg exponent
15RESERVED
16RESERVED
17RANGER/WThe number indicating the exponent of the least sensitive range of detectable leak. The reported leak rates are in this range and three lower (more negative exponent) ranges.Word 5: yy (exponent), Word 6: 0 = neg, 1 = pos exponent
18RANGESTOPR/WThe Rangestop variable.Word 5: yy (exponent), Word 6: 0 = neg, 1 = pos exponent
191REJECTR/WThe reject 1 leak rate setpoint.Word 5: xxx (mantissa X 1000), Word 6: yy (exponent), Word 7: 0 = neg, 1 = pos exponent
202REJECTR/WThe reject 2 leak rate setpoint.Word 5: xxx (mantissa X 1000), Word 6: yy (exponent), Word 7: 0 = neg, 1 = pos exponent
213REJECTR/WThe reject 3 leak rate setpoint.Word 5: xxx (mantissa X 1000), Word 6: yy (exponent), Word 7: 0 = neg, 1 = pos exponent
224REJECTR/WThe reject 4 leak rate setpoint.Word 5: xxx (mantissa X 1000), Word 6: yy (exponent), Word 7: 0 = neg, 1 = pos exponent
23RESERVEDReports the currently stored Test Sequencer rough time used as part of setting up the Auto Sequencer.
24STDLEAKR/WThe calibration standard leak rate.Word 5: xxx (mantissa X 1000), Word 6: yy (exponent), Word 7: 0 = neg, 1 = pos exponent
25STDLEAKtRReports the temperature corrected value of the internal calibrated leak. If the value is out of range or the temp cannot be read, the STDLEAK value is returned.Word 5: xxx (mantissa X 1000), Word 6: yy (exponent), Word 7: 0 = neg, 1 = pos exponent
26RESERVED
27SYSPWDRThe value of the system password.
28RESERVED
29TEMPFACTORR/WCoefficient of temperature change of the cal leak (Determined at the factory when the cal leak was calibrated).Word 5: tt (temp X 10), Word 6: 1 = neg
30RESERVED
31VERRThe CPU software version in the form LC02.05.Word 5: xxxx (Revision X 100), Word 6: yy (Chksum MSByte), Word 7: Chksum LSByte ex
32CHKSUMRThe checksum of the software in the CPU PCB.Word 5: Hi Byte, Word 6: Lo Byte
33WHYNOCALRThe calibration failure diagnostics.Word 5: 0 to 5 0 = Cal Ok, 1 = No Ion Peak, 2 = No E Peak, 3 = GainTooHigh, 4 = GainTooLow, 5 = LostPower

Table 1-1 Installation Requirements

RequirementSpecification
Mains Supply Voltage FluctuationsMust not exceed ±10% of the nominal voltage.
PollutionPollution Degree 2 per EN61010-1 3rd Edition Standard.
Foreline Pump❑ A primary vacuum pump with a minimal pumping 285-300 l/m (10 cfm) for use as a Foreline/(Roughing) Test Pump for a single pump configuration system. Recommended: Agilent's DS302 Rotary Vane Mechanical pump. ❑ A primary vacuum pump with a minimal pumping speed of 100 l/m (4 cfm) for use as a Dedicated Foreline Pump only in a two pump configuration system. Recommended: Agilent's DS102 Rotary Vane Mechanical pump.
Fixed Mains Socket OutletConnect to a building protective earthing system.

PE wire: position 2 (middle terminal) of the receptacle; to provide grounding between the VS C15 units and the external power supply system.

Table 2-2 General Description

ScreenPurpose
HOMEAccesses the Control Panel and Menu screens.
Activates the Zero function. The LED illuminates while zeroing occurs.
Control PanelActivates/deactivates: ❑ Std Leak ❑ Offset ❑ Tune ❑ Calibrate ❑ Zero Enable
Displays the type of leak: Internal or external, associated with Calibrate.
Enables the leak rate vs. time graphic display
Adjusts the speaker volume.
MenusAccesses the following screens for tailoring leak detector operation: ❑ Calibration Set-Up – Configures leak type, mode and parameters (Section 2.2.5.1 “Calibration Set-Up Screen” on page 2-23). ❑ Reject & Audio Set Points – Configures parameters and toggles operations on/off for each reject set point for both pressure and leak rate and for the audio threshold (Section 2.2.5.2 “Reject and Audio Set Points Screen” on page 2-25). ❑ Transfer Points with Stabilization Wait Timer - Configures the transfer points wait timer (Section 2.2.5.3 “Transfer Points with Stabilization Wait Timer” on page 2-28). ❑ Leak Ranging – Configures the Range Stop, Manual Range values, and range of operation (Section 2.2.5.4 “Leak Rate Ranging Screen” on page 2-29). ❑ System Information – Displays the status of configured and active items (Section 2.2.5.5 “System Information” on page 2-32). ❑ Control Panel Access – Panel Lockout. ❑ Set-Up - Accesses all the screens shown in Section 2-6 “First Menu Screen” on page 2-22.

Engineering Facts

12 facts · every claim linked to a page
Installation

The base unit must not be placed on the floor and needs 4 inches of clearance for ventilation and operator access.

Prevents overheating and ensures safe access for connections and operation.

Source
“Do not mount, place or operate the base unit on a floor.”
Installation

The vacuum connection should be made via a tee between the backing pump and test chamber, with the largest possible diameter and shortest possible length to achieve fast cycle time.

Minimizes gas conduction limitations and ensures the system is not restricted by piping.

Source
“To achieve fast test cycle time, the diameter of the interconnecting pipe should be as large as possible. The length of the interconnecting pipe should be a short as possible.”
Selection

A foreline pump with a minimum pumping speed of 285–300 l/m (10 cfm) is required for a single-pump configuration.

Ensures adequate roughing capacity for proper leak detector operation.

Source
“A primary vacuum pump with a minimal pumping 285-300 l/m (10 cfm) for use as a Foreline/(Roughing) Test Pump for a single pump configuration system.”
Installation

The 24 VDC power cable must be wired with +24 V on position 1, PE on position 2, and 24VRTN on position 3 of the Molex receptacle.

Incorrect wiring can damage the unit or defeat grounding, affecting safety and EMC.

Source
“+24 VDC: position 1 (D-shape terminal) of the receptacle.”
Conformité

A Snap-It ferrite (Fair-Rite p/n 0443164151 or equivalent) should be placed on cables at 2.0–3.0 inches from the plug on 24 VDC input, Analog Output, Profinet IO, and Pressure Gauge connectors to improve EMC performance.

Ensures electromagnetic compatibility and prevents interference in industrial environments.

Source
“place a Snap-It-ferrite, Fair-Rite's p/n 0443164151 or equivalent, onto the cables and wires at the distance of 2.0" to 3.0" from a plug connected to the following I/O connectors”
Safety

The leak detector is not designed for hazardous gases; the system under test must be purged and the exhaust connected to a scrubbed or toxic containment exhaust.

Prevents exposure to toxic or flammable gases that could cause serious injury or death.

Source
“The leak detector is not designed for use with hazardous gases. Verify that the system to be tested has been purged of all hazardous gases prior to using the leak detector.”
Maintenance

Do not use alcohol, methanol, or other solvents on O-rings as they cause deterioration and reduce vacuum sealing ability.

Preserves O-ring integrity and vacuum performance during maintenance.

Source
“Do not use alcohol, methanol or other solvents on O-rings. Doing so causes deterioration and reduces their ability to hold a vacuum.”
Operation

When Zero Enable is ON, discrete I/O pin 16 becomes inactive.

Affects PLC integration and signal availability for background suppression.

Source
“If Zero Enable is turned ON in the display screen, discrete I/O pin 16 is inactive.”
Operation

Leak rate set points are not monitored when the system is not ready or in Busy mode, but pressure set points are always monitored regardless of system status.

Critical for configuring PLC interlocks and process control logic correctly.

Source
“Leak Rate set points are not monitored while the system is not ready or in the Busy mode. Pressure set points are always monitored regardless of system status.”
Operation

The External Calibration ok signal (pin 12) is disabled if the internal calibration routine fails after a successful external calibration.

Prevents false pass signals in automated test systems relying on calibration status.

Source
“The External Calibration ok signal is disabled if the internal calibration routine fails after a successful external calibration routine.”
Selection

The VS C15 requires a 24 VDC power supply with at least 5 A capacity.

Undersized power supply can cause unreliable operation or shutdown during peak load.

Source
“The power supply requirements for VS C15 Component leak detector: 24 VDC, 5 A min.”
Installation

The equipment must be operated indoors in an industrial or laboratory installation only, at altitudes up to 2000 m.

Defines the allowed environment and prevents use in uncontrolled or outdoor conditions.

Source
“Indoors use in an industrial and laboratory installation only.”