Installation
The unit requires a three-phase 400 V supply with a 3x16 A fuse, so the site must provide a suitable three-phase feed.
Determines electrical infrastructure and breaker sizing before installation.
Source
“Power supply requirement (3 phase) 400V 3~N 50Hz”
Installation
The cooling water circuit needs a differential pressure of at least 3 bar and must not exceed 6 bar.
Ensures the water supply can achieve the required cooling performance without overpressurizing the unit.
Source
“min. cooling water differential pressure 3 bar”
Installation
Cooling water inlet temperature must stay between 5 and 20 °C, and pipes must be insulated if below the condensation point.
Prevents condensation and ensures rated cooling power is achievable.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, so actual capacity may be lower than the rated values.
Affects sizing of the unit for the required heat load at site conditions.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected”
Installation
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality should be used.
Guides water treatment and material compatibility to avoid corrosion.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Defines acceptable power quality for reliable operation.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices under 63 A are delivered with cable but without a plug, so a plug must be sourced separately.
Affects electrical installation planning and procurement.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.
Confirms built-in safety protection required for unattended operation.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
The pump delivers up to 25 l/min at 0.7 bar pressure and 18.5 l/min suction at 0.4 bar, with M16x1 connections.
Determines whether the pump can serve the external circuit's flow and pressure needs.
Source
“max. delivery pressure 0,7 bar”
Operation
Maximum permissible kinematic viscosity of the medium is 50 mm²/s.
Limits which heat transfer fluids can be used with the pump.
Source
“max. permissible kin. viscosity 50 mm²/s”
Selection
The bath holds 17 l, with a minimum filling of 15 l and 10 l when a displacement rack is used.
Affects fluid volume, expansion volume and displacement accessory choice.
Source
“Bath capacity with displacement rack 10 l”
Installation
The unit is IP20 and must operate between 5 and 40 °C ambient, so it needs a protected indoor location.
Defines environmental limits and enclosure protection for siting.
Source
“max. ambient temperature 40 °C”
Conformité
The refrigerant R-452A has a GWP of 2141 with 1.9 kg charge, giving 4.1 t CO2 equivalent subject to F-gas rules.
Relevant for refrigerant leak monitoring and environmental compliance obligations.
Source
“CO2 equivalent 4,1 t”
Conformité
The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011.
Class A equipment may require EMC measures in residential environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”