CS 100-H Immersion Circulator Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and general description of the CS 100-H circulator
  2. Spec p. 2
    Performance and physical specifications according to DIN 12876
  3. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...85 °C
temperature set point / displaycolour LED Touchscreen
Internal temperature sensorPt100
Temperature stability at -10°C0,5 K
Interface digitalRS232
Safety classificationI / NFL
Heating power6 kW
Cooling power at ambient temperature 20°C
at 15°C18 kW
at 10°C10 kW
at 0°C9,5 kW
at -10°C7,5 kW
at -20°C4,5 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Refrigerant quantity4,5 kg
CO2 equivalent6,29 t
Circulation pump:
max. delivery70 l/min
max. delivery pressure3,7 bar
Pump connectionG3/4 male
min. filling capacity175 l
max. filling capacity195 l
Overall dimensions WxDxH **980x820x1770 mm
Net weight335 kg
sound pressure level +/- 4 dB(A)71 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)25 A
Fuse (3 phase)3x32A
Degree of ProtectionIP52
min. ambient temperature5 °C
max. ambient temperature32 °C
from Serial-No.:1.0/24

** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller's cooling capacity is rated at 20°C ambient and will drop if ambient temperature rises, so selection must account for actual site ambient conditions.

Cooling capacity must be derated for higher ambient temperatures to avoid undersizing.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit requires a 400V 3-phase 50Hz supply with a 3x32A fuse and draws up to 25A, so the electrical installation must be sized accordingly.

Electrical supply and protection must match the specified three-phase requirements.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices under 63A are delivered with cable but without a plug, so the installer must provide a suitable plug or hard-wired connection.

Planning must include procurement and installation of a plug or connection device.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are interdependent: -5% voltage with +2% frequency is not allowed, so supply quality must be verified.

Incompatible voltage/frequency deviations can damage the equipment or void compliance.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit must be installed in an ambient temperature between 5°C and 32°C, which constrains the plant room design.

Site ambient conditions must be within the allowed range for reliable operation.

Source
“min. ambient temperature 5 °C”
Operation

The flow rate can be adjusted via a manual bypass valve on the backside, so access to the rear of the unit is needed for balancing.

Installation layout must allow rear access for flow adjustment during commissioning.

Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Operation

The recommended thermofluid is a 50:50 water-monoethylene glycol mixture, which must be used to match performance and material compatibility.

Using the correct fluid ensures heat transfer performance and avoids corrosion or freezing issues.

Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation precautions are required.

Personnel must be aware of the refrigerant hazard class for safe installation and maintenance.

Source
“R-449A (A1, H280)”
Safety

The unit has IP52 protection, so it is protected against dust and dripping water but not against water jets or immersion.

Installation location must avoid water spray or flooding to maintain safety.

Source
“Degree of Protection IP52”
Performance

The circulation pump delivers up to 70 l/min at 3.7 bar with G3/4 male connections, which determines the external piping and pressure drop budget.

External circuit pressure drop must be checked against the available pump pressure.

Source
“max. delivery pressure 3,7 bar”
Performance

The tank filling capacity is 175 to 195 liters, so the fluid volume and expansion provisions must be planned accordingly.

Fluid volume affects commissioning cost and expansion vessel sizing.

Source
“min. filling capacity 175 l”
Conformité

The equipment emits Class A Group 1 disturbances under EN55011, so it is intended for industrial environments and may cause interference in residential areas.

EMC classification determines the permissible installation environment and need for filtering.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Conformité

The US-SNAP end-use for this equipment is industrial process refrigeration, which limits its permitted application in the US.

Application must fall within the allowed end-use to remain compliant.

Source
“The US-SNAP end-use for this equipment is the industrial process refrigeration.”
Maintenance

The condenser is air-cooled with a high-efficiency fan, so adequate air flow and cleaning access are needed for sustained cooling performance.

Maintenance planning must include condenser cleaning and unobstructed air flow.

Source
“Condenser in air-cooled design, performance-optimized by a built-in high-efficiency fan motor.”