Product introduction and general description of the Petite Fleur unit
Specp. 2
Technical specifications according to DIN 12876 standard
p. 2table, 12 rows
p. 1table, 32 rows
Technical data according to DIN 12876
Parameter
Value
Filling capacity expansion tank
2 l
Overall dimensions WxDxH **
260x450x504 mm
Net weight
45 kg
sound pressure level +/- 4 dB(A)
58 dB(A)
Power supply requirement
110-120V 1~ 60Hz
max. current
14,5 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S610246
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-40...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
1,5 kW
Cooling power with Thermooil at 200°C
0,48 kW
Cooling power with Thermooil at 150°C
0,48 kW
Cooling power with Thermooil at 100°C
0,48 kW
Cooling power with Thermooil at 50°C
0,48 kW
Cooling power with Thermooil at 20°C
0,48 kW
Cooling power with Ethanol at 0°C
0,45 kW
Cooling power with Ethanol at -20°C
0,27 kW
Cooling power with Ethanol at -30°C
0,16 kW
Cooling power with Ethanol at -40°C
0,04 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,08 kg
CO2-Äquvivalent
0,0 t
UN-number
UN 3358
Circulation pump: max. delivery
25 l/min
Circulation pump: max. delivery pressure
0,9 bar
Pump connection
M16x1 male
max. permissible kin. viscosity
50 mm²/s
min. filling capacity
1,5 l
Engineering Facts
12 facts · every claim linked to a page
Selection
The unit is rated for an operating temperature range of -40 to 200 °C, which defines the process window it can serve.
Confirms whether the circulator covers the required process temperatures before selection.
Source
“Operating temperature range -40...200 °C”
Selection
Cooling capacity drops sharply at low temperatures with ethanol, from 0.45 kW at 0 °C to 0.04 kW at -40 °C, so low-temperature heat loads must be sized accordingly.
Prevents oversizing expectations and ensures the process heat load is removable at the target temperature.
Source
“Cooling power with Ethanol at 0°C 0,45 kW at -20°C 0,27 kW at -30°C 0,16 kW at -40°C 0,04 kW”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects installation location and ventilation requirements.
Flammable refrigerant classification drives site safety and ventilation planning.
Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Installation
The unit requires a 110-120 V 1~ 60 Hz supply with a maximum current of 14.5 A and a 16 A fuse.
Determines electrical supply, breaker sizing and plug configuration at the installation site.
Source
“Power supply requirement 110-120V 1~ 60Hz max. current 14,5 A Fuse 16 A”
Installation
The device is IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in wet or hot environments.
Sets environmental limits for the installation location and enclosure requirements.
Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance
Cooling capacity is specified at 20 °C room temperature and may drop if ambient temperature rises.
Engineers must derate cooling performance in warm rooms to avoid undersized cooling.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but not simultaneously in opposite directions.
Defines acceptable mains quality for reliable operation and warranty compliance.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Conformité
The equipment is EMC classified as Class A, Group 1 to EN55011, which is intended for industrial environments.
Class A equipment may cause interference in residential areas and affects EMC planning.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.
Ensures the correct plug is supplied and avoids delays or rework at commissioning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The pump delivers up to 25 l/min at 0.9 bar and accepts fluids up to 50 mm²/s kinematic viscosity.
Confirms hydraulic capability for the external circuit and compatible heat transfer fluids.