Unichiller GL 075 — Technical Data

Sommaire

  1. Overview p. 1
    Product name and general introduction to the Unichiller GL 075
  2. Spec p. 2
    Technical data according to DIN 12876, including performance and dimensions
  3. p. 1 table, 23 rows

Technical data according to DIN 12876 — Unichiller GL 075 (Order-No.: 5506.0006.01)

CharacteristicValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,2 K
Safety classificationI / NFL
Cooling power at 20°C7,5 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
UN-numberUN 2857
max. delivery50 l/min
max. delivery pressure3,5 bar
Pump connectionG1 1/4 male
Volume of expansion9,5 l
Overall dimensions WxDxH **550x760x1455 mm
Power supply (3 Phase)460V 3~ 60Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.1.0/26

** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 40 °C, so process requirements outside this window cannot be met.

Defines the process temperature envelope for which the unit can be selected.

Source
“Operating temperature range -20...40 °C”
Performance

Cooling capacity is 7.5 kW at 20 °C, but it may drop as ambient temperature rises, so sizing must account for actual room conditions.

Prevents undersizing when the chiller is installed in a warm environment.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit requires a 460 V three-phase 60 Hz supply, so site power must match this configuration.

Determines electrical infrastructure and whether a transformer or different supply is needed.

Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the installer must provide a suitable connection.

Affects electrical installation planning and required site hardware.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Conformité

External branch circuit protection according to UL 489 is required, so the electrical design must include a UL 489 listed breaker.

Ensures the installation meets the manufacturer's electrical protection requirement.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but not all combinations are allowed, so supply quality must be verified.

Avoids invalid supply conditions that could damage the chiller or void compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The refrigerant is R-744 (CO2) with GHS hazard H280, so ventilation and safe handling procedures are required.

Informs site safety planning and refrigerant handling requirements.

Source
“R-744 (A1, H280)”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.

Restricts installation environment and prevents unsuitable outdoor or washdown locations.

Source
“Degree of Protection IP20”
Operation

The chiller is intended for externally closed applications, so the external circuit must be closed to avoid problems.

Guides the design of the external cooling loop and prevents open-loop misuse.

Source
“For externally closed applications.”
Installation

The pump connection is G1 1/4 male, so mating hoses or fittings must match this thread.

Ensures correct hydraulic connection and avoids adapters or leaks.

Source
“Pump connection G1 1/4 male”
Performance

Maximum delivery is 50 l/min at up to 3.5 bar, so external circuit pressure drop must be checked against these limits.

Prevents selecting a circuit with excessive flow resistance or required flow beyond pump capability.

Source
“max. delivery 50 l/min”
Installation

Ambient temperature must stay between 5 and 40 °C, so the installation room must be controlled within this range.

Avoids derating or malfunction caused by unsuitable room conditions.

Source
“min. ambient temperature 5 °C”
Conformité

The equipment is Class A, Group 1 for electromagnetic disturbance, so it may require EMC measures in residential or sensitive environments.

Informs EMC planning and potential restrictions on installation location.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

The expansion tank is atmospheric open and the unit has a digital level indicator, so fluid level must be monitored during operation.

Supports routine maintenance and prevents low-level operation faults.

Source
“With digital level indicator.”