Unichiller P007 Technical Datasheet

Sommaire

  1. Overview p. 1
    Introduces the Unichiller P007 chiller unit and its main features
  2. Spec p. 2
    Lists technical data according to DIN 12876 standard
  3. p. 2 table, 8 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current5 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S591725

Unichiller P007 — Technical data according to DIN 12876

ParameterValue
Order-No.3012.0317.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 15°C0,7 kW
at 0°C0,55 kW
at -10°C0,4 kW
at -20°C0,2 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,11 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Circulation pump:B
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **350x496x622 mm
Net weight56 kg
sound pressure level +/- 4 dB(A)58 dB(A)

** — сноска в исходном тексте не раскрыта; значение Cooling power at ambient temperature +20°C в тексте отсутствует (пустая строка).

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 40 °C, so process requirements outside this window cannot be met.

Defines the process temperature envelope the unit can support during selection.

Source
“Operating temperature range -20...40 °C”
Selection

Cooling capacity drops with lower temperature, from 0.7 kW at 15 °C to 0.2 kW at -20 °C, so the load must be checked at the actual set point.

Prevents oversizing or undersizing based on a single nominal capacity figure.

Source
“at 15°C 0,7 kW at 0°C 0,55 kW at -10°C 0,4 kW at -20°C 0,2 kW”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects room ventilation and ignition-source planning.

Flammable refrigerant classification drives installation safety and site requirements.

Source
“R-290 (A3, H220)”
Installation

The pump connection is G3/4 male, so mating hoses and fittings must match this thread.

Determines the hose coupling and adapter selection for the hydraulic circuit.

Source
“Pump connection G3/4 male”
Operation

The pump delivers 25 l/min max and 2.5 bar max pressure, with flow falling to 6 l/min at 2.0 bar, so system pressure drop must be matched to the pump curve.

Allows checking that the pump can overcome the external circuit pressure drop.

Source
“max. delivery 25 l/min max. delivery pressure 2,5 bar”
Installation

Minimum filling capacity is 3.8 l with 1.7 l expansion volume, so the loop must be sized and filled accordingly.

Ensures adequate heat transfer fluid volume and expansion allowance.

Source
“min. filling capacity 3,8 l Volume of expansion 1,7 l”
Installation

The unit requires 208-240 V single/two-phase 50/60 Hz supply with max 5 A current and a 16 A fuse.

Defines electrical supply and protection sizing for site preparation.

Source
“208-240V 1~/2~ 50/60Hz max. current 5 A Fuse 16 A”
Installation

The degree of protection is IP20, so the unit must be installed in a dry, protected environment.

Restricts installation location regarding moisture and dust ingress.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 and 40 °C, and cooling capacity may drop if ambient rises.

Sets the room conditions required for reliable operation and rated performance.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Operation

The plastic pump is not resistant to acetone and polyglycols, so these fluids must be avoided or a more resistant plastic requested.

Prevents chemical attack and premature pump failure from incompatible heat transfer fluids.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Guides correct heat transfer fluid preparation for sub-zero operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Conformité

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Ensures the site power quality stays within the permitted operating envelope.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Conformité

The equipment is EMC classified as Class A, Group 1 to EN55011, which has implications for residential or sensitive environments.

Class A equipment may require additional EMC measures in non-industrial locations.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.

Avoids delays and mismatched plugs by specifying the correct power connection at order.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”