Unichiller P025w-H OLÉ Datasheet

Sommaire

  1. Overview p. 1
    Product name and general description of the chiller unit
  2. Spec pp. 1-2
    Technical data, performance values and operating parameters
  3. Spec p. 2
    Detailed specifications according to DIN 12876 standard
  4. p. 2 table, 9 rows
  5. p. 1 table, 38 rows

Technical data according to DIN 12876

ParameterValue
sound pressure level +/- 4 dB(A)60 dB(A)
Power supply requirement220-240V 1~ 50Hz
max. current15,9 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.1/24

Technical data according to DIN 12876 — Unichiller P025w-H OLÉ (Order-No.: 3052.0002.98)

ParameterValue
Operating temperature range-10...100 °C
temperature set point / displaydigital
Internal temperature sensorPt100
Resolution of display0,1 K
Interface digitalUSB (Device), RS232
Interface
Temperature stability at -10°C0,2 K
Alarm messageoptic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at 15°C2,5 kW
Cooling power at 0°C2 kW
Cooling power at -10°C1,2 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,4 kg
CO2 equivalent0,6 t
Circulation pumpB
at 0,5 bar21 l/min
at 1,0 bar17 l/min
at 1,5 bar11 l/min
at 2,0 bar6 l/min
max. delivery25 l/min
max. delivery pressure2,5 bar
Pump connectionG3/4 male
Consumption at water 15°C, flow 15°C84 l/h
Consumption at water 15°C, flow 0°C84 l/h
Consumption at water 15°C, flow -10°C72 l/h
Cooling water connectionG1/2 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,8 l
expansion tank1,7 l
Overall dimensions WxDxH **420x487x579 mm
Net weight67 kg

** см. сноску в оригинале (Overall dimensions WxDxH **)

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water with a differential pressure of at least 1 bar and a maximum of 6 bar at the G1/2 male connection.

Ensures the water supply system can meet the minimum pressure for proper cooling and avoid overpressure damage.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is water-cooled and requires a cooling water intake temperature between 5°C and 20°C; below the condensation point, cooling water pipes must be insulated.

Defines the allowable cooling water temperature range and insulation requirement to prevent condensation and ensure performance.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Operation

Cooling power drops as cooling water temperature increases, and the cooling water flow rate may need to be increased.

Helps predict performance changes and size the cooling water supply accordingly.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Safety

The plastic pump is not resistant to acetone and polyglycols; for freeze protection, use water mixed with glysantine or ethylene glycol instead.

Prevents chemical attack on the pump and ensures compatibility of the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Installation

The power supply must be 220-240V 1~ 50Hz with a maximum current of 15.9 A and a 16 A fuse.

Ensures the electrical supply and protection are correctly rated for safe operation.

Source
“Power supply requirement 220-240V 1~ 50Hz”
Installation

Voltage tolerance is ±5% with a simultaneous frequency tolerance of ±2%, but combinations like -5% voltage and +2% frequency are not allowed.

Guides the acceptable power quality and prevents invalid supply conditions that could damage the unit.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation

The maximum ambient temperature for operation is 40°C and the minimum is 5°C.

Ensures the installation location maintains ambient conditions within the allowed range for reliable operation.

Source
“max. ambient temperature 40 °C”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so suitable cooling water must be used.

Helps assess water quality and corrosion compatibility with the wetted materials.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Selection

The chiller has a maximum delivery pressure of 2.5 bar and a maximum flow of 25 l/min, with flow decreasing at higher pressures.

Allows proper sizing of the external circuit and ensures the pump can overcome system pressure drops.

Source
“max. delivery pressure 2,5 bar”
Selection

The refrigerant is R-449A with a GWP of 1397 and a charge of 0.4 kg, corresponding to 0.6 t CO2 equivalent.

Relevant for environmental compliance, refrigerant handling, and leak reporting obligations.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Installation

The unit has a protection degree of IP20, so it must be installed in a dry environment protected from water ingress.

Determines the required environmental protection at the installation site.

Source
“Degree of Protection IP20”
Operation

The OLÉ controller includes an autostart function for power failure, which can restart the unit automatically after power restoration.

Affects operational safety and process continuity; may require additional precautions depending on the application.

Source
“Autostart function for power failure”