Installation
The unit requires a 460 V three-phase 60 Hz supply, so the site electrical infrastructure must be verified before ordering.
Mismatch with available site power would prevent commissioning and require a transformer or different model.
Source
“Power supply requirement 460V 3~ 60Hz”
Installation
Cooling water must be supplied at 3–6 bar differential pressure with a flow of 5000 l/h at 15 °C, which dictates the facility water system sizing.
Insufficient cooling water pressure or flow will degrade cooling capacity and may trip the unit.
Source
“Consumption at water 15°C, flow 0°C 5000 l/h”
Installation
The cooling water intake temperature must stay between 5 °C and 20 °C, and pipes must be insulated if below the condensation point.
Out-of-range cooling water temperature reduces cooling power and risks condensation and corrosion.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Aggressive or incompatible water chemistry can corrode or foul the circuit and void performance.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité
External branch circuit protection according to UL 489 is required and must be specified from the electrical schematics.
The installer must provide a correctly rated UL 489 breaker; the unit does not include it.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Three-phase devices drawing more than 63 A are delivered without cable and without plug, so the customer must supply and terminate the power feed.
This affects procurement of cabling, glands and the electrical installation scope.
Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Installation
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and combinations such as −5% voltage with +2% frequency are not allowed.
Site power quality must be checked; an out-of-tolerance combination can damage the compressor and electronics.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The unit is classified IP20, so it must be installed in a dry, protected environment free of water ingress.
IP20 offers no water protection; washdown or outdoor placement is not permissible.
Source
“Degree of Protection IP20”
Operation
Ambient temperature must remain between 5 °C and 40 °C for rated operation.
Operating outside this range affects cooling performance and reliability.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection
Cooling capacity falls sharply at low temperatures, from 45 kW at 0 °C to 6 kW at −60 °C with ethanol, which must be matched to the process heat load.
Sizing at the lowest operating temperature is critical; the unit cannot deliver high capacity at −60 °C.
Source
“at -60°C 6 kW”
Selection
The pump delivers 343 l/min maximum and 300 l/min at 1.0 bar, with pressure falling as flow rises, so system pressure drop must be checked.
External circuit pressure drop determines achievable flow and heat transfer.
Source
“max. delivery 343 l/min”
Installation
Pump connections are flange DN32 PN10/40 per EN1092-1, and hoses must use ball socket per DIN 3863 with sleeve nut per DIN 3870.
Correct mating flanges and fittings must be procured to avoid leaks and delays.
Source
“Pump connection flange DN32 PN10/40 EN1092-1”
Safety
The refrigerant is R-507 (A1, H280) with GWP 3985, which carries flammability and high global warming potential obligations.
Refrigerant handling, leak detection and environmental reporting requirements apply.
Source
“R-507 (A1, H280)”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, limiting fluid choice at low temperatures.
Too-viscous fluid at low temperature overloads the pump and reduces circulation.
Source
“max. permissible kin. viscosity 50 mm²/s”