Unistat GL 540w Temperature Control System Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and general description of the Unistat GL 540w unit
  2. Spec pp. 1-2
    Technical data, performance figures and operating parameters per DIN 12876
  3. p. 2 table, 13 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **918x963x1771 mm
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

CharacteristicValue
Operating temperature range-50...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 200°C28 kW
Cooling power with Thermooil at 100°C28 kW
Cooling power with Thermooil at 20°C28 kW
Cooling power with Ethanol at 0°C25 kW
Cooling power with Ethanol at -20°C17 kW
Cooling power with Ethanol at -40°C6,8 kW
Cooling power with Ethanol at -50°C4 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
Circulation pump: MK pump max. delivery196 l/min
Circulation pump: MK pump max. delivery pressure2,5 bar
Delivery at 0,5 bar175 l/min
Delivery at 1,0 bar152 l/min
Delivery at 1,5 bar123 l/min
Delivery at 2,0 bar86 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply, so the site must provide a 3~ 50 Hz feed sized for the 12 kW heater and 28 kW cooling load.

Electrical supply and protection must be planned before installation; single-phase power is not acceptable.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

Cooling water must be supplied at 1–4 bar differential pressure and no more than 6 bar static pressure, so the building water system must be checked against these limits.

Insufficient differential pressure reduces cooling capacity; excessive pressure can damage the water circuit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The cooling water inlet temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation occurs, which affects chiller/water supply design.

Cooling performance ratings are valid only within this water temperature window; outside it, capacity drops and condensation risk rises.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

The published cooling powers are valid only at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, so selection must account for actual site conditions.

If site water or ambient conditions differ, the 28 kW cooling power cannot be assumed.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and - outlet.”
Installation

The pump connection is M38x1.5 male and the cooling water connection is G3/4 male, so mating hoses and adapters must be ordered accordingly.

Mechanical interfaces determine which hoses, fittings and adapters are needed at installation.

Source
“Pump connection M38x1,5 male”
Operation

The circulation pump delivers 196 l/min at zero head but only 86 l/min at 2.0 bar, so external circuit pressure drop directly limits achievable flow.

External loop resistance must be calculated to confirm the required flow is available at the application.

Source
“max. delivery 196 l/min”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s, which constrains fluid selection at low temperatures.

Using a too-viscous fluid at low temperature can overload the pump and reduce heat transfer.

Source
“max. permissible kin. viscosity 100 mm²/s”
Safety

The unit uses R-744 (CO2) refrigerant with GWP 1 and is water-cooled, which affects refrigerant handling and ventilation requirements.

R-744 is a high-pressure refrigerant; service and room ventilation must be planned for it.

Source
“Refrigerant (ASHRAE, GHS) R-744 (A1, H280)”
Safety

The safety classification is III / FL and overtemperature protection is adjustable per DIN 12876, so the protection settings must be configured for the application.

Class III/FL determines the required safety strategy and protection settings for flammable or high-temperature operation.

Source
“Safety classification III / FL”
Installation

The unit is IP20 and must be installed at ambient temperatures between 5 °C and 40 °C, so it cannot be placed in a wet or condensing environment.

IP20 offers no water protection; installation location must be dry and within the ambient temperature range.

Source
“Degree of Protection IP20”
Installation

The overall dimensions are 918x963x1771 mm, so access and clearance for a unit of this footprint and height must be reserved.

Space planning and transport route must accommodate the unit's physical size.

Source
“Overall dimensions WxDxH ** 918x963x1771 mm”
Conformité

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are explicitly not allowed.

Site power quality must be verified; out-of-tolerance supply can void performance or damage the unit.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide the final connection.

The electrical installer must plan for a suitable disconnect and plug or hard-wired connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The minimum filling capacity is 8.9 l and the expansion tank holds 19.9 l, which sets the minimum fluid charge and expansion allowance.

Fluid volume and expansion tank sizing must be checked against the external loop volume.

Source
“min. filling capacity 8,9 l”