The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant, which affects safety classification and handling requirements.
Flammable refrigerant classification impacts installation location, ventilation, and safety compliance.
Source
“R-1270 (A3, H220)”
Installation
Cooling water differential pressure must be at least 1 bar and must not exceed 6 bar at the connection.
Ensures adequate cooling water flow and protects the unit from overpressure during installation.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit is designed for cooling water intake temperatures between 5°C and 20°C; below the condensation point the cooling water pipes must be insulated.
Defines acceptable cooling water supply conditions and required insulation to prevent condensation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops significantly at low temperatures with ethanol, from 1.3 kW at 0°C to 0.1 kW at -40°C.
Critical for selecting the unit for low-temperature applications where cooling capacity is limited.
Source
“at -40°C 0,1 kW”
Installation
The electrical supply must be 460V 3~ 60Hz, and voltage/frequency tolerances of ±5% and ±2% apply simultaneously.
Ensures correct power supply design and prevents invalid combinations such as -5% voltage with +2% frequency.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Conformité
External branch circuit protection according to UL 489 is required for this equipment.
Mandatory for electrical installation compliance in UL jurisdictions.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Three-phase devices with current consumption less than 63A are supplied with cable but without plug.
Installer must provide a suitable plug or hardwired connection at the site.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The expansion tank is thermally passive, preventing HTF vapour and moisture absorption, and for external open baths it is blocked off so the thermostat is atmospherically sealed.
Allows flexible placement below or above the application level and avoids moisture ingress.
Source
“No HTF vapour and no moisture absorption because the expansion tank is thermally passive.”
Installation
Pump connections use M30x1.5 male threads with bore shape Y (60°) according to DIN 3863, requiring matching ball socket and sleeve nut.
Ensures correct selection of hoses and fittings for leak-free connection.
Source
“Pump connection M30x1,5 male”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.
Limits choice of heat transfer fluid to ensure proper pump operation and cooling performance.
Source
“max. permissible kin. viscosity 50 mm²/s”
Installation
The unit has IP20 protection and must be operated in ambient temperatures between 5°C and 40°C.
Restricts installation environment to dry indoor locations and defines ambient temperature limits.