Product introduction and general description of the Unistat P425w
Specpp. 1-2
Technical data and performance specifications per DIN 12876
p. 2table, 21 rows
p. 1table, 30 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 2,0 bar
48 l/min
Delivery at 2,5 bar
30 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 0°C
150 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,6 l
Filling capacity expansion tank
3,3 l
Overall dimensions WxDxH **
460x554x1453 mm
Net weight
178 kg
sound pressure level +/- 4 dB(A)
61 dB(A)
Power supply requirement
400V 3~ 50Hz
max. current
10 A
Fuse
3x16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S469298
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unistat P425w (Order-No.: 1050.0033.01)
Parameter
Value
Operating temperature range
-40...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
2 kW
Cooling power with Thermooil at 250°C
2,8 kW
Cooling power with Thermooil at 200°C
2,8 kW
Cooling power with Thermooil at 100°C
2,8 kW
Cooling power with Thermooil at 20°C
2,8 kW
Cooling power with Ethanol at 0°C
2,5 kW
Cooling power with Ethanol at -20°C
1,8 kW
Cooling power with Ethanol at -40°C
0,1 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
0,97 kg
CO2 equivalent
2,1 t
Circulation pump: MK pump
max. delivery
97 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
86 l/min
Delivery at 1,0 bar
75 l/min
Delivery at 1,5 bar
63 l/min
Значения приведены дословно по тексту страницы; строки «Cooling power with Thermooil» и «Cooling power with Ethanol» объединены с соответствующими температурными точками, так как в исходном тексте они напечатаны как отдельные подзаголовки.
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply and breaker/fuse rating needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
The cooling water circuit must be supplied at 1–6 bar with 150 l/h at 15 °C, and the intake water must stay between 5 °C and 20 °C.
Cooling water pressure, flow and temperature limits dictate the facility water supply design.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water is colder than the condensation point, the cooling water pipes inside the unit must be insulated.
Prevents condensation and corrosion/electrical risk when using cold facility water.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and material compatibility to avoid corrosion or fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
The pump connection is M30x1.5 male and the cooling water connection is G1/2 male, so matching hoses/fittings must be ordered.
Ensures correct plumbing adapters and hoses are available at installation.
Source
“Pump connection M30x1,5 male”
Installation
Pump connections use bore shape Y (60°) per DIN 3863 with ball socket and sleeve nut per DIN 3870, so standard DIN fittings are required.
Specifies the exact fitting standard to avoid leaks or mismatched connections.
Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Installation
The unit is IP20 and must be installed where ambient temperature stays between 5 °C and 40 °C.
Defines environmental limits and protection against dust/water at the installation site.
Source
“Degree of Protection IP20”
Installation
The unit weighs 178 kg and measures 460x554x1453 mm, so floor loading and access clearance must be planned.
Needed for rigging, floor loading and space planning during installation.
Source
“Overall dimensions WxDxH ** 460x554x1453 mm”
Operation
The operating temperature range is -40 to 250 °C with 0.01 K stability, defining the process window.
Confirms the unit can cover the required process temperatures with high stability.
Source
“Operating temperature range -40...250 °C”
Performance
Cooling power with ethanol drops to 0.1 kW at -40 °C, so low-temperature capacity is very limited.
Critical for sizing heat loads at the cold end of the range.
Source
“at -40°C 0,1 kW”
Performance
The pump delivers 97 l/min max and 3 bar max, with 86 l/min at 0.5 bar, so pressure-drop calculations are needed.
Determines whether the pump can overcome the external circuit pressure drop.
Source
“max. delivery 97 l/min”
Safety
Leakage current exceeds 3.5 mA, so protective earth and RCD selection must account for it.
Affects electrical safety design and residual current device selection.
Source
“Attention: leakage current > 3,5mA”
Conformité
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and -5% voltage with +2% frequency is not allowed.
Prevents operation outside EN60034-1 tolerances that could damage the compressor/pump.
Source
“-5% voltage and +2% frequency -> not allowed!”
Conformité
The refrigerant is R-452A with GWP 2141 and 0.97 kg charge (2.1 t CO2e), relevant for F-gas reporting.
Needed for refrigerant compliance, leak reporting and CO2-equivalent accounting.