Unistat P610w Technical Data

Sommaire

  1. Overview p. 1
    Product introduction and general description of the Unistat P610w
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 2 table, 20 rows
  4. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 2,5 bar36 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C400 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity6,5 l
Filling capacity expansion tank26 l
Overall dimensions WxDxH **630x704x1520 mm
Net weight358 kg
sound pressure level +/- 4 dB(A)62 dB(A)
Power supply requirement400V 3~ 50Hz
max. current16 A
Fuse3x20A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S452288

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

CharacteristicUnistat P610w
Order-No.1052.0001.01
Operating temperature range-60...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 200°C7 kW
Cooling power with Thermooil at 100°C7 kW
Cooling power with Ethanol at 0°C7 kW
Cooling power with Ethanol at -20°C6,4 kW
Cooling power with Ethanol at -40°C2,6 kW
Cooling power with Ethanol at -60°C0,05 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity2,55 kg
CO2 equivalent5,5 t
Circulation pump: MK pump max. delivery82 l/min
Circulation pump: MK pump max. delivery pressure3 bar
Delivery at 0,5 bar75 l/min
Delivery at 1,0 bar68 l/min
Delivery at 1,5 bar59 l/min
Delivery at 2,0 bar49 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 16 A maximum current and 3x20 A fusing, so the electrical installation must be sized accordingly.

Determines the site power supply, cable sizing and breaker rating needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 16 A Fuse 3x20A”
Installation

The cooling water circuit must be supplied at 5–20 °C with at least 1 bar differential pressure and no more than 6 bar, and the pipes must be insulated if condensation can occur.

Defines the cooling water utility requirements and condensation protection needed at the installation site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Prevents corrosion or damage by selecting suitable cooling water quality and materials.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

The pump connection is M30x1.5 male and the cooling water connection is G1/2 male, so matching hoses and fittings must be provided.

Ensures correct mechanical interface for process and cooling water piping during installation.

Source
“Pump connection M30x1,5 male”
Installation

The unit weighs 358 kg and measures 630x704x1520 mm, so floor loading and access space must be planned.

Needed for transport, rigging and allocating sufficient installation space.

Source
“Overall dimensions WxDxH ** 630x704x1520 mm Net weight 358 kg”
Operation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be used in wet or condensing environments.

Defines the permissible operating environment and protection against ingress.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Operation

The cooling capacity drops sharply at low temperatures, from 2.6 kW at -40 °C to only 0.05 kW at -60 °C, so the -60 °C limit is only usable with very small loads.

Critical for selecting the unit when low-temperature operation with real heat loads is required.

Source
“at -40°C 2,6 kW at -60°C 0,05 kW”
Operation

The pump delivers 82 l/min maximum and 36 l/min at 2.5 bar, so the external circuit pressure drop must be checked against the pump curve.

Ensures the circulation pump can overcome the pressure drop of the external application.

Source
“max. delivery 82 l/min max. delivery pressure 3 bar”
Safety

The refrigerant is R-452A with GWP 2141 and a CO2 equivalent of 5.5 t, so refrigerant leak monitoring and F-gas obligations apply.

Affects environmental compliance, leak checks and service requirements for the refrigeration circuit.

Source
“Refrigerant (ASHRAE, GHS) R-452A (A1, H280) Global Warming Potential (GWP) 2141”
Safety

The leakage current exceeds 3.5 mA, so a suitable protective earth and residual current device must be provided.

Important for electrical safety and selection of the RCD and earthing arrangement.

Source
“Attention: leakage current > 3,5mA”
Conformité

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.

Prevents operation outside the permitted supply quality and avoids invalid tolerance combinations.

Source
“-5% voltage and +2% frequency -> not allowed!”
Conformité

The equipment is EMC class A, group 1, so it may cause radio interference in residential areas and is intended for industrial environments.

Determines EMC compatibility and suitability for the intended industrial installation site.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Three-phase devices below 63 A are supplied with a cable but without a plug, so the customer must provide the final connection.

Clarifies the scope of supply and the need to arrange a suitable plug or direct connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The unit has a water-cooled refrigeration machine and requires 400 l/h cooling water at 15 °C with 0 °C flow, so the cooling water supply must be sized for this flow.

Needed to verify the available cooling water flow rate at the installation site.

Source
“Consumption at water 15°C, flow 0°C 400 l/h”