Quick Reference Guide: Gas Chromatography Detectors
Technical Reference Summary
Covered Models & Document Scope
Descripción general de los detectores de GC de Agilent: FID, NPD (N), NPD (P), µECD, TCD, FPD (S), FPD (P), NCD, SCD. Abarca la descripción, el rango dinámico lineal, la destructividad y los analitos típicos de cada detector.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Linear dynamic range | >107 (±10%) p.1 ↗ | FID |
| Destructive | Yes p.1 ↗ | FID |
| Linear dynamic range | 105 (±10%) p.1 ↗ | NPD (N) |
| Destructive | No p.1 ↗ | NPD (N) |
| Linear dynamic range | >5 × 104 p.1 ↗ | µECD |
| Linear dynamic range | >104 p.2 ↗ | NCD |
| Destructive | Yes p.2 ↗ | NCD |
| Linear dynamic range | >104 p.2 ↗ | SCD |
| Destructive | Yes p.2 ↗ | SCD |
| Relative sensitivity | 10–15 g g p.2 ↗ | NPD (P), 1 ppt |
| Relative sensitivity | 10–12 g g p.2 ↗ | NPD (N), 1 ppb |
| Relative sensitivity | 10–9 g g p.2 ↗ | µECD, 1 ppm |
| Relative sensitivity | 10–6 g g p.2 ↗ | FID, 0.1% |
| Relative sensitivity | 10–3 g g p.2 ↗ | TCD, % |
Operating Limits
- Detection limits and ranges depend on compound structure and analytical conditions: See GC data sheets for actual specifications p.2 ↗