MO302
Technical Reference Summary
Covered Models & Document Scope
Fiche technique d'un système de spectrométrie de masse MALDI contrôlé par le logiciel LAUNCHPAD™ ; couvre la manipulation des échantillons, la visualisation des échantillons, la source d'ionisation, le laser, l'analyseur, le détecteur, le système de contrôle/Données et le logiciel. Aucune désignation de modèle, taille ou code de commande n'est imprimé.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Mass range | 1 Da to 500 kDa p.2 ↗ | linear mode |
| Mass accuracy | <30 ppm p.2 ↗ | linear mode, with internal calibration |
| Sensitivity | 250 fmol p.2 ↗ | linear mode, bovine serum albumin |
| XY stage step | 10 µm p.1 ↗ | microtiter plate footprint MALDI target |
| XY stage repeatability | 10 µm p.1 ↗ | microtiter plate footprint MALDI target |
| Target thickness | 2 mm p.1 ↗ | plain, 96 and 384 sample targets |
| Maximum target thickness | 10 mm p.1 ↗ | with optional adaptor |
| Turbomolecular pump pumping speed | 250 l/s p.1 ↗ | nominal, sample introduction, rotary backing |
| Turbomolecular pump pumping speed | 250 l/s p.1 ↗ | nominal, analyzer, two pumps with rotary backing |
| Camera magnification | 25 x p.1 ↗ | monochrome CCD camera |
| Ion extraction energy | +25 kV/-20 kV kV p.1 ↗ | linear mode, variable, software controlled |
| Ion extraction energy | +20 kV/-20 kV kV p.1 ↗ | reflectron mode, variable, software controlled |
| Laser wavelength | 337 nm p.1 ↗ | nitrogen laser, fixed focus |
| Laser pulse width | 3 ns p.1 ↗ | |
| Laser energy | 100 µJ p.1 ↗ | nominal, per laser shot |
| Maximum laser pulse rate | 50 Hz p.1 ↗ | 50 laser shots per second |
| Linear flight tube drift length | 1.2 m p.1 ↗ | linear mode |
| Reflectron effective drift length | 2.0 m p.1 ↗ | reflectron mode |
| Transient recorder sampling rate | 2 GHz p.1 ↗ | 8 bit, 16 bit accumulator |
| Second transient recorder sampling rate | 125 MHz p.1 ↗ | neutral detection, 8 bit, 64 kB RAM |
| RAM | 16 GB p.1 ↗ | control PC |
| HDD storage | 256 GB p.1 ↗ | control PC |
| SSD storage | 1 TB p.1 ↗ | control PC |
| Monitor size | 24 in p.1 ↗ | control PC |
| Operating system | Microsoft® Windows® 10 64-bit p.1 ↗ | control PC and LAUNCHPAD™ |
Operating Limits
- Maximum laser pulse rate: 50 Hz p.1 ↗
- Maximum target thickness with optional adaptor: 10 mm p.1 ↗
- Mass range upper limit: 500 kDa p.2 ↗