High-Sensitivity Analysis of Aldosterone in Low-Volume Serum Samples using Micro-Flow LC-MS/MS for Clinical Research

Technical Reference Summary

Covered Models & Document Scope

Nota de aplicación / folleto de método para el análisis de alta sensibilidad de aldosterona en muestras de suero de bajo volumen utilizando un sistema LC-MS/MS de microflujo Shimadzu Nexera Mikros (LC-Mikros, LC-30AD, SIL-30AC, FCV-32AH, CTO-Mikros) acoplado a un espectrómetro de masas de triple cuadrupolo LCMS-8060 con fuente micro-ESI, en configuración trap-and-elute. Research Use Only; no para procedimientos de diagnóstico; no disponible en China.

Technical Data

Parameter Value Condition
Mass spectrometer system LCMS-8060 p.1 ↗
Ionization Micro-ESI p.1 ↗ negative ionization
Probe voltage -2.4 kV p.1 ↗ negative ionization
Interface temperature 125 °C p.1 ↗
Desolvation line temperature 250 °C p.1 ↗
Heater block temperature 300 °C p.1 ↗
Nebulizing gas flow 3 L/min p.1 ↗
Heating gas flow 3 L/min p.1 ↗
Drying gas Off p.1 ↗
Dwell time 115 ms p.1 ↗
MRM transition (quantifier), aldosterone 359.20 > 189.30 m/z p.1 ↗ negative
MRM transition (qualifier), aldosterone 359.20 > 331.35 m/z p.1 ↗ negative
MRM transition (quantifier), d4-aldosterone 363.00 > 190.20 m/z p.1 ↗
Trapping column CERI C8 5 µm 5*0.3 mm p.1 ↗
Trapping flow rate 100 µL/min p.1 ↗
Microbore column Shimadzu PLONAS C18 2.7 µm 100*0.2 mm p.1 ↗
Analytical flow rate 4 µL/min p.1 ↗
Serum sample volume 10 µL p.1 ↗
Calibration range 1 to 2 500 pg/mL p.1 ↗
Lower limit of quantitation (LOQ) 1 pg/mL p.1 ↗ QC, n=3 replicates
Intra-day precision (%RSD) < 8 % p.1 ↗ real samples
LOQ accuracy 103 (mean), 94.4 (min), 114 (max) % p.1 ↗ QC LOQ 1 pg/mL
LOQ precision (%RSD) 5.2 intra-day, 4.9 inter-day % p.1 ↗ QC LOQ 1 pg/mL
Matrix effect recovery 99 (QC), 103 (solution) % p.1 ↗

Operating Limits

  • Lower limit of quantitation (LOQ): 1 pg/mL p.1 ↗
  • Calibration range: 1 to 2 500 pg/mL p.1 ↗
  • Minimum serum sample volume: 10 µL p.1 ↗
  • Organic solvent addition to plasma: max 0.5% (v/v) p.1 ↗
  • Intra-day precision (%RSD) for real samples: < 8% p.1 ↗
  • LOQ acceptance criteria: accuracy 80-120%, %RSD < 20% p.1 ↗