Agilent Technologies 1220 Infinity II LC System
Agilent Technologies 1220 Infinity II LC System — Liquid chromatography.
The 1220 Infinity II LC System is a compact, integrated HPLC platform for analytical-scale liquid chromatography, including cannabinoid potency testing in industrial hemp. It is positioned below the 1260 Infinity II series, which adds higher-pressure and bio-inert variants. The system is used for routine quantification and method development where a simple, self-contained LC is sufficient.
Agilent Technologies Analytical HPLC Systems Series — Model Comparison & Selection
| Model | Flow rate, liter/minute | Sample volume, milliliter | Column temperature, °C | Data rate, Hz | Operating altitude, m | Weight, kg |
|---|---|---|---|---|---|---|
| 1220 Infinity II LC System Current | — | 1 | 50 | 80 | 2000 | 30 |
| 1260 Infinity III LC System | 0.6 | 0.005 | 40 | 120 | — | — |
| 1260 Infinity III Prime LC System | 9×10⁻⁴ | 0.002 | 40 | 40 | 3000 | 11.5 |
Full specifications
| Wavelength range | 190...600 nm |
| Centrifugation Speed | 5000 rpm |
| Column temperature | 50 °C |
| Data Rate | 80 Hz |
| Extraction Solvent Volume | 20 mL |
| Filter Membrane Diameter | 47 mm |
| Filter Membrane Pore Size | 0.2 μm |
| Filter Pore Size | 0.45 μm |
| Internal Volume | 6 µL |
| Light source | Deuterium lamp |
| Linearity | 2 limit |
| Maximum Sampling Rate | 80 Hz |
| Ptfe Frits Per Pm Kit | 5 pk |
| Sample Capacity | 100 tray |
| Sample volume | 1 microvial |
| Solvent Bottle Volume | 1 L |
| Solvent Filtration Assembly Flask Volume | 1 L |
| Solvent Filtration Assembly Funnel Volume | 250 mL |
| Solvent Inlet Filter Pore Size | 10 μm |
| Temperature Accuracy | 0.8 °C |
| Well-plate Closing Mat Quantity | 50 pk |
| Well-plate Volume | 0.5 mL |
| Target compounds quantified mfr | 11 ytocannabinoids |
| Target cannabinoid mfr | Cannabidivarin Cannabidivarin |
| Regulatory THC limit for industrial hemp mfr | 0.3 |
| Total THC calculation factor mfr | 0.877 |
| Quick Connect heat exchanger flow rate mfr | 2.5 mL/min |
| Capillary dimensions mfr | 0.12 mm |
| Preventive maintenance cost reduction mfr | 30 |
| Pump seals per PM kit mfr | 4 |
| Seal cap assembly sets per PM kit mfr | 2 |
| Gold seals for outlet valve per PM kit mfr | 2 |
| Seal washes per PM kit mfr | 4 |
| Solvent bottles per convenience kit mfr | 3 amber |
| Safety standards mfr | IEC, CSA, UL UL |
| Hydraulic system mfr | Dual plunger in series pump with proprietary ser |
| Flow range mfr | 0.001...10 mL/min |
| Flow precision mfr | 0.07 greater |
| Flow accuracy mfr | 1 greater |
| Pressure operating range mfr | 40 mL/min |
| Pressure pulsation mfr | 2 greater |
| Compressibility compensation mfr | User-selectable, based on mobile phase compressi |
| Recommended pH mfr | 1.0...12.5 |
| Gradient formation (optional) mfr | Low pressure binary mixing/gradient capability u |
| Delay volume mfr | 600...900 pressure |
| Composition precision mfr | 0.2 greater |
| Gradient formation mfr | Low pressure binary mixing/gradient capability u |
| Injection range mfr | 0.1...100 increments |
| Replicate injections mfr | 1...99 vial |
| Carry over mfr | 0.1 cleaning |
| Injection cycle time mfr | 50 volume |
| Column capacity mfr | 1 cm |
| Detection type mfr | 1024 array |
| Wavelength accuracy mfr | 1 nm |
| Flow cells mfr | 14 maximum |
| Number of signals mfr | 8 |
| Flow cell mfr | 13 maximum |
| Time programmable mfr | Wavelength, polarity, peak width, lamp bandwidth |
| Number of target cannabinoids mfr | 11 |
| Precision mfr | 1 μL |
| extraction shake time mfr | 10 minutes |
| oil aliquot volume mfr | 100 μL |
| oil dilution flask volume mfr | 10 mL |
| ethanol volume for oil dilution mfr | 8 mL |
| second dilution aliquot mfr | 100 μL |
| second dilution solvent volume mfr | 900 μL |
| autosampler vial volume mfr | 2 mL |
| Control and data evaluation mfr | Agilent EZChrom Compact, Agilent ChemStation, Ag |
| Communications mfr | -232 LAN |
| GLP features mfr | Early maintenance feedback , electronic records |
| IP address: mfr | 192.168.254.11 |
| Subnet Mask: mfr | 255.255.255.0 |
| Default Gateway mfr | not specified specified |
| ? mfr | displays syntax and descriptions of commands com |
| ip <x.x.x.x> mfr | sets new ip address address |
| sm <x.x.x.x> mfr | sets new subnet mask mask |
| gw <x.x.x.x> mfr | sets new default gateway gateway |
| exit mfr | exits shell and saves all changes changes |
| Recommended flow mfr | 0.2...10.0 mL/min |
| Slit Width | 6.5 range |
| Weight | 30 kg |
| Dimensions (height × width × depth) mfr | 640 mm |
| Diode width mfr | 1 nm |
| Line voltage mfr | 100...240 |
| Line frequency mfr | 50 |
| Power consumption mfr | 240 BTU |
| Quick Turn Fitting Pressure Limit | 800 bar |
| Solvent bottle thread mfr | 45 |
| Sample weight (plant material) mfr | 200 mg |
| Recommended pH range mfr | 1.0...12.5 |
| Sample viscosity range mfr | 0.2...50 cp |
| sample weight (plant material) mfr | 200 mg |
| Noise | 1 nm |
| Operating Altitude | 2000 m |
| Humidity mfr | 95 °C |
| Non-operating altitude mfr | 4600 m |
| Short term noise (ASTM) mfr | 0.25 nm |
| Short term noise (ASTM) Single and Multi-Wavelength mfr | 0.7 nm |
| Instrument part number mfr | 8417 AA |
Description
Construction and positioning
- The 1220 Infinity II LC System is an integrated analytical-scale HPLC, distinct from the 1260 Infinity II GPC/SEC and purification siblings listed. It is the entry-level member of the Infinity II family, intended for standard-bore analytical work rather than high-performance GPC/SEC or preparative purification.
- The system is supplied as a complete LC package; the application brief references the 1220VL LC/VWD configuration for hemp potency testing, indicating a variable-wavelength detector is part of the standard analytical setup.
Behavior and application scope
- The system is applied to the quantification of 11 cannabinoids in industrial hemp, with a sample weight of 200 mg plant material and a regulatory THC limit of 0.3% for industrial hemp. Total THC is calculated using a factor of 0.877.
- No digitized curve points or performance ranges (flow, pressure, temperature) are provided for this model; the only temperature fact is a 50 °C column temperature, which should be treated as an application condition rather than a system limit.
Installation and operation
- The application brief describes sample preparation for both plant material and oils/concentrates, but no installation, connection, or control specifics are given in the supplied digests. The troubleshooting guide (5994-0709EN) covers general LC best practices without model-specific instructions.
- The supplies flyer lists compatible consumables for the 1220 Infinity II series, including inline filters, fittings, capillaries, solvent bottles, filtration assemblies, and preventive maintenance kits, but does not define installation requirements.
Integration & Operating Notes
The PM kit contains 2 pump seals, 2 seal cap assembly sets, 2 gold seals for the outlet valve, and 4 seal washes. Plan seal replacement according to the maintenance schedule; the kit is not part of the base delivery and must be ordered separately.
Selection & Integration Q&A
Choose the 1220 for routine analytical HPLC, including cannabinoid potency testing in hemp, where a compact integrated system is sufficient. The 1260 siblings add specialized capabilities: amino acid analysis, bio-inert flow paths, or analytical-scale purification. Select the 1260 when your method requires those dedicated configurations.
Use 10 µm solvent inlet filters, 47 mm/0.2 µm filter membranes for mobile-phase filtration, and quick-turn fittings rated to 800 bar. Solvent bottles are 1 L with a 45 mm thread. Capillaries of 0.17 mm ID are listed. These supplies are not included in the base delivery and must be ordered separately.
The supplied documents do not specify mounting, ambient temperature, humidity, or duty-cycle limits for this model. The only temperature fact is a 50 °C column temperature, which is an application condition, not a system limit. Consult the site preparation guide or Agilent for environmental requirements.
The preventive maintenance kit includes 2 pump seals, 2 seal cap assembly sets, 2 gold seals for the outlet valve, and 4 seal washes. These are not included in the base delivery. Also stock 10 µm inlet filters and 47 mm/0.2 µm filter membranes. No service interval is stated in the supplied documents.