Agilent Technologies 4200 TapeStation
Agilent Technologies 4200 TapeStation — Automated electrophoresis.
The Agilent 4200 TapeStation is an automated electrophoresis instrument for DNA and RNA sample quality control, using ScreenTape consumables to process up to 16 samples per device with a 1 µL sample requirement. It provides sizing from 35 to 1,000 bp and quantitative range of 25–500 ng/µL, with 16-minute analysis time. This model is distinguished within the Tapestation series by its 96-well plate compatibility and robotics for higher-throughput workflows.
Agilent Technologies Tapestation Instruments Series — Model Comparison & Selection
| Model | Sensitivity | Dimensions w d h | Weight, kg | Storage temperature, °C | Analysis time, min | Application |
|---|---|---|---|---|---|---|
| 4200 TapeStation Current | 5 | 435 | 25.5 | -40...70 | 16 | — |
| 4150 TapeStation | 5 | 254 | 17.8 | -40...70 | 16 | — |
| 4150 TapeStationDx | — | — | — | — | — | automated electrophoresi |
| 4200 TapeStationDx System | — | — | — | — | — | automated electrophoresi |
Full specifications
| Sensitivity | 5 ng/µL |
| Analysis time | 16 min |
| Dimensions (w × D × H) | 435 mm |
| Kit Stability | 6 months |
| Quantitative precision mfr | 10 % CV |
| Quantitative accuracy mfr | 20 |
| Quantitative range mfr | 25...500 ng/μL |
| RINe functional range mfr | 25...500 ng/μL |
| Maximum buffer concentration in sample mfr | 200 NaCl |
| Sample volume required mfr | 1 μL |
| Kit size mfr | 112 samples |
| Sizing range mfr | 35...1,000 bp |
| Safety standards mfr | 2 only. |
| ISM Classification mfr | 1 |
| Computer Specifications mfr | Checks the computer specifications and configura |
| Insufficient Mixing mfr | Buffer and sample was not sufficiently vortexed. |
| Sample concentration outside recommended range for application mfr | Linear concentration estimation cannot be achiev |
| Agilent D1000 ScreenTape Assay mfr | 2991 |
| Agilent High Sensitivity D1000 ScreenTape Assay mfr | 2991 |
| Agilent RNA ScreenTape Assay mfr | 2991 |
| Agilent High Sensitivity RNA ScreenTape Assay mfr | 2991 |
| Agilent Genomic DNA ScreenTape Assay mfr | 2991 |
| Agilent D5000 ScreenTape Assay mfr | 2991 |
| Agilent High Sensitivity D5000 ScreenTape Assay mfr | 2991 |
| Agilent Cell-free DNA ScreenTape Assay mfr | 2991 |
| Needle To Pump mfr | Yes Yes |
| Tape Handling mfr | Yes Yes |
| Sample concentration outside recommended range for application. mfr | Linear concentration estimation cannot be achiev |
| Sample prepared too long before analysis mfr | Evaporation or settling may alter the sample to |
| Incorrect Marker peaks picked up mfr | Depending on the sample pattern, wrong peaks can |
| Clear space around instrument mfr | 75 mm |
| Additional space above instrument for lid mfr | 25 cm |
| Minimum wait before unpacking after shipment mfr | 5 h |
| Vortex mixing speed mfr | 2000 rpm |
| Vortex mixing time mfr | 1 minute |
| Minimum distance from magnets for pacemaker wearers mfr | 55 mm |
| Minimum distance from magnets for sensitive devices mfr | 25 mm |
| Weight | 25.5 kg |
| Power consumption | 50 W |
| Line voltage mfr | 100...240 |
| Line frequency mfr | 50 |
| Interface mfr | USB USB |
| Incorrect protocol used mfr | Differences exist between Standard and High Sens |
| Samples per consumable mfr | 16 |
| Operating Altitude | 3000 m |
| Sound pressure | 70 decibel |
| Storage temperature | -40...70 °C |
| Operating humidity mfr | 15...80 condensing |
| Software Environment mfr | Checks the compatibility of TapeStation Controll |
Description
Construction and positioning within the series
The 4200 TapeStation (models G2991A and G2991B) is a benchtop automated electrophoresis system that integrates robotics, fluidics, and optical detection to analyze nucleic acid samples. It sits above the 4150 TapeStation in the Tapestation series, adding a 96-well plate format and extended sample capacity. The system uses ScreenTape devices—single-use consumables with 16 lanes each—and processes up to 16 samples per consumable. The instrument includes a piercing needle, electrode cartridge, and a camera-based detection system. It is controlled by the TapeStation Controller software running on a connected laptop.
Behavior under load and assay performance
- The 4200 supports a portfolio of ScreenTape assays: D1000, High Sensitivity D1000, D5000, High Sensitivity D5000, RNA, High Sensitivity RNA, Genomic DNA, High Sensitivity Genomic DNA, and Cell-free DNA. Sizing range is 35–1,000 bp, and quantitative range is 25–500 ng/µL for both DNA and RNA (RINe functional range). Sensitivity is 5 ng/µL. Analysis time is 16 minutes per run. Quantitative precision is 10% CV and accuracy 20%. The maximum buffer concentration in sample is 200 mM NaCl. Kit stability is 6 months.
- For Genomic DNA, D5000, and High Sensitivity D5000 assays, correct ladder peak assignment is mandatory; otherwise no sizing or quantification results are obtained. In Genomic DNA, a DIN value is not presented if a lane contains more than 14 peaks. RNA assays require sufficient denaturation to avoid extra peaks at the 28S position.
Installation, connection, and control specifics
The 4200 TapeStation connects to a laptop via USB. The TapeStation Controller software manages instrument communication and includes System Diagnostics with Software Installation Verification, System Verification, and Hardware Troubleshooting test suites. These tests check mechanical, electrical, and optical components using the TapeStation Test Tape (PN 5067-5601). Hardware Troubleshooting takes approximately 30 minutes. The optional liquid handling test requires a foil-sealed 96-well plate with specific water volumes in designated wells and two tube strips. The recommended foil seal for sample plates is PN 5067-5154. If no valid TapeStation system is connected, only Software Installation Verification tests can be run. Tape Handling, Teach Pump-ML, and Teach Pump-ML-Z tests are specific to G2991A and G2991B instruments.
Integration & Operating Notes
If the system detects insufficient lanes during a run, a second ScreenTape device with sufficient unused lanes must be added within 15 minutes to finalize the analysis. Keep spare ScreenTape devices on hand and respond promptly to lane-insufficiency messages to avoid losing the run.
Selection & Integration Q&A
Choose the 4200 when you need 96-well plate compatibility and higher throughput; it processes up to 16 samples per ScreenTape and supports robotics for automated plate handling. The 4150 is limited to 16 samples per run without plate automation. For lower sample numbers, the 4150 may suffice.
The instrument connects via USB to a laptop running TapeStation Controller software. Ensure the laptop meets the software specifications and that only one instance of the software is running. Use the Windows Device Manager to verify USB communication. If connection fails, check USB cable and port, and restart the software.
Replace the piercing needle before 4480 samples (warning at 3840) and the electrode cartridge annually or when messages appear. ScreenTape devices, Test Tape (5067-5601), foil seals (5067-5154), and sample plates are consumables not included. The Test Tape is required for diagnostics. Schedule maintenance every 6 months for kit stability.