Agilent Technologies 850-DS Sampling Station
Agilent Technologies 850-DS Sampling Station — Dissolution testing.
The Agilent 850-DS G7930A is a benchtop dissolution sampling station with an integrated 8-syringe rotary piston pump, 0.1–14 mL sample volumes and up to 36 timepoints per method. It differs from the optional G7931A filtration module by omitting automated filtration, and serves as the sampling front end for 708-DS, Apparatus 3/7 and NanoDis systems.
Full specifications
| Wavelength range | 190 nm |
| Baseline Flatness | 0.001 Abs |
| Photometric accuracy | 0.005 Abs |
| Printer | Optional internal thermal printer |
| Filtration pore size (optional filter module) mfr | 0.2 μm |
| Sample collection formats mfr | 96 plates |
| Pump type mfr | integrated syringe pump pump |
| Pump speed mfr | variable variable |
| Automated cleaning mfr | rinse port enables automated cleaning cycle cycl |
| Simultaneous system control mfr | four systems systems |
| Software package mfr | Cary WinUV WinUV |
| Number of apparatus mfr | 1...2 |
| Lamp type mfr | Xenon flash flash |
| Instrument design mfr | Scanning Scanning |
| Wavelength accuracy mfr | 0.5 nm |
| Wavelength reproducibility mfr | 0.1 nm |
| Stray light mfr | 1.0 |
| Safety standards mfr | 2 only |
| Pumping system and speed mfr | 8 mL/min |
| Sampling accuracy mfr | 10 |
| Sample volume per vial/tube mfr | 0.1 mL |
| Number of samples per test mfr | 36 change |
| Maximum run duration mfr | 999 hours |
| Filtration options mfr | 1 μm |
| Valves mfr | PEEK body body |
| Needles mfr | Anti-clogging stainless steel tips to prevent cl |
| Volumetric calibration mfr | Simplified routine with stored calibration coeff |
| Vials mfr | 12 mm |
| Test tubes mfr | 16 mm |
| Cleaning volume mfr | 0.1 mL |
| Microtiter plates mfr | 96 tray |
| Method storage mfr | 20 |
| Method export mfr | Printer, SD card or serial interface interface |
| PC instrument control mfr | 1.7 Dissolut |
| Good laboratory practice features mfr | User management via firmware or software, error |
| System and pumping speed mfr | 8 mL/min |
| Sampling precision mfr | 10 |
| Path lengths mfr | 0.2 mm |
| Système et vitesse de pompage mfr | 8 ajustable |
| Précision d’échantillonnage mfr | 10 |
| Volume d’échantillon par flacon/tube mfr | 0.1 mL |
| Fréquence d’échantillonnage mfr | 2 minutes |
| Nombre d’échantillons par test mfr | 36 plateau |
| Durée maximale d’exécution mfr | 999 heures |
| Options de filtration mfr | 1 μm |
| Vannes mfr | Corps en PEEK PEEK |
| Aiguilles mfr | Pointes en acier inoxydable anti-obturation pour |
| Étalonnage volumétrique mfr | Routine simplifiée avec un coefficient d’étalonn |
| Flacons mfr | 12 mm |
| Tubes à essai mfr | 16 mm |
| Types de milieux mfr | Compatible avec tous les milieux de dissolution |
| Remplacement de milieu mfr | 100 |
| Volume de nettoyage mfr | 0.1 chantillonnage |
| Clear (effacer) mfr | Efface toutes les valeurs entrées et remet les v |
| Save (enregistrer) mfr | Enregistre tous les paramètres de la méthode et |
| Cleaning Cycle (cycle de nettoyage) Volume mfr | Saisissez le volume qui sera pompé à travers les |
| # Cycles to Run (nombre de cycles à lancer) mfr | Ce paramètre contrôle le nombre de répétitions d |
| Run (Lancement) mfr | Lance un cycle de nettoyage avec les paramètres |
| Stop mfr | Arrête un cycle de nettoyage. nettoyage. |
| Plaques microtitres mfr | 96 plateau |
| Stockage de méthode mfr | 20 |
| Export de méthode mfr | Imprimante, carte SD ou interface série rie |
| Imprimante mfr | Imprimante thermique interne en option option |
| Fonctionnalités liées aux bonnes pratiques de laboratoire mfr | Gestion des utilisateurs via micrologiciel ou lo |
| Caisson mfr | Tous les matériaux sont recyclables recyclables |
| Spindle Speed (vitesse de l’arbre) mfr | Vitesse de l’arbre souhaitée pour cette méthode. |
| Rotation Start Delay (retard de démarrage de rotation) mfr | Temps après le démarrage de la méthode avant que |
| Apparatus 1/2/5/6 (appareil 1/2/5/6) mfr | Sélection du type d’appareil approprié. |
| Initial Sample (échantillon initial) mfr | Prise d’un échantillon avant de déposer une dose |
| RPM (tours/min) mfr | Vitesse actuelle de l’arbre. arbre. |
| Duration (durée) mfr | Durée pendant laquelle la rotation finale sera e |
| Import mfr | Permet à l’utilisateur d’importer une méthode à |
| Cancel (annuler) mfr | Permet de quitter les écrans de saisie de la mét |
| oC mfr | Température actuelle du bain. bain. |
| Next Sample (échantillon suivant) mfr | Temps avant la prochaine collecte d’échantillon. |
| Sample Time (visualisation des points d’acquisition) mfr | Visualisation des points d’acquisition dans la m |
| Needle Position (Position d’aiguille) mfr | Flèche vers le haut haut |
| Tray Position (Position de plateau) mfr | Eject Eject |
| Pump Control (commande de la pompe) mfr | Volume Volume |
| Maximum acquisition points per method mfr | 36 |
| Closed loop sampling mfr | Yes Yes |
| Path lengths (mm) mfr | 0.2 |
| Automated dosage delivery mfr | Yes Yes |
| Sampling mfr | Simultaneous Simultaneous |
| Sample filtration (min. pore size μm) mfr | 0.2 μm |
| Sample archival mfr | Yes Yes |
| minimum reading interval mfr | 45 seconds |
| lamp lifetime mfr | 10 years |
| photometric linear range mfr | 3.5 Abs |
| Slit Width | 1.5 nm |
| Weight | 27 kg |
| Dimensions (height × width × depth) mfr | 600 mm |
| Enclosure mfr | All materials are recyclable recyclable |
| path length (multicell) mfr | 0.2...10 mm |
| path length (fiber optic tips) mfr | 1...20 mm |
| Mains voltage mfr | 115 |
| Line frequency mfr | 60 Hz |
| ISM classification mfr | 1 A |
| Sampling frequency mfr | 2 minutes |
| Power consumption mfr | 8 A |
| Display mfr | color touchscreen touchscreen |
| Local user interface mfr | 7 inch |
| Interface utilisateur locale mfr | 7 pouces |
| Media types mfr | Compatible with all dissolution media containing |
| Media replacement mfr | 100 |
| Media Change (changement du milieu) mfr | Lorsque cette option est activée, elle permet d’ |
| Media Volume (volume de milieu) mfr | Volume de milieu présent initialement dans les c |
| Operating Altitude | 2000 m |
| Photometric noise mfr | 0.0001 Abs |
| Humidity mfr | 80 % RH |
Description
Construction and series positioning
The 850-DS Sampling Station (model G7930A) is a standalone benchtop unit measuring 400 × 390 × 600 mm and weighing 27 kg, with a color touchscreen and an integrated 8-syringe rotary piston pump. Within the Dissolution Autosamplers series it is the base sampling station; the 850-DS Filtration Module (G7931A) is a separate option that adds automated filtration, increases height to 600 mm and width to 480 mm with the waste tray, and draws 8 A versus 2 A for the base unit. The base station handles sampling, media replacement and cleaning; filtration is enabled only when the G7931A module is installed.
Fluid handling and method behavior
Sample volume per vial or tube is 0.1–14 mL, doubling to 28 mL with dual sample, and sampling accuracy is 10 mL ± 2.5 %. The syringe pump runs at 6–12 mL/min, so timepoints cannot be closer than 2 minutes. Prime and purge volumes are 0–30 mL; cleaning volume is 0.1–15 mL for up to 9 cycles via the rinse port or sampling cannulas. At each timepoint the volume removed from each vessel equals sample volume + waste drop volume + a fixed 0.2 mL needle purge, and media replacement must be programmed to include that purge. Up to 36 timepoints are allowed per method, numbered consecutively, with a maximum run duration of 999 hours; beyond 36 samples the collection tray must be changed manually.
Installation, control and operation
Three 1/4-28 male to 1/8" barbed fittings on the left rear connect waste, media replacement and rinse containers via 1/4" silicone tubing; up to 8 sampling and 8 return lines run to the dissolution apparatus, and unused ports must be fitted with barbed fittings routed to waste. Media containers must sit at or below the 850-DS, and the waste inlet tube must never be below the waste liquid level. The unit accepts 115/230 V~ ±10 %, 60/50 Hz, 2 A through a country-specific Agilent power cord into a grounded outlet; there is no voltage selector or accessible fuse. Methods are stored in 40 memory locations, and the station can control one Agilent/Varian/VanKel dissolution apparatus or run synchronized with a second 850-DS using a 9-pin null modem cable (5075-0449) on the AUX port.
Integration & Operating Notes
The syringe pump flow range of 6–12 mL/min and the minimum 2-minute sampling frequency set the shortest programmable interval. Methods allow up to 36 timepoints, numbered consecutively, and entering zero for the time deletes a timepoint. Runs can last up to 999 hours, but more than 36 samples requires a manual tray change.
Dissolution Apparatus Control must be enabled only when the 850-DS is connected to another Agilent, Varian or VanKel dissolution apparatus. It must not be checked for a third-party apparatus with fixed probes, because the station would then issue control commands that conflict with the apparatus's own controller.
Two-collector synchronized operation requires the same method loaded in the same memory location on both 850-DS units and a 9-pin to 9-pin null modem cable (5075-0449) connected via the AUX port. Time Delayed start cannot be combined with Synchronized Start, so the start option must be chosen accordingly.
A validated cleaning method must be run after each dissolution test, using 0.1–15 mL cleaning volume for up to 9 cycles through the rinse port or the sampling cannulas. The main screen shows a maintenance-required notification and leak alarm; these must be addressed before the next run. Filter plates, where the G7931A module is fitted, are single-use with a maximum of 5 uses and a warning above 2 uses per filter.
Media containers must be at the same level as or below the 850-DS, and the waste inlet tube must never be below the waste liquid level. Otherwise liquid can siphon back into the instrument. Up to 8 sampling and 8 return lines connect to the apparatus, and unused positions must be fitted with barbed fittings routed to waste.
Selection & Integration Q&A
Choose the base G7930A when the method does not require automated filtration before collection; it draws 2 A and stands 400 mm high. Add the G7931A module when filtered samples are needed: it raises height to 600 mm, width to 480 mm with the waste tray, weighs 29 kg and draws 8 A, and must be installed by a certified Agilent representative.
The station connects to the 708-DS, Apparatus 3, Apparatus 7, NanoDis or Multicell UV system, with up to 8 sampling and 8 return lines using 1/4-28 male to 1/8" barbed fittings and 1/4" silicone tubing. Apparatus Control may be enabled only for Agilent/Varian/VanKel apparatus. Collection trays for UV-Vis or HPLC are separate options, and the station can be driven by Dissolution Workstation Software or OpenLab Dissolution Workflow Manager.
A validated cleaning method must be run after each test using 0.1–15 mL for up to 9 cycles. Consumables not included are cannula filters (1, 10, 35, 70 μm) and, with the G7931A module, single-use filter plates (0.22 or 0.45 μm, max 5 uses, warning above 2). A 9-pin null modem cable (5075-0449) is needed for synchronized two-unit operation, and country-specific power cables are ordered separately.