Agilent Technologies AA-Duo

Agilent Technologies AA-Duo — Atomic spectroscopy.

Graphite furnace AAS method card for cadmium in water, using an UltrAA lamp at 4 mA, 228.8 nm, 0.5 nm slit, 10 µL sample and 5 µL modifier. The temperature program runs ash at 568 °C (sample) / 570 °C (standard) and atomization at 1577 °C (sample) / 1533 °C (standard), with 2100 °C as the atomization temperature setpoint. Three replicates yield 101.8 % QC recovery against a 2.50 Cd standard.

Full specifications

Core Performance & Functional Specifications
Wavelength range 228.8 nm
Elements Cd
Fuel Flow 2 L/min
Inert Gas Purity 99.99 %
Modifier Volume 5 μL
Qc Recovery 101.8 %
Sample volume 10 μL
Matrix mfr Water Water
Chemical modifier mfr 4
Instrumentation mfr 240 AAS
Hollow Cathode Lamp mfr Cadmium Cadmium
Graphite tube platform mfr Omega Omega
Mode mfr Emission Emission
Chemical modifier volume mfr 1 g/L
Bandpass mfr 0.2 nm
D2 correction mfr No No
Replicates mfr 3
Lamp type mfr UltrAA lamp lamp
Dimensions, Weight & Mechanical Design
Slit Width 0.5 nm
Electrical, Power & Drive Specifications
Lamp Current 4 mA
Operating Conditions, Environment & Protection
Ash Temperature (sample) 568 °C
Atomization Temperature 2100 °C
Atomize Temperature (sample) 1577 °C
Standards, Safety & General Data
Ash Temperature (standard) 570 °C
Atomize Temperature (standard) 1533 °C
Standard Solution 2.50 Cd
Standard Reference Material mfr 1 Cd

Description

Method architecture

Thermal behavior

Operation

Integration & Operating Notes

Process Limits
Inert gas must meet 99.99 % purity

The method is specified at 99.99 % inert gas purity. Lower-purity gas introduces oxidant and trace-metal background into the graphite furnace, which is not corrected here because D2 correction is switched off.

Factory Specification Sheet
Maintenance
Omega platform tube is a consumable tied to this program

The graphite tube with Omega platform is the atomizer consumable for this method. Tube condition directly affects the 1577/1533 °C atomization signals and the 101.8 % recovery; replacement tubes must be re-conditioned before the QC check is trusted.

Factory Specification Sheet

Selection & Integration Q&A

When should this Cd-in-water graphite furnace method be chosen rather than another element or matrix program?

Pick it for cadmium in water only: 228.8 nm, Cd UltrAA lamp at 4 mA, 0.5 nm slit, 10 µL sample plus 5 µL modifier, 2.50 Cd standard. For other elements or matrices the lamp, wavelength, modifier and temperature stages all change, so this program does not transfer.

What must the gas supply and modifier system provide for this method to run as specified?

Inert gas at 99.99 % purity and a 1 g/L chemical modifier dosed at 5 µL per 10 µL sample. D2 correction is off, so background control relies on the modifier and gas quality rather than on continuum correction.

What duty-cycle and environment limits apply to running this program?

The verified facts give no duty-cycle, ambient or installation limits, and no manual digest was supplied. Only the analytical parameters (temperatures, gas purity, volumes, lamp settings) are confirmed; environmental limits must be taken from the instrument manual, not this card.

What consumables and accessories are needed beyond the base instrument for this method?

A Cd UltrAA lamp, graphite tubes with Omega platform, 1 g/L chemical modifier, 2.50 Cd standard and 99.99 % inert gas. None of these are stated as part of base delivery; the tube in particular is a wear item that must be stocked and re-conditioned.