Agilent Technologies Cary 620-FTIR
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Cary 620-FTIR

Agilent Technologies Cary 620-FTIR — Molecular spectroscopy with Flow rate 10 liter/minute.

The Cary 620-FTIR is an FTIR microscope that couples to Cary 600 Series spectrometers, using a 64×64 FPA (5.5 µm pixels) and a 15× IR objective for chemical imaging. It operates on 100–240 VAC, 50–60 Hz, 60 W, with a 350 µm field of view and 4 cm⁻¹ spectral resolution. It is used for spatially resolved molecular mapping of heterogeneous samples such as tissue sections.

Flow rate
10 liter/minute

Full specifications

Core Performance & Functional Specifications
Flow rate 10 L/min
Temperature 20...26 °C
Wavelength range 4000...900
Dewar Equilibrium Time 10 minutes
Disk Usage 4.3 GB
Field of view 350 µm
Integration Time 12 µs
Magnification 15 x
Rotational speed 5 kHz
Spectral resolution 4 dimensionless
Tissue section thickness mfr 7 μm
Animal age mfr 13 old
Olefinic functional group (C=C-H) absorption band mfr 3012
Lipid carbonyl (C=O) absorption band mfr 1735
Number of scans mfr 256 scan
FPA size mfr 64 pixels
Pixel size mfr 5.5 μm
Co-added scans (sample) mfr 16
Co-added scans (background) mfr 32
Under-sampling (UDR) scan setting mfr 4
Low pass optical filter cut-off mfr 3975
Attenuation setting mfr Open Open
Mosaic configuration mfr 27
Total mosaic size mfr 8.4 mm
Purge source pressure mfr 60 psi
Rear clearance mfr 30 centimeters
Overhead clearance mfr 75 cm
Loaded flatness tolerance mfr 0.4 length
Laser wavelength mfr 632.8 nanometers
Gas outlet distance from instrument mfr 1.5 meters
Cryosection thickness mfr 7 μm
Number of scans for co-addition mfr 256 scans
Spectral cutoff (Fourier transform limit) mfr 3975
Low pass optical filter cutoff mfr 3975
Spectral cutoff mfr 1000
Maximum purge source pressure mfr 60 psi
Minimum rear clearance mfr 30 cm
Scans mfr 256
Spatial resolution mfr 5.5 μm
Mosaic mfr 7 tiles
cryosection thickness mfr 7 μm
number of scans mfr 256 scans
pixel resolution mfr 5.5 micron
detector array size mfr 64 pixel
olefinic band mfr 3012
carbonyl band mfr 1735
Single FPA tile FOV (μm) mfr 1230
Nominal pixel size (μm) mfr 19.2
Mosaic config. mfr 7
Total mosaic size (mm) mfr 8.6
Collection time (hh:mm) mfr 1
sample co-added scans mfr 16
background co-added scans mfr 32
UDR scan setting mfr 4
low-pass filter cutoff mfr 3975
4x objective spectral cutoff mfr 1000
tissue section size mfr 8 mm
total mosaic size 4x mfr 8.6 mm
data size 4x mfr 4.3 GB
data size 15x mfr 42.4 GB
collection time 4x mfr 1
collection time 15x mfr 10
UDR 8 spectral limit mfr 1975
region of interest size mfr 750 μm
General mfr 8
Verifying Safe State mfr 8
Ultraviolet radiation mfr 9
Compressed gas cylinders mfr 9
Electrical hazards mfr 10
Laser safety mfr 11
Other precautions mfr 13
Warnings or symbols mfr 14
Color coding mfr 15
Electromagnetic compatibility mfr 15
Agilent Cary 600 Series FTIR spectrometer mfr 4 equivalent
Agilent Cary 610/620 FTIR microscope (typical) mfr 1 equivalent
Maximum source pressure mfr 60 psi
Dewar initial LN2 fill mfr 475 milliliters
Dewar top-up LN2 mfr 200 milliliters
Dimensions, Weight & Mechanical Design
Depth 68.9 cm
Height 66.8 cm
Weight 50 kg
Width 39.6 cm
Electrical, Power & Drive Specifications
Power consumption 60 W
Microscope voltage mfr 100...240 max
Mains frequency mfr 50...60 Hz
Power circuit rating mfr 10 A
Fuse type (Cary 600 Series FTIR spectrometer) mfr 4 equivalent
Fuse type (Cary 610/620 FTIR microscope, typical) mfr 1 equivalent
Power circuit mfr 10 A
Materials, Media & Process Compatibility
Liquid Nitrogen Temperature 196 °C
Liquid nitrogen initial fill mfr 475 milliliters
Liquid nitrogen top-off mfr 200 milliliters
Liquid nitrogen initial fill volume mfr 475 mL
Liquid nitrogen top-off volume mfr 200 mL
Operating Conditions, Environment & Protection
Operating temperature 20 °C
Purge Gas Dew Point -70 °C
Storage temperature -70 °C
Relative humidity mfr 20 condensing
Cryogenic cooling mfr 9
Standards, Safety & General Data
CE compliance mfr 15

Description

Construction and positioning

Behavior under load and in different modes

Installation, connection and control

Integration & Operating Notes

Mounting
Bench must be flat to 0.4 mm per 300 mm and vibration-free

A warped or sagging support surface misaligns the optical path and degrades imaging; verify loaded flatness before installation. The 50 kg microscope also requires a bench rated for the load and a fork lift or suitable lifting device for moving.

User Manual, Installation requirements / p.23 and Table 2 / p.22
Electrical
Dedicated 10 A circuit required; shared circuits add noise

The auto-ranging 100–240 VAC supply draws 60 W max but must sit on a clean, dedicated grounded branch with no other instruments. Shared circuits risk electrical noise and interference with sensitive IR measurements. Use only the specified time-lag fuses (T2.0 AH for the microscope).

User Manual, Specifications / p.20–21
Process Limits
Purge gas must be dry to −70 °C dew point at ≤10 L/min

Use water-free air dried to −70 °C dew point or dry nitrogen. Source pressure must not exceed 60 psi (420 kPa) unless an additional regulator is fitted. Wet or over-pressured purge compromises the optics. Gas drops should be within 1.5 m of the instrument with stop valves and gauges.

User Manual, Specifications / p.21 and Safety / p.10
Cooling & Media
Never top the Dewar to the brim; o-ring freeze causes vacuum loss

Fill about 475 mL initially, wait 10–20 min for thermal equilibrium, then add about 200 mL. Overfilling freezes the elastomer o-ring seal, causing loss of vacuum and requiring return of the detector assembly for pump-down. Use the half-funnel incremental fill method.

User Manual, Operation / p.39–40
Control & I/O
Confirm FPA type before specifying the control PC serial ports

Type A FPA detectors use two cables and require two PC serial ports (FPA data and motorized stage), with an interface converter on Serial Port 1 and an RS232 pigtail to the FPA/TRS penthouse connector. Type B uses three cables and only one serial port for the stage; its trigger cable runs from a two-pin snap-in at the detector to a 26-pin D-connector at the spectrometer FPA/TRS port.

User Manual, FPA detector cable installation / p.31–34
Environment
Room must hold 20–26 °C with ≤1 °C/hr gradient and 20–50% RH

Laboratory HVAC must hold a tight temperature band; rapid room-temperature swings degrade FTIR microscope performance. Humidity must be non-condensing. The instrument is Class A ISM equipment for professional/non-domestic environments and may cause EMC disturbances in residential settings.

User Manual, Specifications / p.20 and EMC / p.16
Maintenance
Cleaning limited to water or mild detergent; no solvents

Wipe spills immediately with a soft cloth dampened with water or mild detergent. Organic solvents and abrasives must not be used as they can damage exterior surfaces and optics. Routine maintenance procedures are in Resolutions Pro Help; non-listed maintenance must be performed by Agilent-trained representatives.

User Manual, Cleaning / p.46 and Maintenance section
Compliance
ITAR-regulated; U.S. export license required for export

The product is ITAR-regulated (22 CFR 120-130). A U.S. export license is required to export it from the United States. Export, shipment, service and use are subject to ITAR restrictions; verify licensing before international transfer.

User Manual, ITAR statement / p.19

Selection & Integration Q&A

When should I choose the Cary 620-FTIR microscope over other configurations in the Cary 610/620 series?

The Cary 620 is specified for coupling to Cary 600 Series spectrometers and supports FPA imaging with 64×64 and 128×128 configurations. The manual's fuse table distinguishes Cary 610/620 microscope fuses from the spectrometer fuse. Choose the 620 when FPA-based chemical imaging with mosaic capability (e.g., 7-tile, 8.6 mm) is required; the 610 is referenced only in the fuse table, so confirm its imaging capability with Agilent.

What are the requirements for the purge gas supply and the spectrometer interface?

Purge must be water-free air at −70 °C dew point or dry nitrogen, max 10 L/min, source pressure ≤60 psi unless an additional regulator is fitted. The microscope connects to a Cary 600 Series spectrometer; FPA cabling depends on detector type: Type A needs two PC serial ports, Type B needs one serial port for the stage plus a trigger cable to the spectrometer FPA/TRS port.

What mounting and environmental conditions does the manual require?

Bench must be level, vibration-free, rigid, and flat to 0.4 mm per 300 mm loaded. Allow 30 cm rear clearance and 75 cm overhead. Operating environment: 20–26 °C, ≤1 °C/hr gradient, 20–50% RH non-condensing, sea-level operation. Power on a dedicated 10 A circuit, 100–240 VAC, 50–60 Hz.

What maintenance and consumables are required, and what is not included in base delivery?

Liquid nitrogen is a consumable: ~475 mL initial fill plus ~200 mL top-off per cooldown. Routine maintenance procedures are in Resolutions Pro Help; non-listed service must be by Agilent-trained representatives. Cleaning only with water or mild detergent. Spare fuses (T2.0 AH microscope, T4 AH spectrometer) and purge gas regulators/flow meters are not included and must be sourced separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Optimizing Tissue Preparation and Methodology
Brochure ·786,5 КБ ·4 pp.
View PDF
PDF
FTIR Microscopic Imaging of Large Samples
Brochure ·1,8 МБ ·6 pp.
View PDF
PDF
Agilent Cary 620 FTIR Microscope User's Guide
Manual ·50 pp.
View PDF