Agilent Technologies Fluorescence Spectrometer, Fluorometer, Cary Eclipse
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Fluorescence Spectrometer, Fluorometer, Cary Eclipse

Agilent Technologies Fluorescence Spectrometer, Fluorometer, Cary Eclipse — Molecular spectroscopy.

The Cary Eclipse is a benchtop fluorescence spectrophotometer using a xenon flash lamp and a red-sensitive photomultiplier extended to 900 nm. It supports fluorescence, phosphorescence, chemiluminescence and bioluminescence modes, with kinetics data collection at 80 points/s and scan speeds to 24000 nm/min. It is used for trace molecular analysis in laboratory spectroscopy workflows.

Full specifications

Core Performance & Functional Specifications
Kinetics Data Collection Rate 80 points/s
Minimum Reduced-volume Cuvette Volume 5 µL
Red-sensitive Photomultiplier Sensitivity Extension 900 nm
Scan speed 24000 nm/min
Typical Lamp Life 10 years
Wavelength range scan time mfr 3 s
Xenon flash lamp lifetime mfr 3 flashes
Dimensions, Weight & Mechanical Design
Bench footprint mfr 600 mm

Description

Construction and positioning

Measurement behavior

Installation and operation

Integration & Operating Notes

Process Limits
Red-sensitive PMT extends detection to 900 nm only

The photomultiplier sensitivity extension is specified to 900 nm; measurements planned beyond this wavelength are outside the documented detector range and require a different instrument configuration. Confirm the emission wavelength of the analyte before committing the method.

Factory Specification Sheet
Maintenance
Xenon flash lamp rated for 10 years typical life

The typical lamp life is stated as 10 years, which differs from the extra-spec entry listing '3 flashes' for xenon flash lamp lifetime. Because these two figures conflict, neither should be used for a replacement-interval plan without clarification from the manufacturer.

Factory Specification Sheet; 5990 7788EN Cary Eclipse Brochure
Process Limits
Reduced-volume cuvettes must hold at least 5 µL

The minimum reduced-volume cuvette volume is 5 µL. Samples prepared below this volume cannot be reliably measured in the reduced-volume configuration, so dilution or a larger cuvette must be planned when sample quantity is limited.

Factory Specification Sheet
Control & I/O
Kinetics acquisition capped at 80 points per second

Time-resolved kinetics data collection is limited to 80 points/s. For fast reactions, the sampling interval implied by this rate must be checked against the reaction half-life; slower chemistry is unaffected. No controller or interface details are given in the supplied digests.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the Cary Eclipse rather than another fluorescence system in this line?

No sibling models are documented in the supplied facts, so the Cary Eclipse is the only listed configuration. Choose it when you need fluorescence, phosphorescence, chemiluminescence or bioluminescence modes with detection to 900 nm and 5 µL reduced-volume capability.

What adjacent system parts or accessories must be confirmed before integrating this instrument?

The supplied brochure digest contains no actionable facts on cabling, controllers, power supply or cuvette holders. These must be confirmed from the full instrument manual, which was not provided in digest form; do not assume included accessories from the base specification.

What maintenance schedule and consumables apply, and what is not included in base delivery?

Lamp life is stated inconsistently (10 years typical vs '3 flashes'), so no replacement interval can be set from these facts. Consumables and accessories included in base delivery are not documented in the supplied digests and must be confirmed with the manufacturer.

Dimensional Drawings & Technical Documentation

Technical documentation