Agilent Technologies Femto Pulse Capillary Array
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Femto Pulse Capillary Array

Agilent Technologies Femto Pulse Capillary Array — Automated electrophoresis.

The Femto Pulse System with 12-capillary array is a pulsed-field automated electrophoresis instrument built on the Fragment Analyzer platform. It separates DNA fragments up to 165,000 bp in 1.5 hours—10× faster than PFGE—and detects down to 50 fg/µL with ultra-sensitivity kits. It is used for sizing and quantifying high-molecular-weight DNA, BACs, and RNA in genomics workflows.

Full specifications

Core Performance & Functional Specifications
Maximum Separable Dna Fragment Size 165000 bp
Maximum Separable Genomic Dna Size 165000 bp
Minimum Dna Fragment Input Concentration (ultra Sensitivity Kits) 50 fg/µL
Sample Volume Per Well 2 µL
Separation Time For 165,000 Bp Fragments 1.5 hours
Speed advantage vs pulsed-field gel electrophoresis (PFGE) mfr 10 faster
Sizing range mfr 1300 bp
Input concentration range (fragments) mfr 0.3 pg/μL
Input concentration range (smears) mfr 5 pg/μL
Kit size mfr 275 samples
Input concentration range (total RNA) mfr 15 pg/μL
Input concentration range (mRNA) mfr 25 pg/μL
Input concentration range (fragment up to 48 kb) mfr 3 pg/μL
Input concentration range (fragment <2 kb) mfr 1 pg/μL
Input concentration range (fragment <48 kb) mfr 1.6 pg/μL
Input concentration range (fragment >48 kb) mfr 3 pg/μL
Input concentration range (multiple frags) mfr 12.5 pg/μL
Speed advantage versus PFGE mfr 10 faster
Sample dilution mfr single dilution dilution
speed advantage vs PFGE mfr 10 faster
maximum queued samples mfr 288 samples
number of gel types mfr two gel types types
Standards, Safety & General Data
Maximum Extended Warranty 5 years
Standard Warranty 1 year
Maximum extended warranty time mfr five years total total
standard warranty period mfr one-year year

Description

Construction and positioning

Performance envelope

Kits, software and operation

Integration & Operating Notes

Process Limits
Input concentration must match the kit and sample type

Each kit and sample class has a minimum input concentration: 50 fg/µL for ultra-sensitivity NGS kits, 0.1 pg/µL for fragments, 25 pg/µL for smears and mRNA, 15 pg/µL for total RNA, and 1–12.5 pg/µL for various fragment size classes. Loading below these limits risks no or non-quantitative peaks; overloading can compress the sizing range. Quantify samples before dilution to stay within the kit-specific window.

Data Sheet 5994-2452EN, Femto Pulse consumables
Maintenance
12-capillary array is a consumable replaced per kit workflow

The Femto Pulse 12-Capillary Array is listed as a separate consumable alongside the analysis kits, so it must be ordered and tracked independently of the instrument. Plan replacement according to sample load and kit instructions; the array is not part of the base instrument delivery and should be included in the consumables budget from the first run.

Data Sheet 5994-2452EN, Femto Pulse consumables
Control & I/O
ProSize software handles sizing and quantification output

Data analysis is performed in ProSize software, which produces sizing and concentration results from the capillary electropherograms. Integration with laboratory information systems is not described in the supplied documents; confirm export formats and any LIMS connectivity requirements with the vendor before specifying the instrument in a regulated workflow.

Brochure 5994-0410EN, Femto Pulse System overview

Selection & Integration Q&A

When should I choose the Femto Pulse System instead of a 4150 or 4200 TapeStation system?

Choose the Femto Pulse when you need sizing beyond the TapeStation range—up to 165,000 bp—or ultra-high sensitivity down to 50 fg/µL with ultra-sensitivity kits. TapeStation systems are better suited to routine lower-molecular-weight DNA/RNA QC at higher throughput and simpler operation.

What consumables and accessories must be ordered with the instrument to run samples?

The base instrument does not include the 12-capillary array or analysis kits. Order the Femto Pulse 12-Capillary Array plus the appropriate quantitative kit (gDNA 165 kb, Ultra Sensitivity NGS, Ultra Sensitivity RNA) or qualitative kit (55 kb BAC, 165 kb BAC); all kits are 275-sample sizes.

What sample input concentration is required for different sample types?

Minimum inputs vary by kit and sample: 50 fg/µL for ultra-sensitivity NGS kits, 0.1 pg/µL for fragments, 25 pg/µL for smears and mRNA, 15 pg/µL for total RNA, and 1–12.5 pg/µL for fragment size classes. Quantify and dilute samples to the kit-specific window before loading 2 µL per well.

What warranty and service coverage is available for the Femto Pulse System?

The instrument ships with a one-year standard warranty, which can be extended to a total of five years through CrossLab Silver service. Plan the service contract at purchase to cover the capillary array and kit workflow over the instrument's life.