Agilent Technologies Ethylene testing GC/MS application, olefin impurity testing
Agilent Technologies Ethylene testing GC/MS application, olefin impurity testing — Gas chromatography mass spectrometry gc ms.
The Agilent Arsine/Phosphine GC/MS Analyzer is a factory-configured system built on the 7890B GC and 5977B MSD with HES, dedicated to trace arsine, phosphine, H2S and COS in ethylene and propylene. It is distinguished within the GC/MS Application Solutions series by its olefin-specific column and hydrogen-into-source flow, delivering sub-ppb IDLs for these hydride gases.
Full specifications
| Carrier gas | 1.2 mL/min |
| Analyzer model mfr | 3440 B |
| Gas chromatograph mfr | 7890 GC |
| Column mfr | 120 Olefins |
| Hydrogen flow into source mfr | 150 μL/min |
| Linearity (R2), Phosphine (PH3) mfr | 0.9996 |
| Linearity (R2), Hydrogen sulfide (H2S) mfr | 0.9995 |
| Linearity (R2), Arsine (AsH3) mfr | 0.9999 |
| Linearity (R2), Carbonyl sulfide (COS) mfr | 0.9987 |
| RSD, Phosphine (PH3) mfr | 5.14 |
| RSD, Hydrogen sulfide (H2S) mfr | 3.96 |
| RSD, Arsine (AsH3) mfr | 0.62 |
| RSD, Carbonyl sulfide (COS) mfr | 5.06 |
| IDL, Phosphine (PH3) mfr | 0.715 ppb |
| IDL, Hydrogen sulfide (H2S) mfr | 0.456 ppb |
| IDL, Arsine (AsH3) mfr | 0.063 ppb |
| IDL, Carbonyl sulfide (COS) mfr | 0.575 ppb |
| Mass spectrometer detector mfr | 5977 HES |
| Oven program mfr | 35 °C |
Description
System architecture and series positioning
- This is a complete, application-specific analyzer (model G3440B with option SP1 7890-0667), not a general-purpose GC/MS. It comprises an Agilent 7890B gas chromatograph and a 5977B Series GC/MSD equipped with a High-Efficiency Source (HES) and hydrogen cleaning.
- Within the GC/MS Application Solutions series, this modification is dedicated to olefin impurity testing; sibling configurations target pesticide analysis (GC/MS/MS Pesticides Analyzer, MRM transitions database for pesticide analysis by GCMS). Selection should be driven by the target analyte list and matrix, not by hardware alone.
- The system is pre-configured for arsine and phosphine, with additional validated performance for hydrogen sulfide and carbonyl sulfide in ethylene and propylene streams.
Analytical behavior and performance
- The method uses a 120 m olefin-specific column with a 1.2 mL/min carrier gas flow and an oven program starting at 35 °C; hydrogen flow into the source is 150 µL/min.
- Linearity (R²) is 0.9996 for PH3, 0.9995 for H2S, 0.9999 for AsH3 and 0.9987 for COS across the calibrated range.
- Repeatability (RSD) is 5.14% for PH3, 3.96% for H2S, 0.62% for AsH3 and 5.06% for COS.
- Instrument detection limits are 0.715 ppb for PH3, 0.456 ppb for H2S, 0.063 ppb for AsH3 and 0.575 ppb for COS, supporting trace-level monitoring in olefin product streams.
Installation, control and operation
- The analyzer is delivered as an integrated GC/MSD system; installation and operation follow the dedicated application note (5991-6967EN) covering the G3440B with option SP1 7890-0667 configuration.
- Hydrogen is used both as carrier/source gas and for source cleaning; the 150 µL/min hydrogen flow into the source is a fixed method parameter that must be maintained for consistent sensitivity.
- The 120 m column and 35 °C initial oven temperature require adequate laboratory temperature control and carrier-gas supply stability to preserve retention-time reproducibility.
Integration & Operating Notes
The documented application covers arsine, phosphine, H2S and COS in ethylene and propylene. Using other matrices or target analytes requires re-validation of linearity, RSD and IDL, because the column and oven program are optimized for this olefin separation.
The 5977B HES with hydrogen cleaning depends on a stable hydrogen flow into the source. Deviating from the specified 150 µL/min changes ionization efficiency and can degrade IDLs and repeatability; the hydrogen supply must be regulated and monitored accordingly.
The HES with hydrogen cleaning is intended to reduce source contamination, but sustained olefin service still requires periodic source cleaning and column performance checks. Schedule these based on RSD and IDL trending rather than fixed intervals, since no interval is specified in the supplied documents.
Selection & Integration Q&A
Choose this G3440B/option SP1 7890-0667 system when the target analytes are arsine, phosphine, H2S and COS in ethylene or propylene at ppb levels. The pesticide siblings use different columns, MS/MS transitions and methods and are not configured or validated for these hydride gases.
The system requires a regulated hydrogen supply capable of 150 µL/min into the MS source, carrier gas at 1.2 mL/min, and the 120 m olefin column specified in the method. The 7890B GC and 5977B MSD with HES are supplied as the integrated analyzer; lab gases and consumables are not part of the base delivery.
Plan for periodic MS source cleaning and column performance checks, plus replacement of carrier-gas filters and hydrogen supply consumables. No maintenance interval is stated in the supplied documents; trend RSD and IDL values to schedule service. Columns, gases and source consumables are not included in the base analyzer.