Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Maintenance free, JetClean ion source for GC/MS instrument

Agilent Technologies Maintenance free, JetClean ion source for GC/MS instrument — Gas chromatography mass spectrometry gc ms.

The Agilent JetClean is a self-cleaning ion source accessory for Agilent single- and triple-quadrupole GC/MS systems, operating in Acquire & Clean or Clean Only mode. It extends manual source cleaning from every two weeks to once in nine months by bleeding hydrogen through the source at 0.13–3.52 mL/min (Clean Only) or 0.13–0.49 mL/min (Acquire & Clean).

Agilent Technologies GC MS Ion Sources Inlets Series — Model Comparison & Selection

Model Detection limitClean duration lower setpoint, minClean duration upper setpoint, minEmission current lower setpoint, microampereEmission current upper setpoint, microampereHydrogen flow acquire clean mode lower setpoint, mL/min
Maintenance free, JetClean ion source for GC/MS instrument Current —112010350.13
Ion Sources for Gas Chromatography/Mass Spectrometry attogram112010350.13

Full specifications

Core Performance & Functional Specifications
Clean Duration, Lower Setpoint 1 min
Clean Duration, Upper Setpoint 120 min
Hydrogen Flow (acquire & Clean Mode), Lower Setpoint 0.13 mL/min
Hydrogen Flow (acquire & Clean Mode), Upper Setpoint 0.49 mL/min
Hydrogen Flow (clean Only Mode), Lower Setpoint 0.13 mL/min
Hydrogen Flow (clean Only Mode), Upper Setpoint 3.52 mL/min
Cleaning gas mfr hydrogen hydrogen
Manual source cleaning interval without JetClean mfr once every two weeks weeks
Manual source cleaning interval with JetClean mfr once in nine months months
Area %RSD, Benzo[a]anthracene mfr 1.8 %RSD
Area %RSD, Chrysene mfr 1.2 %RSD
Area %RSD, Benzo[b]fluoranthene mfr 1.9 %RSD
Area %RSD, Benzo[a]pyrene mfr 1.7 %RSD
Calibration linearity R2, Benzo[a]pyrene mfr 0.9976
MRM ratio, 252.1 -> 250.1 / 252.1 -> 252.1 mfr 56.2
MRM ratio, 250.1 -> 248.1 / 252.1 -> 252.1 mfr 8.7
Hydrogen flow step size mfr 0.069 sccm
Electrical, Power & Drive Specifications
Emission Current, Lower Setpoint 10 µA
Emission Current, Upper Setpoint 35 µA
Operating Conditions, Environment & Protection
Source Temperature, Lower Setpoint 150 °C
Source Temperature, Upper Setpoint 350 °C

Description

Construction and series positioning

Operating modes and behavior

Operation and maintenance effect

Integration & Operating Notes

Process Limits
Hydrogen flow ceiling depends on the active mode

Acquire & Clean mode is limited to 0.13–0.49 mL/min, while Clean Only permits up to 3.52 mL/min. Setting a Clean Only-level flow while acquiring data exceeds the Acquire & Clean envelope; verify the mode before entering a flow setpoint.

Operating Manual G7077-90052, parameter setpoints
Maintenance
Manual source cleaning still required about every nine months

JetClean extends the manual cleaning interval from roughly every two weeks to once in nine months; it does not eliminate manual cleaning. Plan the nine-month service event into the instrument schedule.

JetClean eBook 5991-9517EN
Mounting
Retrofit availability is system-specific

The JetClean is offered as an option on new 5977A/B, 7000B/C/D, 7010A/B and Intuvo-equipped systems, and as a retrofit to existing systems. Confirm the host MS model and Intuvo configuration before specifying a retrofit.

JetClean eBook 5991-9517EN, scope and variants

Selection & Integration Q&A

When should I specify the JetClean ion source instead of a standard source on a 5977 or 7000-series GC/MS?

Choose JetClean when sample load would otherwise force manual source cleaning about every two weeks; it extends the interval to once in nine months. If source cleanliness is not the bottleneck, the standard source on the 5977C, 7000E or 7010D needs no hydrogen plumbing or cleaning-gas supply.

What does the JetClean require from the surrounding system in terms of gas supply and host instrument?

It needs a hydrogen supply connected through the dedicated plumbing described in the operating manual, and a compatible host: 5977A/B, 7000B/C/D, 7010A/B, or Intuvo-equipped 5977B/7000D. Hydrogen flow is set per mode, 0.13–0.49 mL/min in Acquire & Clean and up to 3.52 mL/min in Clean Only.

What maintenance and consumables should I plan for beyond the base JetClean delivery?

Plan a manual source cleaning about once every nine months even with JetClean, versus roughly every two weeks without it. The hydrogen supply and its plumbing are system-side items to be provided; the operating manual covers configuration, setpoints and troubleshooting.