GC nitrogen chemiluminescence detector, NOx analyzer drawing
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies GC nitrogen chemiluminescence detector, NOx analyzer

Agilent Technologies GC nitrogen chemiluminescence detector, NOx analyzer — Gas chromatography.

The Agilent 8255 nitrogen chemiluminescence detector (NCD) is a GC detector configured here as a dedicated NOx analyzer for trace ammonia in hydrocarbon and hydrogen matrices. Its 900 °C burner, 200 °C base, and 131 Torr burner pressure define the chemiluminescence reaction zone, while a 1 mL gas sample loop and 5 Hz data rate support ppb-level quantitation from 37.1 to 495 ppb. It is positioned in the GC detector series as the nitrogen-selective alternative to the FID, NPD, FPD, and sulfur chemiluminescence detectors.

Full specifications

Core Performance & Functional Specifications
Data Rate 5 Hz
Sample Loop Volume 1 mL
Oxidizer flow during analysis mfr 30 mL/min
Oxidizer flow at ammonia elution mfr 4 mL/min
Valve switch-off time after injection mfr 0.25 minutes
Ammonia run area mfr 264.08
Column mfr Select Low Ammonia Ammonia
Burner pressure mfr 131 Torr
Calibration level L1 concentration mfr 37.1 ppb
Calibration level L2 concentration mfr 49.5 ppb
Calibration level L3 concentration mfr 79.2 ppb
Calibration level L4 concentration mfr 99 ppb
Calibration level L5 concentration mfr 198 ppb
Calibration level L6 concentration mfr 495 ppb
Response factor L1 mfr 0.228
Response factor L2 mfr 0.236
Response factor L3 mfr 0.218
Response factor L4 mfr 0.215
Response factor L5 mfr 0.175
Response factor L6 mfr 0.156
Linearity range mfr 37.1...495 ppb
Quality control requirement mfr 100 ppb
Operating Conditions, Environment & Protection
Burner Temperature 900 °C
Standards, Safety & General Data
Ammonia standard deviation mfr 7.20
Ammonia relative standard deviation (RSD) mfr 2.73

Description

Construction and series positioning

Behavior under load and in different modes

Installation, control and operation

Integration & Operating Notes

Control & I/O
Valve must switch off 0.25 min after injection

The method requires the sample valve to close 0.25 minutes after injection, with oxidizer flow reduced from 30 mL/min to 4 mL/min at ammonia elution. The GC timetable must sequence valve and oxidizer flow together; a missed event will distort the ammonia peak and the run area.

Application note 5991-8061EN
Process Limits
Multi-level calibration required over 37.1–495 ppb

Response factors fall from 0.228 at 37.1 ppb to 0.156 at 495 ppb, so a single-point calibration will bias results across the range. The method defines six calibration levels and a 100 ppb quality-control requirement; bracketing QC checks must be run within this range.

Application note 5991-8061EN
Maintenance
Use Select Low Ammonia column and deactivated stainless surfaces

The method specifies the CP8590 Select Low Ammonia megabore column and UltiMetal Plus deactivated stainless steel surfaces to prevent ammonia adsorption. Substituting an untreated column or unpassivated tubing will cause low recovery and tailing, so these consumables must be maintained as specified.

Application note 5991-8061EN

Selection & Integration Q&A

When should I choose this NCD configuration instead of an FID, NPD, FPD, or sulfur chemiluminescence detector in the same GC detector series?

Choose the NCD for nitrogen-selective trace analysis such as ammonia at ppb levels (37.1–495 ppb calibrated here). The FID is universal and not nitrogen-selective; the NPD is nitrogen/phosphorus-selective but not a chemiluminescence detector; the FPD targets sulfur and phosphorus; the SCD is sulfur-only. The NCD is the dedicated nitrogen chemiluminescence option.

What adjacent system parts and consumables are required to run this NCD ammonia method?

The referenced configuration uses a 6-port gas sample valve with a 1 mL sample loop, a CP8590 Select Low Ammonia megabore column, UltiMetal Plus deactivated stainless steel surfaces, and OpenLAB CDS. The second method uses a 2 mL loop on an 8890 GC, so loop volume and GC platform are method variables to confirm before ordering.

What maintenance and calibration items are not part of the base detector delivery?

The base detector does not include the CP8590 Select Low Ammonia column, UltiMetal Plus deactivated tubing, the 6-port gas sample valve with 1 mL or 2 mL loop, calibration standards for the six levels (37.1–495 ppb), or the OpenLAB CDS data system. These must be specified separately, and the multi-level calibration and 100 ppb QC check must be scheduled as ongoing service tasks.