Agilent Technologies Automated Sampling with NovoSampler Pro
Agilent Technologies Automated Sampling with NovoSampler Pro — Research flow cytometry.
The NovoSampler Pro is an automated sample sampler for the NovoCyte flow cytometer, supporting a 24-tube rack (12 × 75 mm) and 96-well plates. It delivers 10 µL sampling volumes at 120 µL/min with 1 mix cycle (10 s) and 1 rinse cycle. It is the automation front-end for walk-away acquisition in flow cytometry workflows.
Full specifications
| Mix Cycles | 1 cycles |
| Mixing Duration | 10 s |
| Rinse Cycles | 1 cycles |
| Sample Flow Rate | 66 µL/min |
| Sampling Volume | 10 µL |
| Supported tube rack mfr | 24 tubes |
| Supported sample plate type mfr | 96 plates |
| Minimum sample volume, 96-well V-bottom mfr | 90 μL |
| Dead volume, 96-well V-bottom mfr | 50 μL |
| Minimum sample volume, 96-well U-bottom mfr | 90 μL |
| Dead volume, 96-well U-bottom mfr | 50 μL |
| Minimum sample volume, 48-well plates mfr | 210 μL |
| Dead volume, 48-well plates mfr | 170 μL |
| Minimum sample volume, 24-well plates mfr | 310 μL |
| Dead volume, 24-well plates mfr | 270 μL |
Description
Construction and positioning
- The NovoSampler Pro is a dedicated automation accessory for the NovoCyte flow cytometer, not a standalone instrument. It is positioned within the Flow Cytometry Automation series as the sampler that adds tube-rack and microplate handling to the cytometer.
- It supports a 24-position tube rack for 12 × 75 mm tubes and 96-well plates; other plate formats (48- and 24-well) are accommodated with higher minimum volumes.
- Unlike the flow cytometer models in the sibling list, this unit does not perform optical measurement; it handles sample presentation and uptake.
Behavior under load and media
- Sample flow rate is 120 µL/min, with a sampling volume of 10 µL per acquisition. Mixing is performed with 1 cycle of 10 s; rinse is 1 cycle.
- Minimum sample volume and dead volume depend on the plate type: for 96-well V- and U-bottom plates, 90 µL minimum and 50 µL dead volume; for 48-well plates, 210 µL minimum and 170 µL dead; for 24-well plates, 310 µL minimum and 270 µL dead. These values must be respected to avoid aspirating air.
- Acceleration is specified as 0 s, indicating immediate transition to the set flow rate.
Installation, connection and control
- The operator's guide covers hardware overview, installation, software control, sample acquisition, cleaning and troubleshooting. Control is via the NovoCyte software; the sampler is integrated into the acquisition workflow.
- Only the NovoSampler 24-tube rack (12 × 75 mm tubes) is supported; other tube racks are not compatible.
- Cleaning procedures are described in the guide and are required for maintenance.
Integration & Operating Notes
The sampler requires at least 90 µL in 96-well V- or U-bottom plates, 210 µL in 48-well plates, and 310 µL in 24-well plates. Below these volumes, the dead volume (50, 170, or 270 µL respectively) leaves insufficient liquid for reliable aspiration, risking air intake and lost sample.
The system supports only the NovoSampler 24-tube rack designed for 12 × 75 mm tubes. Other tube racks or tube sizes are not compatible and may cause positioning errors or failed aspiration.
Sample acquisition, mixing, and rinsing are initiated and sequenced via the NovoCyte software. The sampler does not have independent front-panel control; integration with the cytometer software is required for all run operations.
The guide includes cleaning procedures to be followed between runs or at specified intervals. Neglecting cleaning can lead to carryover, clogging, or inaccurate sampling volumes.
Selection & Integration Q&A
Choose the NovoSampler Pro when you need automated sample handling for the NovoCyte cytometer, especially for tube racks or 96-well plates. The Advanteon is a cytometer itself; the sampler is an accessory that adds automation but requires a cytometer for detection.
The sampler must be connected to and controlled by the NovoCyte software. It uses the cytometer's fluidics and optics for acquisition. Ensure the cytometer is compatible and that the sampler is installed per the operator's guide, including proper fluidic connections.
The operator's guide does not specify environmental limits beyond standard laboratory conditions. Duty cycle is continuous within the software-controlled acquisition; however, sample volumes must meet plate-specific minimums to avoid air aspiration. Cleaning between runs is required.
Regular cleaning per the operator's guide is required. The 24-tube rack is a specific accessory; other racks are not supported. Sample plates and tubes are consumables not included. No additional cables or accessories are listed in the digest; refer to the guide for installation parts.