Agilent Technologies Automated Sampling with NovoSampler Pro
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Automated Sampling with NovoSampler Pro

Agilent Technologies Automated Sampling with NovoSampler Pro — Research flow cytometry.

The NovoSampler Pro is an automated sample sampler for the NovoCyte flow cytometer, supporting a 24-tube rack (12 × 75 mm) and 96-well plates. It delivers 10 µL sampling volumes at 120 µL/min with 1 mix cycle (10 s) and 1 rinse cycle. It is the automation front-end for walk-away acquisition in flow cytometry workflows.

Full specifications

Core Performance & Functional Specifications
Mix Cycles 1 cycles
Mixing Duration 10 s
Rinse Cycles 1 cycles
Sample Flow Rate 66 µL/min
Sampling Volume 10 µL
Supported tube rack mfr 24 tubes
Supported sample plate type mfr 96 plates
Minimum sample volume, 96-well V-bottom mfr 90 μL
Dead volume, 96-well V-bottom mfr 50 μL
Minimum sample volume, 96-well U-bottom mfr 90 μL
Dead volume, 96-well U-bottom mfr 50 μL
Minimum sample volume, 48-well plates mfr 210 μL
Dead volume, 48-well plates mfr 170 μL
Minimum sample volume, 24-well plates mfr 310 μL
Dead volume, 24-well plates mfr 270 μL

Description

Construction and positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Process Limits
Respect plate-specific minimum volumes to avoid air aspiration

The sampler requires at least 90 µL in 96-well V- or U-bottom plates, 210 µL in 48-well plates, and 310 µL in 24-well plates. Below these volumes, the dead volume (50, 170, or 270 µL respectively) leaves insufficient liquid for reliable aspiration, risking air intake and lost sample.

Factory Specification Sheet; NovoSampler Pro Operator's Guide
Mounting
Use only the NovoSampler 24-tube rack for tube sampling

The system supports only the NovoSampler 24-tube rack designed for 12 × 75 mm tubes. Other tube racks or tube sizes are not compatible and may cause positioning errors or failed aspiration.

NovoSampler Pro Operator's Guide, Supported tube rack
Control & I/O
Sampler operation is controlled through the NovoCyte software

Sample acquisition, mixing, and rinsing are initiated and sequenced via the NovoCyte software. The sampler does not have independent front-panel control; integration with the cytometer software is required for all run operations.

NovoSampler Pro Operator's Guide, Software control
Maintenance
Perform cleaning cycles as described in the operator's guide

The guide includes cleaning procedures to be followed between runs or at specified intervals. Neglecting cleaning can lead to carryover, clogging, or inaccurate sampling volumes.

NovoSampler Pro Operator's Guide, Cleaning

Selection & Integration Q&A

When should I choose the NovoSampler Pro instead of a standalone flow cytometer like the NovoCyte Advanteon?

Choose the NovoSampler Pro when you need automated sample handling for the NovoCyte cytometer, especially for tube racks or 96-well plates. The Advanteon is a cytometer itself; the sampler is an accessory that adds automation but requires a cytometer for detection.

What are the requirements for integrating the NovoSampler Pro with a NovoCyte flow cytometer?

The sampler must be connected to and controlled by the NovoCyte software. It uses the cytometer's fluidics and optics for acquisition. Ensure the cytometer is compatible and that the sampler is installed per the operator's guide, including proper fluidic connections.

What are the environmental or duty-cycle limitations for the NovoSampler Pro?

The operator's guide does not specify environmental limits beyond standard laboratory conditions. Duty cycle is continuous within the software-controlled acquisition; however, sample volumes must meet plate-specific minimums to avoid air aspiration. Cleaning between runs is required.

What maintenance is required and what accessories are needed but not included?

Regular cleaning per the operator's guide is required. The 24-tube rack is a specific accessory; other racks are not supported. Sample plates and tubes are consumables not included. No additional cables or accessories are listed in the digest; refer to the guide for installation parts.