Agilent Technologies 1260 Infinity II Prime LC System
Agilent Technologies 1260 Infinity II Prime LC System — Liquid chromatography with Flow rate 0.005 liter/minute, Pumping speed 3×10⁻⁴ m³/h.
The 1260 Infinity II Prime LC System is a modular HPLC/UHPLC platform built around the Flexible Pump G7104C, delivering up to 5 mL/min at pressures up to 800 bar. It distinguishes itself within the InfinityLab series by covering both HPLC and UHPLC methods on one system, while the 1290 Bio variant extends pressure to 1300 bar for biocompatible UHPLC. It serves as the front-end separation engine in analytical workflows, feeding detectors such as the 1260 Infinity II DAD WR (190–950 nm, 120 Hz).
Full specifications
| Flow rate | 5 mL/min |
| Pumping speed | 5 mL/min |
| Wavelength range | 190...950 nm |
| Maximum Data Acquisition Rate | 120 Hz |
| Maximum Individual Bottle Volume | 2 L |
| Maximum pressure | 800 bar |
| Pressure limit | 800 bar |
| Solvent Cabinet Maximum Volume | 8 L |
| Stroke | Automatic |
| Nominal inline filter pore size mfr | 0.3 μL |
| Compressibility mfr | Use solvent types types |
| Composition mfr | 95 %A |
| Stoptime mfr | as pump pump |
| Timetable mfr | 2.5 %B |
| Posttime mfr | 1 min |
| Autobalance mfr | Prerun Prerun |
| G7120-60007 (Bottle Head Assembly, long (1730 mm)) mfr | Solvent Bottle Bottle |
| G5664-68712 (Analytical tubing kit 0.25 mm i.d. PTFE-ESD) mfr | Detector Detector |
Description
Construction and series positioning
- The system is assembled from stackable modules: pump, autosampler, column thermostat, detector, and solvent cabinet, interconnected by specified capillaries to minimize delay volume.
- The 1260 Infinity II Prime LC System uses the Flexible Pump G7104C (5 mL/min, 800 bar) and is positioned below the 1290 Infinity III Bio Online LC System, which offers up to 1300 bar with biocompatible flow path for UHPLC and bio-applications.
- Recommended stack configurations include the InfinityLab Flex Bench (preferred), single stack (max 4 modules), benchtop, or two stacks; the Flex Bench optimizes flow path and reduces delay volume.
Behavior under load and media
- The pump supports flow rates up to 5 mL/min and pressures up to 800 bar, enabling both standard HPLC and UHPLC methods on the same instrument.
- The solvent cabinet holds a maximum of 8 L total, with individual bottles limited to 2 L; this constrains mobile-phase volume and requires planning for longer runs.
- The leak plane with PVDF leak sensor stops the system upon liquid detection; DMF is incompatible and causes corrosion, so mobile phases must be checked for compatibility.
- Optional inline filter (5067-5407, nominal 0.3 µm pore size) protects downstream components from particulates and precipitates; it requires weekly back-flushing.
Installation, connection, and control
- Network configuration is set via rear DIP switches: all DOWN = default IP 192.168.254.11; switch 4 UP = DHCP; switch 5 UP = STORED address. LAN configuration must be performed from the module with direct LAN connection, typically the detector with highest data rate.
- The Online LC Monitoring Software requires OpenLab CDS v2.x Workstation Plus; the ERI interface timeout (min) defines how long the software waits for a sample delivery device response before marking the sample as missed.
- Daily pump conditioning with the application solvents (e.g., 50/50 A/B at 0.5 mL/min) minimizes pressure ripple by dissolving air bubbles in pump heads.
- For 1260 Infinity III systems, use capillary 5500-1246 (0.17 mm x 500 mm) from pump to sampler and 5500-1191 (Quick Turn 0.12 mm x 280 mm) from column/MCT valve to detector to ensure correct flow path and minimal delay volume.
Integration & Operating Notes
The 1260 Infinity II Prime LC System is rated for a maximum pressure of 800 bar. Operating above this limit risks damage to the pump, column, and flow path components. For methods requiring pressures up to 1300 bar, the 1290 Infinity III Bio Online LC System must be used instead.
Dimethylformamide (DMF) is incompatible with the PVDF leak sensor and causes corrosion, which can lead to false leak detection or sensor failure. Mobile phases containing DMF must be avoided; if used, the leak sensor must be inspected regularly for corrosion.
The rear DIP switches determine the network address mode: all switches DOWN = default IP 192.168.254.11; switch 4 UP = DHCP; switch 5 UP = STORED address. These must be set before attempting LAN configuration, which should be executed from the module with direct LAN connection to the controller software, usually the detector with the highest data rate.
The optional inline filter (part number 5067-5407, nominal pore size 0.3 µm) protects the system from blockages caused by buffers or solvent precipitation. It must be back-flushed weekly to maintain performance and prevent pressure buildup.
Water-based solvents must be exchanged daily and selected to be used up within 1–2 days to prevent algae growth and contamination. Use only HPLC-grade solvents and water filtered through 0.2 µm filters.
Buffer solutions should be placed on lower channels (A and/or D) of the 4CGV gradient valve. All 4CGV channels must be flushed regularly with water to remove salt deposits and prevent precipitation and blockage.
The ERI interface timeout (in minutes) defines how long the software waits for an answer from the sample delivery device. If no answer is received, the sample is marked as missed and the software continues. Set the timeout according to the hardware response time to avoid missed samples.
The recommended waste container is 5043-1221 (6 L waste can with 1 Stay Safe cap GL45 with 4 ports). Waste tubes must not immerse in the liquid to prevent back-siphonage. Ensure leak plane outlets are unobstructed and waste tubes are correctly routed to allow leak detection.
Selection & Integration Q&A
Choose the 1260 Infinity II Prime LC System for HPLC and UHPLC methods requiring pressures up to 800 bar and flow rates up to 5 mL/min. Select the 1290 Infinity III Bio Online LC System when pressures up to 1300 bar or biocompatible flow paths are needed, such as for bio-molecule analysis or high-pressure UHPLC.
For 1260 Infinity III systems, use capillary 5500-1246 (0.17 mm x 500 mm SI/SI) from pump to sampler and 5500-1191 (Quick Turn 0.12 mm x 280 mm) from column/MCT valve to detector. These ensure correct flow path connections and minimal delay volume. For 1290 Bio systems, MP35N capillaries are required.
The leak plane with PVDF leak sensor stops the system when liquid is detected. Ensure leak plane outlets are unobstructed and waste tubes are correctly routed. Do not use DMF as it corrodes the sensor. Use waste container 5043-1221 (6 L) and do not immerse waste tubes in liquid to prevent back-siphonage.
The optional inline filter (5067-5407, 0.3 µm) must be back-flushed weekly. Water-based solvents must be exchanged daily and used within 1–2 days; use HPLC-grade solvents and water filtered through 0.2 µm filters. The inline filter is not included in base delivery and must be ordered separately.