Agilent Technologies 5975C GC MSD
Agilent Technologies 5975C GC MSD — Gas chromatography.
The Agilent 5975C GC MSD is a stand-alone capillary GC detector with an independently heated EI ion source, hyperbolic quadrupole, and performance turbomolecular pump. It accepts column flows up to 4.0 mL/min and columns up to 0.53 mm ID, distinguishing it from standard-turbo and diffusion-pump siblings. It serves as the mass-selective detector in a GC/MS system for EI and CI analyses.
Full specifications
| Foreline Reading | 34.7 Torr |
| Gas Flow | 2 mL/min |
| Maximum Recommended Gas Flow | 1.5 mL/min |
| Micro-ion Gauge Reading At 0.5 Ml/min Column Flow | 3.18 Torr |
| Operating system | Displays the GC/MSD ChemStation operating system version. |
| Optimal He Column Flow | 1 mL/min |
| Max column id mfr | 0.53 mm |
| CI capability mfr | Yes Yes |
| DIP capability (3rd party) mfr | Yes Yes |
| High vacuum pump mfr | Performance turbo turbo |
| Run Method mfr | Displays the current method name and starts an a |
| Run Sequence mfr | Displays the current sequence and starts a seque |
| Release ChemStation mfr | Disassociates the GC/MSD ChemStation from the MS |
| Prepare to vent mfr | Reminds you to shut down the GC then prepares th |
| Pumpdown mfr | Initiates a pumpdown sequence. sequence. |
| Methane/PCI mfr | 17 |
| Methane/NCI mfr | 17 |
| Isobutane/PCI mfr | 39 |
| Isobutane/NCI mfr | 17 |
| Ammonia/PCI mfr | 18 |
| Ammonia/NCI mfr | 17 |
| Control firmware mfr | Displays the MSD firmware version. version. |
| Front panel mfr | Displays the version of the LCP. LCP. |
| Log amplifier mfr | Displays version information. information. |
| Sideboard mfr | Displays the sideboard type. type. |
| Mainboard mfr | Displays the mainboard type. type. |
| CI bakeout time mfr | 2 hours |
| Methane reagent gas flow setting mfr | 20 |
| Depth | 65 cm |
| Height | 42 cm |
| Weight | 25 kg |
| Width | 26 cm |
| Foreline (roughing) pump weight mfr | 11 kg |
| Serial number mfr | Is assigned to the MSD by GC/MSD ChemStation Con |
Description
Construction and series positioning
- The 5975C GC MSD integrates a performance turbomolecular pump, a rotary vane foreline pump, an independently heated electron impact (EI) ion source, a hyperbolic quadrupole mass filter, a high-energy dynode electron multiplier, and a GC/MSD interface.
- Within the 5975 Series, this performance-turbo configuration supports the highest column flows and largest column IDs: up to 4.0 mL/min and 0.53 mm ID, versus 2.0 mL/min and 0.32 mm for standard-turbo models and 1.5 mL/min and 0.25 mm for diffusion-pump models.
- Chemical ionization (CI) capability is provided; a third-party direct insertion probe (DIP) is also supported.
Behavior under column flow
- Foreline pressure rises with column flow: 34.7 Torr at 0.5 mL/min, 39.4 Torr at 0.7 mL/min, 52.86 Torr at 1 mL/min, 60.866 Torr at 1.2 mL/min, 91.784 Torr at 2 mL/min, and 125.76 Torr at 3 mL/min.
- Micro-Ion gauge readings are 3.18 Torr at 0.5 mL/min, 2.48 Torr at 4 mL/min, and 3.75 Torr at 6 mL/min.
- Optimal helium column flow is 1 mL/min; the maximum recommended gas flow for this model is 4.0 mL/min.
Installation and operation
- The MSD is a stand-alone detector for Agilent Series GCs, with a local control panel for status messages, menus, pump down, temperature and flow control, venting, tuning, and verification.
- Column installation requires the capillary to extend 4–6 mm past the ferrule in the split/splitless inlet; the interface nut is tightened an additional 1/4 to 1/2 turn with a 1/4-inch open-end wrench after hand-tightening.
- During pump down, the turbo must reach 80% speed within 10–15 minutes and eventually 95%; otherwise the electronics shut off the foreline pump and a power cycle is required. After “Okay to run,” wait 2 hours for thermal equilibrium before acquiring data.
Integration & Operating Notes
The performance turbo allows up to 4.0 mL/min, but exceeding this degrades spectral performance and sensitivity. Column selection must respect this limit; standard-turbo and diffusion-pump siblings are limited to 2.0 and 1.5 mL/min respectively.
Larger-ID columns require higher flows that exceed the vacuum system capacity. Use 0.53 mm ID or smaller; diffusion-pump and standard-turbo models are restricted to 0.25 mm and 0.32 mm respectively.
Lower-purity carrier gas introduces column bleed and contamination, reducing sensitivity and increasing maintenance. Use ultrapure helium or other inert gas; never condition columns with hydrogen.
Higher ammonia pressure can cause condensation from gas to liquid, disrupting CI operation. When running ammonia for 5 or more hours per day, ballast the foreline pump for at least 1 hour daily and purge with methane afterward.
If the turbo does not reach 80% speed in this window, the MSD electronics shut off the foreline pump and a power cycle is required. Verify foreline pump operation and vacuum seal before restarting.
Heating a column without carrier gas flow will damage the column. Ensure carrier gas is flowing before turning on any GC heated zone.
The MSD must reach thermal equilibrium; data acquired earlier may not be reproducible. Plan analytical runs accordingly after pump down or maintenance.
Plan long-term maintenance and budget for pump replacement. This applies to models equipped with a dry foreline pump.
Tilting or inverting the instrument can damage internal components. Use appropriate lifting equipment and secure the unit during transport.
Selection & Integration Q&A
Choose this performance-turbo model when using column flows up to 4.0 mL/min or column IDs up to 0.53 mm, or when CI capability is required. Standard-turbo models limit flow to 2.0 mL/min and 0.32 mm ID; diffusion-pump models limit to 1.5 mL/min and 0.25 mm ID and lack CI capability.
Install the capillary column in the split/splitless inlet so it extends 4–6 mm past the ferrule. Tighten the interface nut an additional 1/4 to 1/2 turn with a 1/4-inch open-end wrench after hand-tightening, and check tightness after one or two heat cycles. Use carrier gas at least 99.9995% pure.
The MSD must remain upright at all times when moving or storing. During pump down, the turbo must reach 80% speed within 10–15 minutes and eventually 95%; otherwise the electronics shut off the foreline pump and a power cycle is required. After “Okay to run,” wait 2 hours for thermal equilibrium before acquiring data.
Replace foreline pump oil every 6 months (every 3 months for CI with ammonia). Replace dry foreline pump typically every 3 years. When using ammonia for 5+ hours/day, ballast the foreline pump for at least 1 hour daily and purge with methane afterward. Column conditioning requires ultrapure inert gas, not hydrogen.