Agilent Technologies 5977B GC MSD
Agilent Technologies 5977B GC MSD — Gas chromatography.
The 5977B GC MSD is a single-quadrupole mass selective detector for 8890, 9000, 8860 and 7890 GCs, with a 41 × 30 × 54 cm housing and 39 kg mainframe. This turbo-pump configuration supports 1–2 mL/min helium column flow, up to 0.53 mm columns and field CI upgrade. It serves as the mass-selective detector in capillary GC/MS systems.
Full specifications
| Foreline Reading | 34.7 mTorr |
| Gas Flow | 2 mL/min |
| Ion Vacuum Gauge Reading | 3.18 Torr |
| Maximum Column Id | 0.53 mm |
| Maximum Recommended Gas Flow | 1.5 mL/min |
| Optimal He Column Flow | 1 mL/min |
| Max column id mfr | 0.25 mm |
| CI capability mfr | No No |
| Inert ion sources available mfr | Yes Yes |
| GC compatibility mfr | 9000 Series |
| Foreline pumps available mfr | Pfeiffer Duo Duo |
| DIP capability (3rd party) mfr | Yes Yes |
| High vacuum pump mfr | Diffusion Diffusion |
| CI capability‡ mfr | No No |
| DIP** capability (3rd party) mfr | Yes Yes |
| Methane/PCI mfr | 17 |
| Methane/NCI mfr | 17 |
| Isobutane/PCI mfr | 39 |
| Isobutane/NCI mfr | 17 |
| Ammonia/PCI mfr | 18 |
| Ammonia/NCI mfr | 17 |
| Minimum column velocity during conditioning mfr | 30 cm/s |
| Carrier gas purity for conditioning mfr | 99.9995 better |
| Column conditioning hold time mfr | 30 minutes |
| Column extension past ferrule mfr | 4...6 mm |
| Column extension for XTR/SS/Inert/CI mfr | 1 mm |
| Final nut tightening mfr | 4...1 turn |
| Minimum pumpdown time before verification mfr | 60 minutes |
| Injection volume for sensitivity check mfr | 1 μL |
| Thermal equilibrium wait after pumpdown mfr | 2 hours |
| Bakeout time before CI tuning mfr | 2 hours |
| Foreline Pump Weight | 11 kg |
| Weight | 39 kg |
| Housing height mfr | 41 cm |
| Housing width mfr | 30 cm |
| Housing depth mfr | 54 cm |
| MSD housing height mfr | 41 cm |
| MSD housing width mfr | 30 cm |
| MSD housing depth mfr | 54 cm |
| Diffusion pump mainframe weight mfr | 39 kg |
| EI/CI turbo pump mainframe weight mfr | 46 kg |
| Primary fuse rating mfr | 12.5 V |
| EMV threshold for CI autotune concern mfr | 2600 V |
| Sound Pressure Lp | 70 dB |
| Standard EI turbo pump mainframe weight mfr | 44 kg |
Description
Construction and series positioning
- The 5977B is a benchtop single-quadrupole MSD; the mainframe measures approximately 41 cm high × 30 cm wide × 54 cm deep and weighs 39 kg, with a separate foreline pump adding 11 kg.
- The series spans diffusion-pump and turbo-pump mainframes; this turbo configuration is distinguished by higher allowable gas flow (1–2 mL/min optimal helium, 4 mL/min maximum recommended total flow, 6.5 mL/min absolute maximum) and larger column capacity (up to 0.53 mm ID) versus diffusion-pump limits of 1 mL/min optimal, 1.5 mL/min recommended and 0.25 mm ID.
- CI capability is not available on diffusion-pump models; turbo-pump models are field upgradeable to CI, so this modification is the entry point for later PCI/NCI work.
Behavior under load and media
- Typical ion vacuum gauge readings at 1 mL/min helium are 6.26E-06 Torr (turbo) and 3.66E-05 Torr (diffusion); foreline reading for diffusion is 52.86 mTorr. Pressures may vary by as much as 30% between instruments; consistently higher pressures indicate air leaks or vacuum problems.
- Exceeding the maximum recommended total gas flow (4 mL/min turbo) degrades performance; operation above the 6.5 mL/min absolute maximum causes expected degradation of spectral performance and sensitivity.
- For CI operation with ammonia, supply pressure must remain below 5 psig to prevent condensation; a 5% ammonia/95% helium or 5% ammonia/95% methane mixture is frequently used to minimize negative effects.
Installation, control and operation
- The MSD installs on 8890, 9000 (Intuvo), 8860 and 7890 Series GCs; column installation uses either a self-tightening column nut or a standard column nut, with the GC/MSD interface tip seal fitted per the GC-specific procedure.
- Operation is from the data system or the 7890 GC control panel, with WUI network configuration, eModule mini display and front-panel status LED; pumpdown, temperature and flow control, venting, vacuum monitoring, tuning and JetClean setup are covered in EI mode, while CI mode adds CI autotune, reagent gas flow control and PCI/NCI autotune.
- Primary fuses are T12.5A, 250 V (part 2110-1398), two required; the most likely cause of primary fuse failure is a foreline pump problem.
Integration & Operating Notes
Total flow includes column flow plus reagent gas flow. The maximum recommended total gas flow into the MSD is 4 mL/min for turbo-pump models; operation above the 6.5 mL/min absolute maximum causes expected degradation of spectral performance and sensitivity. Size column and reagent gas flows together so the sum stays within the recommended limit.
Higher pressures can cause ammonia to condense to liquid. Keep the tank upright below the flow module and coil tubing into vertical loops. A 5% ammonia/95% helium or 5% ammonia/95% methane mixture is frequently used to minimize negative effects of ammonia while achieving good CI.
Primary fuses are T12.5A, 250 V (part 2110-1398), two required. Use exact replacement fuses; the most likely cause of primary fuse failure is a foreline pump problem, so investigate the pump before replacing fuses.
Selection & Integration Q&A
Choose the turbo configuration when you need 1–2 mL/min optimal helium flow, up to 4 mL/min recommended total flow, columns up to 0.53 mm ID, or field CI upgradeability. Diffusion-pump models are limited to 1 mL/min optimal, 1.5 mL/min recommended, 0.25 mm ID and no CI capability.
The GC must deliver 1–2 mL/min optimal helium column flow and keep total gas flow (column plus reagent gas) at or below 4 mL/min recommended, 6.5 mL/min absolute maximum. Columns up to 0.53 mm ID (30 m) are supported; install via self-tightening or standard column nut with the GC/MSD interface tip seal.
Check foreline pump fluid weekly and replace every 6 months (every 3 months with ammonia CI); diffusion pump fluid and dry foreline pump seals annually; IDP-3 tip seal annually; leaks monthly; calibration vials every 6 months. Primary fuses are T12.5A, 250 V (part 2110-1398), two required; Apiezon L high vacuum grease is needed for the analyzer side plate O-ring.