Agilent Technologies 6495D LC MS
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 6495D LC MS

Agilent Technologies 6495D LC MS — Liquid chromatography.

The Agilent 6495D is a triple quadrupole LC/MS detector for the Ultivo/6475A/6495D series, distinguished by 6495D-only Configure Tune, Advanced Tune, and Diagnostic Tune. It operates in 15–35 °C labs and serves as the MS detector in LC/MS workflows with ESI, AJS ESI, APCI, MMI, or Nanospray sources.

Full specifications

Core Performance & Functional Specifications
Maximum operating pressure 250 PSIG
Software revision mfr 12.1 higher
Steam cleaning LC pump flow mfr 0.5 ml/min
Steam cleaning nebulizer pressure mfr 60 psi
Steam cleaning drying gas flow mfr 13 L/min
Nebulizer flush flow rate mfr 1 minute
Nebulizer flush duration mfr 3...5 minutes
MS40+ oil mist filter distance from cap threads mfr 2 cm
Diagnostic tune duration mfr 3 hours
Maximum manual instrument modes mfr ten ten
ESI mfr ESI-L Low Concentration Tuning Mix Mix
AJS ESI mfr ESI-L Low Concentration Tuning Mix Mix
MMI mfr MMI-L Low Concentration Tuning Mix Mix
APCI mfr APCI-L Low Concentration Tuning Mix Mix
Microflow Nebulizer Assembly mfr 0.003 inch
Tested ESI Nebulizer Assembly (older style) mfr 0.003 inch
Maximum flow rate mfr 8 minute
Gas source supply pressure mfr 10...15 psi
Electrical, Power & Drive Specifications
Line voltage fuse rating mfr 8 amp
Operating Conditions, Environment & Protection
Operating temperature 15...35 °C
Storage temperature -40...70 °C

Description

Series positioning and 6495D-specific architecture

Behavior under tuning and acquisition modes

Installation, control and operation

Integration & Operating Notes

Control & I/O
Manual Tune privilege must be enabled before advanced tuning

Configure Tune and Advanced Tune are inaccessible unless the Manual Tune privilege is enabled in OpenLab Control Panel; set user permissions before attempting 6495D tuning workflows to avoid blocked access.

User Manual, Configure Tune / p.36, Advanced Tune / p.45
Maintenance
Cool the source 20 minutes before daily spray chamber cleaning

Lower drying gas temperature to minimum and, for APCI/multimode, vaporizer heater to minimum, then wait 20 minutes to prevent burns from the high-temperature spray chamber before cleaning the spray shield.

User Manual, Maintenance / p.77
Maintenance
Replace MS40+ pump oil every six months with AVF 60 Gold

Oil should be replaced every six months or sooner if dark/cloudy, filled to the maximum sight-glass level; the oil mist filter must sit no more than 2 cm from outside the cap threads.

User Manual, Maintenance / p.86-90
Process Limits
Use source-specific tuning mix; mixes are not interchangeable

ESI/AJS ESI requires ESI-L (G1969-85000), MMI requires MMI-L (G1969-85020), APCI requires APCI-L (G1969-85010); the wrong mix can cause miscalibration and mass assignment errors.

User Manual, Tuning / p.31, p.79
Maintenance
Refill calibrant bottle before the intake tube draws air

Refill when tuning mix is within a few millimeters of the intake tube end; rinse once with water then twice with ACN, and label the bottle with the tuning mix expiration date to prevent calibrant delivery failure.

User Manual, Maintenance / p.78-79

Selection & Integration Q&A

When should I choose the 6495D over the Ultivo or 6475A in this LC/TQ series?

Pick the 6495D when you need Configure Tune, Advanced Tune, and Diagnostic Tune (6495D-only) and up to ten manual Instrument Modes. Choose Ultivo or 6475A if those advanced tuning workflows are not required; the manual covers all three on shared hardware.

What software and tuning-mix prerequisites must be in place before integrating the 6495D with my LC and workstation?

MassHunter 12.1 or higher is required. Match the tuning mix to the installed source: ESI-L G1969-85000 for ESI/AJS ESI, MMI-L G1969-85020 for MMI, APCI-L G1969-85010 for APCI. Enable Manual Tune privilege in OLCP for advanced tuning.

What environmental and duty-cycle limits apply when installing the 6495D in my lab?

Operate only within 15–35 °C; storage is -40 to 70 °C. Diagnostic Tune occupies about 3 hours and should follow autotune failure. Standby keeps source warm with voltages off; full venting is required before shutdown to protect the turbo pump.

What maintenance schedule and consumables should I budget for beyond base delivery?

Daily/weekly spray chamber cleaning (50:50 LC/MS IPA:water), weekly nebulizer flush (30–70% methanol, ≤1 ml/min, 3–5 min), quarterly rough pump oil check, six-month MS40+ oil/filter change with AVF 60 Gold, plus source-specific tuning mixes and calibrant refills.