Agilent Technologies 715 ICP OES
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 715 ICP OES

Agilent Technologies 715 ICP OES — Atomic spectroscopy.

The Agilent 715 ICP OES is a radial-plasma optical emission spectrometer in the 700 Series, distinguished from the axial-plasma 710/720 by its radial viewing configuration. It operates an argon plasma at approximately 10,000 K, driven by a 40 MHz, 1500 V RMS work coil, for elemental analysis in laboratory process and materials testing. The 715 shares the series' gas, cooling and control architecture but is selected where radial viewing is preferred over axial.

Full specifications

Core Performance & Functional Specifications
Nitrogen (argon Only) 20 ppm
Plasma Temperature 10000 K
Standby Mode Flow 0.75 L/min
Water Vapor 4 ppm
Argon purity mfr 99.996
Nitrogen purity mfr 99.996
Oxygen mfr 5 ppm
Permissible pressure range mfr 400...600 kPa
Recommended pressure mfr 550 kPa
Operational range (plasma on) mfr 9...32.8 L/min
Measuring wavelengths > 200 nm (poly boost off) mfr 13.5...21.75 L/min
Measuring wavelengths < 200 nm (poly boost on) mfr 15.75...24 L/min
Drain vessel minimum capacity mfr 2 liters
Dimensions, Weight & Mechanical Design
Radial induction coil pitch mfr 4.7 mm
Axial induction coil pitch mfr 7.5 mm
Instrument weight mfr 203 kg
Gas supply hose length mfr 1.8 m
Electrical, Power & Drive Specifications
Work Coil Frequency 40 MHz
Fuse 1FS1 and 1FS2 mfr 6.3 mm
Circuit breaker 1CB1 mfr 30 A
Work coil voltage mfr 1500 RMS
Operating Conditions, Environment & Protection
Ambient temperature limit 20 °C
Maximum permissible cooling water pressure mfr 310 kPa
Minimum cooling water flow mfr 1.1 L/min

Description

Construction and series positioning

The 715 is one of four models in the Agilent 700 Series ICP-OES platform. Within the series, the 710 and 720 use an axial plasma configuration, while the 715 and 725 are radial-plasma instruments; the 715 is therefore the radial-viewing modification. The instrument weighs approximately 203 kg and requires mechanical lifting equipment for relocation. The RF generator feeds a work coil operating at 1500 V RMS and approximately 40 MHz, sustaining a plasma at about 10,000 K. Radial induction coil pitch is specified at 4.7 ±0.5 mm center-to-center; coil service is recommended only by an Agilent field service engineer, since a deformed coil distributes RF energy unevenly and can melt the torch or torch clamp.

Gas and cooling behavior

Argon is the primary plasma gas; argon or nitrogen may serve as polychromator purge gas. Gas purity must be 99.996% for both argon and nitrogen, with oxygen below 5 ppm, water vapor below 4 ppm, and nitrogen in argon below 20 ppm. Supply pressure must remain within 400–600 kPa, with 550 kPa recommended. Standby argon flow is 0.75 L/min, while operational flow spans 9–32.8 L/min; with the optional nitrogen purge kit fitted, operational argon flow is 8.3–29.1 L/min and nitrogen flow is 0.45–2.1 L/min. For measurements above 200 nm with Poly Boost off, argon flow is 13.5–21.75 L/min; below 200 nm with Poly Boost on, 15.75–24 L/min. Cooling water flow must stay above 1.1 L/min, below which a flow sensor stops the plasma and camera Peltier assembly; supply pressure must not exceed 310 kPa without regulation.

Installation, control and operation

Installation must be performed by an Agilent-trained or authorized field service engineer and includes ICP Expert II software, the USB-GPIB-HS converter, dark current scan, wavelength calibration and torch scan. The USB-GPIB-HS converter is the only supported instrument-to-software interface for ICP Expert II on Windows 10. Both the rear RF supply circuit breaker and the front-left instrument power switch must be on for operation. The exhaust fan must be running before plasma ignition, and the plasma interlock extinguishes the plasma if mains power fails or the torch compartment door is opened. The torch and torch compartment remain hot for up to five minutes after plasma extinction.

Integration & Operating Notes

Process Limits
Gas purity below 99.996% compromises plasma and purge stability

Argon and nitrogen must meet 99.996% purity with oxygen below 5 ppm, water vapor below 4 ppm, and nitrogen in argon below 20 ppm. Lower-purity gases can destabilize the plasma and degrade optical purge performance, so verify cylinder certificates and regulator/line cleanliness before connection.

User Manual, Gas Supplies, p.20
Cooling & Media
Cooling water flow below 1.1 L/min trips plasma and Peltier

A water flow sensor stops the plasma and the camera Peltier assembly if flow drops below 1.1 L/min (0.3 gpm). The cooling system must be running before plasma ignition, and supply pressure must not exceed 310 kPa (45 psi) without a pressure regulator.

User Manual, Water Chiller System, p.21
Environment
Laboratory temperature must hold 20–25 °C within ±2 °C daily

Optimum analytical performance requires ambient temperature held constant to within ±2 °C throughout the working day, in the 20–25 °C range. HVAC design must account for the instrument's heat load and for exhaust extraction, which must vent to outside air, never within the building.

User Manual, Temperature Control, p.18; Heat, Vapors and Fumes, p.9
Electrical
Two power points must both be on: rear breaker and front switch

The RF supply circuit breaker at the right rear and the instrument power switch at the front left must both be on for the instrument to work. Operator-replaceable fuses 1FS1 and 1FS2 are T6.3 AH 250 V, 5 x 20 mm; circuit breaker 1CB1 is 30 A fast trip. Power boards and extension cables are not recommended.

User Manual, Electrical Power Supplies and Fuses, p.19; Operation, p.28
Maintenance
Drain vessel must be inert, non-glass, ≥2 L and checked hourly

Provide a chemically inert drain vessel, not glass or narrow-necked, holding at least 2 liters, located underneath the sample compartment or on the right side of the instrument. Check and empty it hourly during operation to prevent overflow of sample waste.

User Manual, Drain Vessel, p.22

Selection & Integration Q&A

When should I choose the 715 rather than the 710 or 720 in the 700 Series?

The 715 is the radial-plasma instrument; the 710 and 720 are axial-plasma models. Choose the 715 when radial viewing suits your matrices and detection-limit requirements. The 725 is also radial but is a separate modification; compare its configuration against the 715 before finalizing.

What gas supply and purity infrastructure must be in place before the 715 is installed?

Argon (and nitrogen if the purge kit is fitted) at 99.996% purity, oxygen <5 ppm, water vapor <4 ppm, nitrogen in argon <20 ppm. Regulated pressure 400–600 kPa, 550 kPa recommended. PTFE hoses are 1.8 m with Swagelok fittings; size supply for 9–32.8 L/min operational argon flow, not the 0.75 L/min standby flow.

What are the laboratory environment and duty-cycle constraints for operating the 715?

Ambient temperature 20–25 °C held within ±2 °C over the working day. Exhaust must vent outside and run before every plasma ignition. Cooling water flow above 1.1 L/min and pressure at or below 310 kPa. The torch area stays hot up to five minutes after plasma extinction, so allow cooling before access.

What maintenance and accessories are operator-performable, and what is not included in base delivery?

Operator tasks at hourly/daily/weekly/monthly intervals are listed in the manual; detailed procedures are in ICP Expert II Help. Consumables include torch, cone cleaning, peristaltic pump tubes and fuses. Coil service and non-listed maintenance require Agilent field service. The USB-GPIB-HS converter and ICP Expert II software are installation items; gas regulators, chiller and drain vessel are site-provided.

Dimensional Drawings & Technical Documentation

Technical documentation