Agilent Technologies 7200 GC Q TOF
Agilent Technologies 7200 GC Q TOF — Gas chromatography.
The Agilent 7200 GC Q-TOF (G3851BA) is a quadrupole time-of-flight GC/MS system supplied with both EI and CI removable ion sources. It accepts capillary columns up to 0.32 mm id and 30 m, with total gas flow into the MS limited to 2.0 mL/min recommended (2.4 mL/min maximum). It is used for accurate-mass detection in EI or CI workflows, with source switching under vacuum in under 30 minutes.
Full specifications
| Gas Flow | 2.4 mL/min |
| Maximum Column Id | 0.32 mm |
| Maximum Recommended Gas Flow | 2.0 mL/min |
| Quad Vacuum | 3.41 Torr |
| Rough Vacuum | 1.16 Torr |
| Tof Vacuum | 4.20 Torr |
| Optimal carrier gas flow mfr | 1.0...1.5 mL/min |
| Reagent gas flow mfr | 1...2 mL/min |
| Collision cell gas flow mfr | 1.5 |
| Rough Vac mfr | 1.29 Torr |
| Quad Vac mfr | 7.84 Torr |
| TOF Vac mfr | 2.21 Torr |
| High-vacuum pump 1 mfr | Split flow turbo turbo |
| High-vacuum pump 2 mfr | Split flow turbo turbo |
| High-vacuum pump 3 mfr | Turbo Turbo |
| Optimal carrier gas flow, mL/min mfr | 1.0...1.5 |
| Reagent gas flow, mL/min mfr | 1...2 |
| Flight tube extension over top mfr | 84 cm |
| RIS handle extension from front mfr | 48 cm |
| Carrier gas purity mfr | 99.9995 |
| Minimum column velocity during conditioning mfr | 30 cm/s |
| Foreline pump exhaust hose id mfr | 11 id |
| PCI reagent gas flow mfr | 20 |
| NCI reagent gas flow mfr | 40 |
| EI electron energy mfr | 70 eV |
| Foreline Pump Weight | 21.5 kg |
| Instrument Depth | 89 cm |
| Instrument Height | 48 cm |
| Instrument Width | 71 cm |
| Instrument mainframe weight mfr | 152 kg |
| EI emission current mfr | 35 μA |
| EI repeller voltage mfr | 30 V |
| EI ion focus voltage mfr | 90 V |
| EI entrance lens offset mfr | 25 V |
| Source Temperature | 230 °C |
| PCI mfr | 300 °C |
| NCI mfr | 150 °C |
Description
Construction and series positioning
- The 7200 Q-TOF is a tandem MS architecture: a quadrupole mass filter followed by a time-of-flight analyzer, fed from a GC/MS interface. The model G3851BA ships standard with both an EI source and a CI source using Removable Ion Source (RIS) technology.
- No sibling models are listed for this series; the distinguishing feature of this modification is the dual-source RIS configuration, which allows EI/CI exchange in under 30 minutes while the instrument stays under vacuum.
- Three high-vacuum pumps are fitted: two split-flow turbos and one turbo, backed by a foreline pump.
Behavior under gas load and in different modes
- Optimal carrier gas flow is 1.0–1.5 mL/min. Total flow into the MS above 2.0 mL/min degrades mass spectral and sensitivity performance; above 2.4 mL/min degradation increases further. Columns larger than 0.32 mm id or longer than 30 m are not recommended.
- In EI mode, typical setpoints are MS source 230 °C and MS quad 150 °C. The software enforces hard limits: quadrupole ≤200 °C, ion source ≤350 °C.
- In CI mode, reagent gas flow is 1–2 mL/min and collision cell gas flow is 1.5 mL/min. With ammonia, supply pressure must stay below 5 psig to prevent condensation and flow-module damage.
- Typical vacuum pressures vary by up to 30% between instruments in EI mode; consistent higher readings indicate air leaks or vacuum faults.
Installation, control and operation
- The GC/MS interface typically runs at 280 °C (range 0–400 °C); set it slightly above the maximum GC oven temperature but never above the column maximum. A setpoint below ambient turns the interface heater off.
- The last 4–5 mm of the capillary column must extend past the guide tube into the ionization chamber; insertion depth is critical for sensitivity.
- During pumpdown, turbo speed should reach 80% within 10–15 minutes and eventually 95%; otherwise the electronics shut off the foreline pump and a power cycle is required. After 'Okay to run' appears, wait 2 hours for thermal equilibrium before acquiring data.
- Carrier gas must be at least 99.9995% pure. Hydrogen is specifically prohibited. The top thumbscrews on the analyzer side plates must be finger-tight only; over-tightening causes air leaks. Collision cell top-plate shipping brackets must remain fastened during normal operation.
Integration & Operating Notes
Optimal carrier flow is 1.0–1.5 mL/min; 2.0 mL/min is the maximum recommended total flow and 2.4 mL/min the absolute maximum. Exceeding these degrades mass spectral and sensitivity performance, so column id (≤0.32 mm) and length (≤30 m) must be chosen to stay within this budget.
The manual explicitly prohibits hydrogen with this product. Carrier gas must be helium or another approved gas at ≥99.9995% purity, used with recommended traps to minimize column bleed and ion source contamination.
MassHunter enforces hard temperature limits: quadrupole 200 °C and ion source 350 °C. Setpoints beyond these are rejected, so method development must stay within them; EI starting points are source 230 °C and quad 150 °C.
Higher ammonia supply pressure can cause the gas to condense to liquid and damage the reagent gas flow module. When running ammonia 5 or more hours per day, ballast the foreline pump at least 1 hour per day and purge the MS with methane after ammonia flow.
Over-tightening the top thumbscrew on the front and rear analyzer side plates can cause air leaks that compromise vacuum. The collision cell top-plate shipping brackets must remain fastened during normal operation to secure the plate in the event of an explosion.
Check foreline pump fluid color and level weekly. Replace pump oil and oil mist filter every 6 months or as needed, and replace RIS probe seals yearly or as needed. Skipping these intervals degrades performance and can damage the instrument.
If turbo speed does not reach 80% in 10–15 minutes (and eventually 95%), the MS electronics shut off the foreline pump and a power cycle is required. Monitor pumpdown and troubleshoot vacuum faults before proceeding; after 'Okay to run', wait 2 hours for thermal equilibrium.
Venting while a turbo pump spins at more than 50% of normal operating speed will damage it. Use the automated venting program and never vent by allowing air through the foreline hose.
Selection & Integration Q&A
Pick the G3851BA when methods require both EI and CI: the RIS lets you switch sources in under 30 minutes while staying under vacuum, avoiding full vent cycles. If only EI is ever needed, a single-source setup avoids CI reagent-gas plumbing and the extra ammonia maintenance burden.
Use a capillary column of 0.32 mm id or less and 30 m or less, with carrier gas at ≥99.9995% purity and recommended traps. Keep total flow into the MS at 1.0–1.5 mL/min optimal, 2.0 mL/min maximum recommended, 2.4 mL/min absolute maximum. Hydrogen is prohibited.
Set the GC/MS interface slightly above maximum oven temperature but never above the column maximum (typical 280 °C, range 0–400 °C). Insert the column so 4–5 mm extends past the guide tube. Fasten side-plate thumbscrews finger-tight only and keep collision cell shipping brackets fastened.
Weekly: check foreline pump fluid color and level. Every 6 months: pump oil and oil mist filter. Yearly: RIS probe seals. Ammonia CI adds daily foreline ballasting (≥1 h) and methane purge. Plan for these consumables and the RIS probe extraction tool as part of ownership.