Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 7200 GC Q TOF

Agilent Technologies 7200 GC Q TOF — Gas chromatography.

The Agilent 7200 GC Q-TOF (G3851BA) is a quadrupole time-of-flight GC/MS system supplied with both EI and CI removable ion sources. It accepts capillary columns up to 0.32 mm id and 30 m, with total gas flow into the MS limited to 2.0 mL/min recommended (2.4 mL/min maximum). It is used for accurate-mass detection in EI or CI workflows, with source switching under vacuum in under 30 minutes.

Full specifications

Core Performance & Functional Specifications
Gas Flow 2.4 mL/min
Maximum Column Id 0.32 mm
Maximum Recommended Gas Flow 2.0 mL/min
Quad Vacuum 3.41 Torr
Rough Vacuum 1.16 Torr
Tof Vacuum 4.20 Torr
Optimal carrier gas flow mfr 1.0...1.5 mL/min
Reagent gas flow mfr 1...2 mL/min
Collision cell gas flow mfr 1.5
Rough Vac mfr 1.29 Torr
Quad Vac mfr 7.84 Torr
TOF Vac mfr 2.21 Torr
High-vacuum pump 1 mfr Split flow turbo turbo
High-vacuum pump 2 mfr Split flow turbo turbo
High-vacuum pump 3 mfr Turbo Turbo
Optimal carrier gas flow, mL/min mfr 1.0...1.5
Reagent gas flow, mL/min mfr 1...2
Flight tube extension over top mfr 84 cm
RIS handle extension from front mfr 48 cm
Carrier gas purity mfr 99.9995
Minimum column velocity during conditioning mfr 30 cm/s
Foreline pump exhaust hose id mfr 11 id
PCI reagent gas flow mfr 20
NCI reagent gas flow mfr 40
EI electron energy mfr 70 eV
Dimensions, Weight & Mechanical Design
Foreline Pump Weight 21.5 kg
Instrument Depth 89 cm
Instrument Height 48 cm
Instrument Width 71 cm
Instrument mainframe weight mfr 152 kg
Electrical, Power & Drive Specifications
EI emission current mfr 35 μA
EI repeller voltage mfr 30 V
EI ion focus voltage mfr 90 V
EI entrance lens offset mfr 25 V
Operating Conditions, Environment & Protection
Source Temperature 230 °C
PCI mfr 300 °C
NCI mfr 150 °C

Description

Construction and series positioning

Behavior under gas load and in different modes

Installation, control and operation

Integration & Operating Notes

Process Limits
Keep total gas flow into the MS at or below 2.0 mL/min

Optimal carrier flow is 1.0–1.5 mL/min; 2.0 mL/min is the maximum recommended total flow and 2.4 mL/min the absolute maximum. Exceeding these degrades mass spectral and sensitivity performance, so column id (≤0.32 mm) and length (≤30 m) must be chosen to stay within this budget.

Operation Manual, Table 2, page 24
Process Limits
Hydrogen carrier gas is prohibited on this instrument

The manual explicitly prohibits hydrogen with this product. Carrier gas must be helium or another approved gas at ≥99.9995% purity, used with recommended traps to minimize column bleed and ion source contamination.

Operation Manual, pages 16, 25, 29, 37, 39
Control & I/O
Software rejects quad >200 °C and source >350 °C setpoints

MassHunter enforces hard temperature limits: quadrupole 200 °C and ion source 350 °C. Setpoints beyond these are rejected, so method development must stay within them; EI starting points are source 230 °C and quad 150 °C.

Operation Manual, Table 4, page 43; page 44
Process Limits
Ammonia CI requires supply pressure below 5 psig

Higher ammonia supply pressure can cause the gas to condense to liquid and damage the reagent gas flow module. When running ammonia 5 or more hours per day, ballast the foreline pump at least 1 hour per day and purge the MS with methane after ammonia flow.

Operation Manual, page 64
Mounting
Analyzer side-plate thumbscrews must be finger-tight only

Over-tightening the top thumbscrew on the front and rear analyzer side plates can cause air leaks that compromise vacuum. The collision cell top-plate shipping brackets must remain fastened during normal operation to secure the plate in the event of an explosion.

Operation Manual, page 16
Maintenance
Weekly pump-fluid check; oil and mist filter every 6 months

Check foreline pump fluid color and level weekly. Replace pump oil and oil mist filter every 6 months or as needed, and replace RIS probe seals yearly or as needed. Skipping these intervals degrades performance and can damage the instrument.

Operation Manual, page 94
Maintenance
Pumpdown must reach 80% turbo speed within 10–15 minutes

If turbo speed does not reach 80% in 10–15 minutes (and eventually 95%), the MS electronics shut off the foreline pump and a power cycle is required. Monitor pumpdown and troubleshoot vacuum faults before proceeding; after 'Okay to run', wait 2 hours for thermal equilibrium.

Operation Manual, page 55
Maintenance
Never vent a turbo spinning above 50% of operating speed

Venting while a turbo pump spins at more than 50% of normal operating speed will damage it. Use the automated venting program and never vent by allowing air through the foreline hose.

Operation Manual, page 39

Selection & Integration Q&A

When should I choose this dual-source RIS 7200 Q-TOF rather than a single-source EI configuration?

Pick the G3851BA when methods require both EI and CI: the RIS lets you switch sources in under 30 minutes while staying under vacuum, avoiding full vent cycles. If only EI is ever needed, a single-source setup avoids CI reagent-gas plumbing and the extra ammonia maintenance burden.

What column and carrier-gas setup must accompany this MS for compliant operation?

Use a capillary column of 0.32 mm id or less and 30 m or less, with carrier gas at ≥99.9995% purity and recommended traps. Keep total flow into the MS at 1.0–1.5 mL/min optimal, 2.0 mL/min maximum recommended, 2.4 mL/min absolute maximum. Hydrogen is prohibited.

What installation and environment conditions does the manual impose on the analyzer and interface?

Set the GC/MS interface slightly above maximum oven temperature but never above the column maximum (typical 280 °C, range 0–400 °C). Insert the column so 4–5 mm extends past the guide tube. Fasten side-plate thumbscrews finger-tight only and keep collision cell shipping brackets fastened.

What maintenance schedule and consumables are required beyond base delivery?

Weekly: check foreline pump fluid color and level. Every 6 months: pump oil and oil mist filter. Yearly: RIS probe seals. Ammonia CI adds daily foreline ballasting (≥1 h) and methane purge. Plan for these consumables and the RIS probe extraction tool as part of ownership.