Agilent Technologies Cary 670 FTIR
Agilent Technologies Cary 670 FTIR — Molecular spectroscopy.
The Cary 670 FTIR is a laboratory Fourier-transform infrared spectrometer with an embedded 632.8 nm continuous helium-neon laser (maximum accessible power 600 µW, 1 mW embedded) used for interferometer referencing. It requires a dry purge gas supply (dew point -70 °C) at up to 10 L/min and 30 minutes warm-up to operating temperature. Its 20-minute dewar equilibrium time and -196 °C liquid-nitrogen dewar support cooled detector operation.
Full specifications
| Dewar Equilibrium Time | 20 minutes |
| Maximum Accessible Laser Power | 600 μW |
| Maximum Beam Divergence (embedded Laser) | 1.3 milliradians |
| Warm-up Time To Operating Temperature | 30 minutes |
| Wavelength (embedded Laser) | 632.8 nm |
| Purge flow rate (maximum) mfr | 10 L/min |
| Purge flow rate (recommended) mfr | 10 L/min |
| Regulator required above source pressure mfr | 60 psi |
| Laser pulse duration mfr | Continuous Continuous |
| Pulse duration (embedded laser) mfr | Continuous Continuous |
| Initial Dewar fill volume mfr | 500 milliliters |
| Dewar top-off volume mfr | 200 milliliters |
| Minimum side clearance mfr | 50 mm |
| Minimum rear clearance mfr | 150 mm |
| Maximum Accessible Power (embedded Laser) | 1 mW |
| Fuse type (spectrometer) mfr | 4 equivalent |
| Fuse type (microscope) mfr | 1 equivalent |
| Purge Gas Dew Point | -70 °C |
Description
Construction and series positioning
- The Cary 670 FTIR is a bench-top FTIR spectrometer; no sibling models are defined in the verified facts, so positioning within a series cannot be stated.
- The optical bench uses a continuous 632.8 nm He-Ne laser for reference; maximum accessible laser power is 600 µW (1 mW at the embedded laser), with beam divergence 1.3 mrad (embedded laser) and a separate 15° maximum beam divergence figure.
- The instrument includes a liquid-nitrogen dewar (initial fill 500 mL, top-off 200 mL) for cooled detector operation at -196 °C.
Behavior under load and media
- Purge gas must be dry to -70 °C dew point; recommended and maximum purge flow are both 10 L/min.
- A regulator is required above the source pressure at 60 psi.
- Warm-up to operating temperature takes 30 minutes; dewar equilibrium time is 20 minutes.
- No digitized curve points or application data are available, so spectral performance cannot be characterized from the verified facts.
Installation, connection and control
- Minimum clearances: 50 mm at the side and 150 mm at the rear.
- Fuse type: 4 equivalent for the spectrometer, 1 equivalent for the microscope.
- No PDF digests are attached, so control interfaces, electrical supply details and operating modes cannot be described.
Integration & Operating Notes
The purge inlet must not see source pressure above 60 psi without a downstream regulator; direct connection to a high-pressure cylinder or house line can overpressure the instrument's internal purge path. Install the regulator at the source and set delivery to the spectrometer's inlet requirement.
Liquid nitrogen at -196 °C is required for cooled detector operation. The initial fill volume is 500 mL; subsequent top-offs use 200 mL. Allow 20 minutes for dewar equilibrium before acquiring data, and handle liquid nitrogen with appropriate cryogenic PPE and venting.
The spectrometer uses 4 equivalent fuses; the microscope uses 1 equivalent fuse. Verify the correct fuse type and rating before replacing any fuse, and isolate power before opening the fuse holder. No voltage or current rating is stated in the verified facts.
Selection & Integration Q&A
The verified facts define no sibling models, so a direct comparison cannot be made. The Cary 670 FTIR is specified with a liquid-nitrogen dewar (-196 °C, 500 mL initial fill) and a 632.8 nm He-Ne reference laser; select it when these features match the application. Confirm alternatives against their own datasheets.
The purge gas must be dried to a -70 °C dew point and supplied at 10 L/min (recommended and maximum). If source pressure exceeds 60 psi, a regulator must be installed upstream. The instrument does not include a gas dryer or regulator in the base delivery.
Liquid nitrogen for the dewar (500 mL initial fill, 200 mL top-off) and a dry purge gas source at -70 °C dew point are required but not supplied. The spectrometer uses 4 equivalent fuses and the microscope 1 equivalent fuse; spare fuses are not listed as included. No maintenance schedule is given in the verified facts.