Agilent Technologies Cary 8454 UV Vis

Agilent Technologies Cary 8454 UV Vis — Molecular spectroscopy with Temperature 40 °C.

The Agilent Cary 8454 is a single-beam diode-array UV-Visible spectrophotometer covering 190–1100 nm with quartz cells, 350–1100 nm with glass, and 400 nm upward with disposable plastic. Its open sample area accepts quick-exchange cell holders and accessories, and it runs from 100–240 V AC at 50/60 Hz with 220 VA maximum consumption and no voltage selector. It is intended for routine and research molecular spectroscopy where fast, reproducible absorbance measurements and flexible cell configurations are required.

Temperature
40 °C

Full specifications

Core Performance & Functional Specifications
Temperature 40 °C
RAM 4 GB
Warm-up Time 1 hour
Clearance required at rear for electrical connections mfr 8 cm
Wavelength range with quartz cells mfr 190...1100 nm
Wavelength range with glass cells mfr 350...1100 nm
Wavelength range with disposable plastic cells mfr 400 nm
Turn-on and self-test duration mfr 1...2 minutes
Electronic self-test duration before lamp ignition mfr 20 seconds
Side clearance for air circulation mfr 2.5 cm
Rear clearance for electrical connections mfr 8 cm
Maximum cell wall angle mfr 0.1 arc
Processor mfr 3 GHz
A (in), B (out) mfr Without stirring stirring
C (in), D (out) mfr Without cell thermostatting thermostatting
A (in), B to D, C (out) mfr In series with cell thermostatting thermostattin
A and C (in), B and D (out) mfr In parallel with cell thermostatting thermostatt
Side clearance mfr 2.5 cm
Rear clearance mfr 8 cm
Hard disk mfr 40 GB
Dimensions, Weight & Mechanical Design
Depth 56.0 cm
Height 18.5 cm
Weight 14 kg
Width 34.4 cm
Recommended cell path length mfr 10 mm
Mass storage mfr 40 ROM
Electrical, Power & Drive Specifications
Power consumption 220 VA
Line voltage mfr 100...240 AC
Line frequency mfr 50 Hz
Interfaces, Connections & Control
LAN interface card buffer handling settings mfr 145
Screen resolution mfr Super VGA VGA
Operating Conditions, Environment & Protection
Operating Altitude 2000 m
Operating temperature 5 °C
Clearance required at sides for cooling mfr 2.5 cm
Operating humidity mfr 95
Non-operating altitude mfr 4600 m
Non-operating humidity mfr 90 %RH
Humidity (%RH) non-condensing mfr 90

Description

Construction and positioning

Optical behavior and media

Installation, control and operation

Integration & Operating Notes

Electrical
No voltage selector or user-accessible fuses

The instrument accepts 100–240 V AC at 50/60 Hz and uses automatic electronic fuses; there is no voltage selector to set and no externally accessible fuse to replace. Site power must simply fall within this range and provide 220 VA maximum.

Installation Guide, Section 2, page 17
Mounting
Maintain 2.5 cm side and 8 cm rear clearance

Bench space must accommodate the 34.4 × 56.0 × 18.5 cm envelope plus 2.5 cm clearance on either side for air circulation and approximately 8 cm at the rear for electrical connections. Blocking these clearances risks overheating or inaccessible cabling.

Installation Guide, Section 2, page 18
Process Limits
Cell material must match the wavelength range

Quartz cells are required for the full 190–1100 nm range; glass cells limit operation to 350–1100 nm; disposable plastic cells are usable only from 400 nm and are generally not recommended. Selecting the wrong cell material invalidates measurements below its cutoff.

Installation Guide, Section 3, page 33
Process Limits
Cell wall parallelism below 0.1 degrees of arc

Use 10 mm path length cells whose opposite walls are parallel to below 0.1 degrees of arc. Non-parallel or wedge-shaped cells cause spatial shift of the light beam and degrade photometric accuracy in this diode-array instrument.

Installation Guide, Section 3, page 34
Control & I/O
Wait for green indicator before launching ChemStation

After power-on, electronic self-tests run about 20 seconds before lamps ignite, and the full turn-on and self-test takes about 1–2 minutes. Start ChemStation only after the front-panel indicator turns green, otherwise the software may not recognize the spectrophotometer.

Installation Guide, Section 2, page 28
Control & I/O
Online session must be launched before offline

A maximum of two ChemStation sessions per instrument can run in parallel, and the online session must always be the first launched. Launching offline first can cause connection conflicts.

Installation Guide, Section 2, page 30
Control & I/O
No DHCP server in the instrument subnet

A DHCP server in the same subnet may respond to bootp requests, assign different IP addresses each time, and fail to set buffer handling correctly. Use fixed IP addresses reserved for instruments and set LAN card buffers to T145=64:T146=01:T147=01.

Installation Guide, Section 6, page 95
Maintenance
Blank every half hour on this single-beam instrument

Because the Cary 8454 is a single-beam spectrophotometer, a blank measurement should be taken every half hour even in a thermally stable environment. Skipping this degrades accuracy as lamp and detector drift accumulate.

Installation Guide, Section 3, page 32
Maintenance
Allow one hour warm-up before automated self-test

For best results, let the spectrophotometer warm up at least 1 hour after passing self-tests. The automated self-test in the diagnostic software can only be executed after this warm-up period.

Installation Guide, Section 2, pages 29-30
Cooling & Media
Waste tubing must slope continuously downward

Where waste tubing is connected, it must have a constant down slope from the spectrophotometer to the waste container. Improper slope can cause liquid accumulation and potential spillage or instrument damage.

Installation Guide, Section 2, page 25

Selection & Integration Q&A

When should I choose the Cary 8454 rather than another UV-Vis configuration in this family?

The verified facts define no sibling models, so selection is against application needs rather than variants: choose the Cary 8454 when you need 190–1100 nm coverage with quartz cells, an open sample area for quick cell-holder exchange, and single-beam operation with frequent blanking. For work confined above 350 nm, glass cells suffice; below 400 nm, quartz is mandatory.

What are the requirements for the PC, network and cell accessories that surround this instrument?

The PC runs Agilent ChemStation (software revision B.05.xx) and connects over LAN using a LAN interface card in the spectrophotometer. Use fixed IP addresses reserved for instruments, run no DHCP server in the same subnet, and set buffer handling to T145=64:T146=01:T147=01. RS 232 is inactive unless enabled by the configuration switch module next to the GPIB connector. Cells must be 10 mm path length with walls parallel below 0.1 degrees of arc.

What mounting, environment and duty-cycle limits apply to the Cary 8454?

Install below 2000 m altitude, within 5–40 °C and <95 %RH non-condensing, with 2.5 cm side and 8 cm rear clearance. For optimum performance hold the lab at 20–25 °C ±2 °C through the working day, away from direct sunlight and variable air circulation. Allow 1 hour warm-up after the 1–2 minute turn-on and self-test before running the automated self-test.

What maintenance routine and accessories are needed beyond the base instrument delivery?

Take a blank measurement every half hour. Optional accessories installed separately include the thermostattable cell holder, cell stirring module, long path cell holder and optical filter wheel. The computer, printer, LAN cabling and ChemStation software are not supplied with non-preconfigured systems; quartz, glass or plastic cells and waste tubing are also user-supplied. Lamp replacement is protected by safety light switches but still requires disconnecting line power.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
Manual ·104 pp.