Agilent Technologies IDP 15 Dry Scroll Pump
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies IDP 15 Dry Scroll Pump

Agilent Technologies IDP 15 Dry Scroll Pump — Liquid chromatography with Motor power 0.559 kW, Noise level 50 dB, Ultimate pressure 1.33 mbar.

The IDP-15 is a single-hermetic dry scroll pump with a 0.75 HP (550 W) motor, delivering 257 L/m at 60 Hz (213 L/m at 50 Hz) and an ultimate pressure of 1 × 10⁻² Torr. It is distinguished within the series by its 300 Torr maximum continuous inlet pressure and 20.7 g/hr water-vapor handling at gas ballast position 1. It serves as a roughing/backing pump for turbo or diffusion stages and for oil-free vacuum processes.

Motor power
0.559 kW
Noise level
50 dB
Ultimate pressure
1.33 mbar

Full specifications

Core Performance & Functional Specifications
Motor power 0.75 HP
Noise level 50 dBA
Ultimate pressure 1 Torr
Coverslip type No corrosive, explosive or particulate forming gases
Inlet Pressure 1.0 atmosphere
Operating Speed 1725 RPM
Water Vapor Handling Per Pneurop 6602 20.7 g/hr
Peak pumping speed mfr 60 hr
Maximum outlet pressure mfr 6.5 psig
Maximum recommended continuous inlet pressure mfr 300 Torr
Exhaust connection mfr 16
Gas ballast mfr 1 NPTF
Motor thermal protection mfr Automatic thermal protection protection
Leak rate (with exhaust and gas ballast sealed) mfr 1 helium
Vibration level at inlet, per mfr 9 rms
Installation Category mfr II II
Pollution mfr 2
Installation mfr Category II II
Europe mfr 10 m
Denmark mfr 10 m
Switzerland mfr 10 m
UK/Ireland mfr 13 m
India mfr 10 m
Israel mfr 10 m
Japan mfr 12 m
North America mfr 15 m
NORTH AMERICA mfr 1 agilent.com
PACIFIC RIM mfr please visit our website for individual office i
Danemark mfr 10 m
Suisse mfr 10 m
Royaume-Uni/Irlande mfr 13 m
Inde mfr 10 m
Israël mfr 10 m
Japon mfr 12 m
Amérique du Nord mfr 15 m
Will not start mfr Motor thermal protector open open
Poor ultimate pressure mfr System leak leak
Pump runs intermittently mfr Motor thermal protector is cycling open and clos
X3815-67000 mfr Replacement Tip Seal Set Set
X3815-69000 mfr -15 Standard
X3815-69010 mfr -15 Valve
Peak pumping speed at 60 Hz mfr 257 L/m
Peak pumping speed at 50 Hz mfr 213 L/m
Water vapor handling, position 2 mfr 235 g/hr
Vibration level at inlet mfr 9 rms
Dimensions, Weight & Mechanical Design
Shipping Weight 54 kg
Weight 30 kg
Weight, pump only mfr 30 kg
Electrical, Power & Drive Specifications
Maximum Starting Current 22.6 A
Typical Full Load Run Current 8.05 A
Fuse type mfr 215 D
Maximum starting current duration mfr 1 sec.
Operating voltage, low range mfr 100 VAC
Operating voltage, high range mfr 200 VAC
Interfaces, Connections & Control
Inlet flange 25
Interface dimensions mfr 1
Operating Conditions, Environment & Protection
Cooling System Air-cooled
Storage temperature -20...60 °C
Pollution Degree mfr 2
Altitude mfr 2000 m
Hammering noise mfr Pump overheated overheated
Standards, Safety & General Data
Vibration level at inlet, per ISO 10816-1 mfr 9 rms
Noise level, per ISO 11201 mfr 50 dBA

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Electrical
Match voltage selector to mains before first power-up

The pump accepts 1-phase 100–120 VAC or 200–240 VAC, 50/60 Hz, ±10%. The voltage selector switch must be set to the actual mains voltage; incorrect setting can damage the motor or control electronics and is not covered by the auto-reset thermal protector.

User Manual, Installation, p.21; Table 1, p.12
Electrical
Use a 15 A circuit for 100/120 V service

Starting current reaches 25 A at 120 V/50 Hz and 23.5 A at 100 V/60 Hz for less than 1 second. A 15 A circuit with motor-start-rated breaker prevents nuisance trips; smaller circuits may trip at start, especially below 100 V.

User Manual, Electrical Connections, p.23; Table 4, p.25
Process Limits
Do not exceed 300 Torr continuous inlet pressure

Continuous operation above 300 Torr inlet pressure overloads the scroll mechanism and motor, causing overheating and potential damage. Use the pump only within this rough-vacuum range and transition to a backing stage for higher-vacuum work.

User Manual, Table 1, p.12
Process Limits
Select gas ballast port by water-vapor load

Water-vapor handling is 20.7 g/hr at gas ballast position 1 and 235 g/hr at position 2. For high vapor loads, use position 2; for dry nitrogen purge, supply approximately 5 l/min via the 1/4 NPTF gas ballast inlet.

User Manual, Table 1, p.12; Gas Ballast, p.35
Mounting
Disengage shipping locks before operation

Shipping locks must be engaged for shipping, long-distance moving or long-term storage, and disengaged before operation. Operating with locks engaged can damage the pump. Also raise the L-brackets so the pump rests on its rubber feet and leave a minimum 2" gap front and back for airflow.

User Manual, p.17, p.21, p.33; Startup, p.22
Mounting
Use a bellows to isolate vibration from the chamber

Insert a bellows between the pump and vacuum chamber to reduce transmission of pump vibration to sensitive vacuum systems and to provide strain relief. Vibration level at the inlet is specified per the manual; rigid connection may couple vibration into the process.

User Manual, Mechanical Connections, p.33
Control & I/O
Integrated Isolation Valve timing affects restart sequence

On pumps fitted with the factory-installed Integrated Isolation Valve, the valve opens approximately 30 seconds after startup and has a typical burst pressure around 50 mTorr. EC-stamped pumps engage the valve on both power loss and motor thermal shutdown; pre-EC pumps only on power loss. Verify burst pressure on the specific system through testing.

User Manual, Operation, p.37-38
Maintenance
Allow several hours of run-in after tip seal replacement

Newly installed tip seals may require several hours of run time to seat properly. Do not judge pump performance immediately after tip seal replacement; ultimate pressure and pumping speed may improve over the first hours of operation.

User Manual, Tip Seal Replacement, p.47
Maintenance
Replace fuse only with Littelfuse Series 215, 10 A, 250 V

The specified fuse is Littelfuse Series 215, 10 A, 250 V, 20 L × 5.2 mm D. Using any other fuse may compromise protection and certification. Investigate the cause of fuse failure before replacement.

User Manual, Table 1 continued, p.13
Maintenance
Auto-reset thermal protector restarts pump when cooled

The IDP-15 has an auto-reset thermal motor protector that shuts down on overload and automatically restarts when cooled. Intermittent running may indicate an overload condition that must be identified and corrected; do not rely on the protector as a process control.

User Manual, Troubleshooting, p.39

Selection & Integration Q&A

When should I choose the IDP-15 over other dry scroll pumps in the Agilent series?

The IDP-15 is the only model described in the supplied facts; no sibling models are listed. Choose it when you need 257 L/m at 60 Hz, 1 × 10⁻² Torr ultimate, and 300 Torr maximum continuous inlet pressure. For higher water-vapor loads, use gas ballast position 2 (235 g/hr).

What are the requirements for the electrical supply and circuit protection for the IDP-15?

The motor is 0.75 HP (550 W), 1-phase 100–120 VAC or 200–240 VAC, 50/60 Hz, ±10%. For 100/120 V service, connect to a 15 A circuit; starting current is 25 A at 120 V/50 Hz and 23.5 A at 100 V/60 Hz for <1 sec. Use a motor-start-rated breaker to avoid nuisance trips.

What are the mounting and environmental limits for installing the IDP-15?

Install indoors, protected from rain, steam and excessive humidity, with ambient 5–40 °C. Leave a minimum 2" gap front and back for airflow; pump surface above 65 °C is potentially damaging. Disengage shipping locks before operation. Use NW25 or larger inlet hardware and short 16 mm ID or larger exhaust tubing.

What maintenance is required and what consumables or accessories are not included in the base delivery?

Tip seal replacement is required periodically; new seals need several hours of run time to seat. Service kits and advance exchange pumps (X3815-69000 standard, X3815-69010 with VPI valve) are available separately. The Integrated Isolation Valve is a factory-installed option, not included in the base pump. Fuse: Littelfuse Series 215, 10 A, 250 V.