Agilent Technologies IDP 7 Dry Scroll Pump

Agilent Technologies IDP 7 Dry Scroll Pump — Liquid chromatography with Motor power 2.24 kW.

The IDP-7 is a single-hermetic dry scroll vacuum pump delivered in low-voltage (100–120 VAC) and high-voltage (200–240 VAC) configurations, with an optional integral inlet isolation valve. It reaches 20 mT at 60 Hz (30 mT at 50 Hz) and pumps 152 L/min at 60 Hz (120 L/min at 50 Hz), serving as an oil-free backing or roughing stage for analytical instruments and clean vacuum systems.

Motor power
2.24 kW

Full specifications

Core Performance & Functional Specifications
Motor power 3 HP
Coverslip type No corrosive, explosive or particulate forming gases
Dimensions 248 mm
Inlet Pressure 1.0 atm
Operating Speed 1450 RPM
Water Vapor Handling Per Pneurop 6602 120 g/hr
Peak pumping speed mfr 50 L/min
Maximum recommended continuous inlet pressure mfr 150 T
Maximum outlet pressure mfr 5 psig
Exhaust connection mfr 16
Gas ballast mfr 1
Motor thermal protection mfr Automatic thermal protection protection
Leak rate (with exhaust sealed) mfr 1 helium
Hour meter mfr 1 hour
Emissions and Immunity mfr 61236 device
Safety mfr 61010 Canada
Installation mfr 2
Ultimate pressure mfr 30 mT
Europe mfr 10 m
Denmark mfr 10 m
Switzerland mfr 10 m
UK/Ireland mfr 13 m
India mfr 10 m
Israel mfr 10 m
Japan mfr 12 m
North America mfr 15 m
Operating voltages** mfr 100 phase
Will not start mfr Fuse blown blown
Poor ultimate pressure mfr System leak leak
Pump runs intermittently mfr Motor thermal protector is cycling open and clos
Service, Maintenance, Tip Seal, or Part Replacement mfr Wear or damage damage
X3807-67002 mfr Front Cowling Assembly Assembly
X3807-67003 mfr Rear Cowling Assembly Assembly
X3807-68008 mfr Gas Ballast Kit Kit
Fax: 00 39 011 9979 330 mfr 1
Fax Free: 00 800 345 00 mfr 800
Toll Free: 00 800 234 00 mfr vpl-ra@agilent.com agilent.com
Clearance around pump mfr 60 cm
Extension cord wire gauge mfr 16 wire
Water vapor handling with both gas ballast ports mfr 120 g/hr
Dry nitrogen flow for gas ballast mfr 5 l/min
Tip seal M5 bolt torque mfr 4 m
Dimensions, Weight & Mechanical Design
Shipping Weight 28.1 kg
Weight 24.5 kg
Extension cord maximum length mfr 25 feet
Electrical, Power & Drive Specifications
Typical Maximum Starting Current 22 A
Operating voltages mfr 100 VAC
Fuse Type mfr 5 equivalent
Power cord capacity mfr 10 A
Interfaces, Connections & Control
Inlet flange 25
Operating Conditions, Environment & Protection
Cooling System Air-cooled
Operating temperature 5 °C
Storage temperature 20...60 °C
Ambient Relative Humidity mfr 5 condensing
Altitude mfr 2000 m
Hammering noise mfr Pump overheated overheated
Standards, Safety & General Data
Noise Level (per ISO 11201) mfr 52 dB
Vibration level at inlet (per ISO 10816-1) mfr 1 mm/s

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Electrical
Verify voltage selector before first start

The pump leaves the factory configured for 200–240 VAC. Applying 100–120 VAC without moving the selector to 115, or the reverse, can damage the motor and invalidate the installation. Confirm the selector position and the correct 10 A Slo-Blo fuse (5×20 mm) before energizing.

Instruction Manual, Installation / Electrical Connections, p.12; Technical Information / Table 1, p.9
Electrical
Size circuit for 12–25 A starting inrush

Starting current can reach 25 A at 120 V 60 Hz and 12 A at 200 V 60 Hz. Breakers, wiring, and any extension cord must be rated for this inrush; use only 16-gage or heavier 3-wire grounded cords up to 25 ft (7.6 m). Undersized or long cords cause voltage drop, loss of power, and overheating.

Instruction Manual, Installation / Start Current Table 3, p.14; Extension Cords, p.13
Process Limits
Keep continuous inlet pressure below 150 T (200 mbar)

Continuous operation above 150 T (200 mbar) overheats the pump and accelerates tip-seal wear. For processes that routinely run at higher pressure, throttle the inlet or select a larger pump rather than relying on the IDP-7 to hold the load.

Instruction Manual, Technical Information / Table 1, p.9
Process Limits
Exhaust backpressure derates with line voltage

Maximum outlet pressure is 5 psig at 115 VAC or greater, falling linearly to 0 psig at 90 VAC; similarly 5 psig at 230 VAC or greater, falling to 0 psig at 180 VAC. Any exhaust restriction, filter, or long exhaust line must be sized so that backpressure stays within the value corresponding to the actual line voltage.

Instruction Manual, Technical Information / Table 1, p.9
Cooling & Media
Maintain 60 cm clearance on all sides

The pump is air-cooled and its surface can exceed 65 °C (150 °F), which is potentially damaging. Install with at least 60 cm (24 in) of clear space on all sides for ventilation and access, keep fan ducts unobstructed, and operate only within 5–40 °C ambient and 5–95% relative humidity (up to 2000 m altitude).

Instruction Manual, Environmental Conditions, p.4; Installation / Safety, p.12
Process Limits
Select gas ballast port by water-vapor load

Use gas ballast port 1 for moderate water-vapor loads and port 2 for high loads; with both ports in use the pump can handle up to 120 g/hr. For dry-nitrogen purge, supply approximately 5 L/min at the gas ballast inlet through a 1/4 NPT connection. After exposure to atmosphere, open GB1 for 15 minutes and/or flush with dry nitrogen before expecting ultimate pressure.

Instruction Manual, Installation / Gas Ballast, p.15; Technical Information footnote, p.9
Control & I/O
Isolation valve opens 30 s after start

With the optional integral inlet isolation valve, the valve opens 30 seconds after pump start and closes immediately on power loss or switch-off. Sequence chamber exposure and venting around this delay; the pump must reach operating speed before the chamber is connected, and the valve prevents backstreaming on shutdown.

Instruction Manual, Installation / Isolation Valve, p.15
Maintenance
Tip-seal kit X3807-67000; torque frame bolts to 4 N·m

Tip-seal replacement uses kit P/N X3807-67000 (seals and O-rings). During reassembly, torque the four M5 frame bolts to 4 N·m (40 in-lbs) to ensure proper sealing and avoid damage to the scroll housing. New tip seals may require several hours of running to seat before ultimate pressure is recovered.

Instruction Manual, Maintenance / Tip Seal Replacement, p.19; Table 5, p.18
Maintenance
Allow 24 h run-in before judging ultimate pressure

To reach maximum vacuum, leave the pump running for at least 24 hours with the inlet sealed. If the pump or system is new or has been exposed to atmosphere, first open GB1 for 15 minutes and/or flush with dry nitrogen to remove water; otherwise the specified 20 mT (60 Hz) or 30 mT (50 Hz) ultimate pressure will not be achieved.

Instruction Manual, Instructions for Use / Use, p.3; Technical Information footnote, p.9

Selection & Integration Q&A

When should I choose the IDP-7 rather than another configuration of the same pump?

Choose the IDP-7 when your required flow and base pressure match its frequency point: 152 L/min and 20 mT at 60 Hz, or 120 L/min and 30 mT at 50 Hz. Select the low-voltage (100–120 VAC) or high-voltage (200–240 VAC) build to suit your mains, and add the optional inlet isolation valve if the chamber must be protected from backstreaming on power loss.

What electrical supply and protection must be provided for the IDP-7?

Provide a grounded supply matching the selected voltage configuration (100–120 VAC or 200–240 VAC) and a circuit rated for 12–25 A starting inrush. Fit the specified 5×20 mm, 10 A Slo-Blo fuse (Littelfuse 215010 or equivalent); fast-blow fuses will nuisance-trip. Extension cords must be 16-gage minimum, 3-wire with ground, and no longer than 25 ft (7.6 m).

What maintenance parts and accessories are needed beyond the base pump?

The base delivery does not include the tip-seal replacement kit P/N X3807-67000 (seals and O-rings), the optional inlet isolation valve, the exhaust silencer, or country-specific power cords. Plan for tip-seal replacement with the kit and torque the four M5 frame bolts to 4 N·m (40 in-lbs); new seals may need several hours to seat.