Agilent Technologies NovoCyte Penteon
Agilent Technologies NovoCyte Penteon — Research flow cytometry.
The NovoCyte Penteon is a benchtop flow cytometer supporting up to five lasers, with detection channels defined in Appendix A. It operates in 15–32 °C laboratories and is a Class I laser product (IEC/EN 60825-1:2014). It serves research immunophenotyping and multi-parameter cell analysis, with fluidics and sampling handled by the NovoCyte Fluidics Station II and optional NovoSampler Q.
Full specifications
| Carryover | 0.1 percent |
| Sheath Flow Rate | 6.5 mL/min |
| Clean the External Surface of the Instrument mfr | Keep the external surface of the instrument clea |
| Decontamination mfr | When suspecting contamination in the system. sys |
| Prime Fluidics System before Long-Term Storage mfr | No plan to run the instrument for more than two |
| Prime Fluidics System after Long-Term Storage mfr | Instrument has not been used for more than two w |
| Purge Fluidics System before Shipment mfr | The instrument needs to be shipped. shipped. |
| Calibrate the NovoCyte Fluidics Station II or NovoCyte Fluidics Cart mfr | Abnormal warning about the liquid level in the r |
| Clean the Optical Filters mfr | The optical filters get dirty. dirty. |
| Clean Fluidics System mfr | • The CV of the data is larger than normal. • QC |
| Sample Aspiration Volume mfr | 5 mL |
| Sample Recovery mfr | Available Available |
| Compatibility to Autosampler mfr | No fluidic tubing disassembly or re-connection r |
| Sample Injection Probe (SIP) Rinse mfr | Automated flying collar wash of inner and outer |
| Data Acquisition Software mfr | NovoExpress NovoExpress |
| Parameters mfr | Height and Area for FSC, SSC and all Fluorescent |
| Dynamic Range mfr | 24 scale |
| Compensation mfr | Full inter-beam matrix, during or post acquisiti |
| File Export mfr | 3.0 report |
| Data Report mfr | Automatic report. Customizable report. Batch PDF |
| Usage Monitor mfr | Comprehensive Transaction Log and System Log Log |
| Quality Control mfr | Each instrument is verified for assembly integri |
| Sampling mfr | Automatic Sample Loading with NovoSampler Q Q |
| Regulatory Compliance mfr | IPXO IPXO |
| NovoCyte Fluidics Station II LED Status Indicator mfr | Green Green |
| Manual sample flow rate range mfr | 5 μL/min |
| Slow preset flow rate mfr | 14 μL/min |
| Medium preset flow rate mfr | 35 μL/min |
| Fast preset flow rate mfr | 66 μL/min |
| ADC sampling speed mfr | 80 second |
| ADC resolution levels mfr | 16,777,216 levels |
| NovoFlow low-level threshold mfr | 0.9 L |
| NovoRinse low-level threshold mfr | 150 mL |
| NovoClean low-level threshold mfr | 150 mL |
| Waste full threshold mfr | 2.8 L |
| Power button hold time for leak shutdown mfr | 6 seconds |
| Instrument weight mfr | 51 lb |
| Fluid Level Sensing mfr | Active sensing using weight sensors with automat |
| Operating temperature | 15 °C |
| Sound pressure | 60 dB |
| Storage temperature | 1 °C |
Description
Construction and series positioning
The NovoCyte Penteon is the five-laser-capable member of the NovoCyte flow cytometer family. Its architecture separates the optical/electronic instrument from the fluidics supply: the NovoCyte Fluidics Station II is a required companion, and an optional Fluidics Cart and NovoSampler Q automate sheath/waste handling and sample introduction. The rear panel carries a power port, four communication ports, and five fluidic connection ports, so the workstation, fluidics station, and sampler are all cabled to the instrument itself. Customized laser and detection configurations beyond the standard build are available; the exact channel map for any unit is defined in Appendix A of the manual, not by the model name.
Behavior and operating envelope
Operation is specified for 15–32 °C ambient; storage is 1–40 °C. The instrument is a Class I laser product under IEC/EN 60825-1:2014, meaning all laser energy is contained during normal operation and no special work area is required except during service by qualified Agilent personnel. EMC compliance is CISPR 11/EN 55011 Group 1, Class A and IEC/EN 61326-1, so the unit is intended for non-domestic establishments; potential EMC difficulties may exist in other environments. No flow-rate, sensitivity, or detection-efficiency curves are provided in the verified facts, so performance must be taken from the Appendix A configuration and the instrument QC test rather than from a published curve.
Installation, control, and operation
Startup follows a fixed sequence: pre-inspection, power on instrument and workstation, add reagents, empty waste, verify or modify configuration, run the instrument QC test, then proceed to sample acquisition. The power switch light is steady green when powered and flashing green while the shutdown cleaning procedure runs; the LED status indicator is green for normal, orange for warning, and red for error, with details reviewed by clicking the status bar in NovoExpress. Sample tubes accepted are 12 × 75 mm or 1.5 mL and 2.0 mL tubes. If a leak is detected, power must be cut by holding the front power button for about 6–7 seconds. The instrument weighs more than 51 lb (23 kg), so moving it requires assistance and proper lifting technique.
Integration & Operating Notes
The instrument weight is more than 51 lb (23 kg). Bench placement, unpacking, and any relocation must be planned with sufficient assistance and proper lifting technique to avoid injury and damage to the optical bench and fluidic connections.
If leaks are detected, power must be removed immediately by pressing and holding the power button on the front of the instrument for about 6–7 seconds. Operators should be trained on this procedure before routine use, because the normal shutdown path is not appropriate during a fluid release.
The rear of the instrument has a power port, four communication ports, and five fluidic connection ports. Communication ports link the instrument to the NovoCyte Workstation, the NovoCyte Fluidics Station II, and accessories such as the NovoSampler Q; fluidic ports connect reagent containers, so port assignment must be followed exactly during installation.
Compatible sample tubes are 12 × 75 mm tubes or 1.5 mL and 2.0 mL tubes. Using another tube type may prevent the tube from fitting the sample tube holder and can cause SIP alignment issues or failed acquisition, so tube selection must be fixed before an experiment is designed.
Do not use solvents with an auto-ignition temperature below 200 °C. Solvent selection for sheath additives, cleaning, or decontamination must account for this threshold to reduce fire risk near the instrument's heated or electrical areas.
The LED status indicator is green for normal, orange for warning, and red for error; warnings and errors must be reviewed by clicking the status bar in the NovoExpress software. A flashing green power switch light means the shutdown cleaning procedure is still running, so power should not be cut until it finishes.
Selection & Integration Q&A
Choose the Penteon when the panel requires up to five lasers; the exact laser and detection channel map is set per unit in Appendix A, and customized models beyond the standard configuration are available. If fewer lasers cover the panel, a smaller configuration avoids paying for unused channels.
A NovoCyte Workstation and Monitor plus a NovoCyte Fluidics Station II are required companions; the optional Fluidics Cart and NovoSampler Q connect via the four rear communication ports. Five rear fluidic ports must be plumbed to reagent containers, and the bench must carry more than 23 kg.
Operate only at 15–32 °C ambient; storage is 1–40 °C. It is Class A EMC equipment (CISPR 11/EN 55011 Group 1, IEC/EN 61326-1), so it belongs in non-domestic establishments. Class I laser containment means no special work area is needed except during qualified service.
Plan for fluidic system consumables, the optional 0.05 µm sheath fluid ultrafiltration filter, debubbling, decontamination, and calibration of the NovoCyte Fluidics Station II. Priming is required before and after long-term storage, and purging before shipment. Sample tubes (12 × 75 mm or 1.5/2.0 mL) are user-supplied.