Agilent Technologies NovoCyte Penteon
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies NovoCyte Penteon

Agilent Technologies NovoCyte Penteon — Research flow cytometry.

The NovoCyte Penteon is a benchtop flow cytometer supporting up to five lasers, with detection channels defined in Appendix A. It operates in 15–32 °C laboratories and is a Class I laser product (IEC/EN 60825-1:2014). It serves research immunophenotyping and multi-parameter cell analysis, with fluidics and sampling handled by the NovoCyte Fluidics Station II and optional NovoSampler Q.

Full specifications

Core Performance & Functional Specifications
Carryover 0.1 percent
Sheath Flow Rate 6.5 mL/min
Clean the External Surface of the Instrument mfr Keep the external surface of the instrument clea
Decontamination mfr When suspecting contamination in the system. sys
Prime Fluidics System before Long-Term Storage mfr No plan to run the instrument for more than two
Prime Fluidics System after Long-Term Storage mfr Instrument has not been used for more than two w
Purge Fluidics System before Shipment mfr The instrument needs to be shipped. shipped.
Calibrate the NovoCyte Fluidics Station II or NovoCyte Fluidics Cart mfr Abnormal warning about the liquid level in the r
Clean the Optical Filters mfr The optical filters get dirty. dirty.
Clean Fluidics System mfr • The CV of the data is larger than normal. • QC
Sample Aspiration Volume mfr 5 mL
Sample Recovery mfr Available Available
Compatibility to Autosampler mfr No fluidic tubing disassembly or re-connection r
Sample Injection Probe (SIP) Rinse mfr Automated flying collar wash of inner and outer
Data Acquisition Software mfr NovoExpress NovoExpress
Parameters mfr Height and Area for FSC, SSC and all Fluorescent
Dynamic Range mfr 24 scale
Compensation mfr Full inter-beam matrix, during or post acquisiti
File Export mfr 3.0 report
Data Report mfr Automatic report. Customizable report. Batch PDF
Usage Monitor mfr Comprehensive Transaction Log and System Log Log
Quality Control mfr Each instrument is verified for assembly integri
Sampling mfr Automatic Sample Loading with NovoSampler Q Q
Regulatory Compliance mfr IPXO IPXO
NovoCyte Fluidics Station II LED Status Indicator mfr Green Green
Manual sample flow rate range mfr 5 μL/min
Slow preset flow rate mfr 14 μL/min
Medium preset flow rate mfr 35 μL/min
Fast preset flow rate mfr 66 μL/min
ADC sampling speed mfr 80 second
ADC resolution levels mfr 16,777,216 levels
NovoFlow low-level threshold mfr 0.9 L
NovoRinse low-level threshold mfr 150 mL
NovoClean low-level threshold mfr 150 mL
Waste full threshold mfr 2.8 L
Power button hold time for leak shutdown mfr 6 seconds
Dimensions, Weight & Mechanical Design
Instrument weight mfr 51 lb
Materials, Media & Process Compatibility
Fluid Level Sensing mfr Active sensing using weight sensors with automat
Operating Conditions, Environment & Protection
Operating temperature 15 °C
Sound pressure 60 dB
Storage temperature 1 °C

Description

Construction and series positioning

The NovoCyte Penteon is the five-laser-capable member of the NovoCyte flow cytometer family. Its architecture separates the optical/electronic instrument from the fluidics supply: the NovoCyte Fluidics Station II is a required companion, and an optional Fluidics Cart and NovoSampler Q automate sheath/waste handling and sample introduction. The rear panel carries a power port, four communication ports, and five fluidic connection ports, so the workstation, fluidics station, and sampler are all cabled to the instrument itself. Customized laser and detection configurations beyond the standard build are available; the exact channel map for any unit is defined in Appendix A of the manual, not by the model name.

Behavior and operating envelope

Operation is specified for 15–32 °C ambient; storage is 1–40 °C. The instrument is a Class I laser product under IEC/EN 60825-1:2014, meaning all laser energy is contained during normal operation and no special work area is required except during service by qualified Agilent personnel. EMC compliance is CISPR 11/EN 55011 Group 1, Class A and IEC/EN 61326-1, so the unit is intended for non-domestic establishments; potential EMC difficulties may exist in other environments. No flow-rate, sensitivity, or detection-efficiency curves are provided in the verified facts, so performance must be taken from the Appendix A configuration and the instrument QC test rather than from a published curve.

Installation, control, and operation

Startup follows a fixed sequence: pre-inspection, power on instrument and workstation, add reagents, empty waste, verify or modify configuration, run the instrument QC test, then proceed to sample acquisition. The power switch light is steady green when powered and flashing green while the shutdown cleaning procedure runs; the LED status indicator is green for normal, orange for warning, and red for error, with details reviewed by clicking the status bar in NovoExpress. Sample tubes accepted are 12 × 75 mm or 1.5 mL and 2.0 mL tubes. If a leak is detected, power must be cut by holding the front power button for about 6–7 seconds. The instrument weighs more than 51 lb (23 kg), so moving it requires assistance and proper lifting technique.

Integration & Operating Notes

Mounting
Plan a two-person lift: instrument exceeds 23 kg

The instrument weight is more than 51 lb (23 kg). Bench placement, unpacking, and any relocation must be planned with sufficient assistance and proper lifting technique to avoid injury and damage to the optical bench and fluidic connections.

User Manual, Ergonomics, page 19
Electrical
Leak response: hold front power button 6–7 s for emergency off

If leaks are detected, power must be removed immediately by pressing and holding the power button on the front of the instrument for about 6–7 seconds. Operators should be trained on this procedure before routine use, because the normal shutdown path is not appropriate during a fluid release.

User Manual, Electrical Safety, page 20
Control & I/O
Rear panel provides four communication and five fluidic ports

The rear of the instrument has a power port, four communication ports, and five fluidic connection ports. Communication ports link the instrument to the NovoCyte Workstation, the NovoCyte Fluidics Station II, and accessories such as the NovoSampler Q; fluidic ports connect reagent containers, so port assignment must be followed exactly during installation.

User Manual, Instrument rear view, page 47
Process Limits
Use only 12 × 75 mm or 1.5/2.0 mL sample tubes

Compatible sample tubes are 12 × 75 mm tubes or 1.5 mL and 2.0 mL tubes. Using another tube type may prevent the tube from fitting the sample tube holder and can cause SIP alignment issues or failed acquisition, so tube selection must be fixed before an experiment is designed.

User Manual, Troubleshooting Table 11, page 137
Process Limits
Avoid solvents with auto-ignition temperature below 200 °C

Do not use solvents with an auto-ignition temperature below 200 °C. Solvent selection for sheath additives, cleaning, or decontamination must account for this threshold to reduce fire risk near the instrument's heated or electrical areas.

User Manual, General Safety, page 18
Maintenance
Read orange/red LED states in NovoExpress before running samples

The LED status indicator is green for normal, orange for warning, and red for error; warnings and errors must be reviewed by clicking the status bar in the NovoExpress software. A flashing green power switch light means the shutdown cleaning procedure is still running, so power should not be cut until it finishes.

User Manual, Table 5, page 47

Selection & Integration Q&A

When should I choose the NovoCyte Penteon rather than another NovoCyte configuration?

Choose the Penteon when the panel requires up to five lasers; the exact laser and detection channel map is set per unit in Appendix A, and customized models beyond the standard configuration are available. If fewer lasers cover the panel, a smaller configuration avoids paying for unused channels.

What must be in place around the instrument before installation?

A NovoCyte Workstation and Monitor plus a NovoCyte Fluidics Station II are required companions; the optional Fluidics Cart and NovoSampler Q connect via the four rear communication ports. Five rear fluidic ports must be plumbed to reagent containers, and the bench must carry more than 23 kg.

What environmental and duty-cycle limits apply to the Penteon?

Operate only at 15–32 °C ambient; storage is 1–40 °C. It is Class A EMC equipment (CISPR 11/EN 55011 Group 1, IEC/EN 61326-1), so it belongs in non-domestic establishments. Class I laser containment means no special work area is needed except during qualified service.

What maintenance and consumables are needed beyond the base delivery?

Plan for fluidic system consumables, the optional 0.05 µm sheath fluid ultrafiltration filter, debubbling, decontamination, and calibration of the NovoCyte Fluidics Station II. Priming is required before and after long-term storage, and purging before shipment. Sample tubes (12 × 75 mm or 1.5/2.0 mL) are user-supplied.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
NovoCyte Penteon Flow Cytometer User Guide
Manual ·150 pp.
View PDF