Agilent Technologies PCG 750

Agilent Technologies PCG 750 — Vacuum gauges.

The PCG 750 is a combination Pirani/capacitance diaphragm gauge with a DN 16 ISO-KF flange, covering 5×10⁻⁵ to 1500 mbar using both sensors continuously. It is the tungsten-filament base version of the PCG-75x series, distinguished from the PCG-752 by filament material and from setpoint/display/Profibus variants by its plain analog output. It serves as a compact, fast-response vacuum sensor for rough-to-fine vacuum monitoring and control.

Full specifications

Core Performance & Functional Specifications
Measurement range 5 mbar
Data Rate 12 Mbaud
Internal Volume 4.7 microliter
Operation 10 °C
Response Time 30 ms
Switching Time 30 ms
Vacuum connection 1.4435 millimeter
Measurement principle mfr thermal conductance acc. to Pirani Pirani
Accuracy (N2) mfr 2.5 reading
Repeatability (N2) mfr 1 reading
Output impedance mfr 2 proof
Load impedance mfr 10
Gauge identification mfr 71.5
HV adjustment mfr 10 mbar
ATM adjustment mfr 100 mbar
Switching functions setting range mfr 5.0 mbar
Hysteresis mfr 10 threshold
Switching characteristics mfr Low Trip Point Point
Switching function type mfr 1 function
Contact rating closed mfr 30 resistive
Contact rating open mfr LED lit solid solid
Ripple mfr 1 Vpp
Sensor cable mfr 0.14 conductor
Grounding concept vacuum connection to signal common mfr 10 nF
RS232C data format mfr 8 handshake
Materials exposed to vacuum - feedthrough mfr glass glass
Materials exposed to vacuum - orifice mfr stainless steel steel
Materials exposed to vacuum - diaphragm mfr ceramic ceramic
Materials exposed to vacuum - further materials mfr 1.4301 SnAg
Permissible pressure (absolute) mfr 5 bar
Bursting pressure (absolute) mfr 10 bar
Use mfr 2000 NN
Grounding concept mfr 10 nF
Materials exposed to vacuum - Feedthrough mfr glass glass
Materials exposed to vacuum - Orifice mfr stainless steel steel
Materials exposed to vacuum - Diaphragm mfr ceramic ceramic
Materials exposed to vacuum - Further materials mfr 1.4301 SnAg
AGPS45INT mfr Power supply, PVG-PCG, international plug. plug.
PCG750KF16 mfr -750 KF.
PCG750KF16RS mfr -750 KF.
PCG750KF16SD1 mfr -750 mbar.
PCG750KF16SD2 mfr -750 Torr.
PCG750KF16SD3 mfr -750 Pascal.
PCG750KF16SP mfr -750 mbar.
PCG752KF16 mfr -752 KF.
PCG752KF16SD1 mfr -752 mbar.
PCG752KF16SD2 mfr -752 Torr.
PCG752KF16SD3 mfr -752 Pascal.
PCG752KF16SP mfr -752 mbar.
Default setting mfr 0.1 Chex
PCG-752 mfr Ni Ni
Feedthrough mfr glass glass
Orifice mfr stainless steel steel
Diaphragm mfr ceramic ceramic
Further materials mfr 1.4301 SnAg
Year's mean mfr 65
During 60 days mfr 85
Grounding concept Vacuum connection to signal common mfr 10 nF
Data format mfr 8 handshake
About this manual mfr 6
Validity mfr 6
Definitions and terminology mfr 7
Definition of Caution, Warning and Note mfr 7
Warning Symbols mfr 8
Safety mfr 9
Proper use mfr 9
Improper use mfr 10
Safety guideline for PCG-750, PCG-752 mfr 11
Disposal mfr 12
Service mfr 13
Dimensions, Weight & Mechanical Design
Cable length 100 m
RS232C operation cable length mfr 30 m
RS232C transmission rate mfr 57600 baud
Mounting orientation mfr any any
Cable length, system wiring mfr according to Profibus specifications specificati
Cable length FCC 68 mfr 30 m
Transmission rate mfr 57600 baud
RS232C cable length mfr 30 m
Electrical, Power & Drive Specifications
Voltage vs. pressure mfr 1.286 logarithmic
Supply voltage at the gauge mfr 15 VDC
Fuse to be connected mfr 1 AT
Fuse to be connected 3) mfr 1 AT
Interfaces, Connections & Control
Profibus Data Rate 12 Mbaud
Output signal measurement range mfr 0.61 V
Diagnostic port mfr 2.5 pin
Power consumption without fieldbus mfr 2.5 W
Power consumption with fieldbus mfr 3 W
Profibus interface physical mfr 485
Profibus node address default setting mfr 0.1 Chex
Profibus connection cable mfr 9 cable
Profibus cable length, system wiring mfr according to Profibus specifications specificati
Weight w/o fieldbus interface mfr 115 g
Weight with fieldbus interface mfr 230 g
Profibus connection mfr 9
w/o fieldbus interface mfr 115 g
with fieldbus interface mfr 230 g
Specification, data format, communication protocol mfr 485
Weight without fieldbus interface mfr 115 g
Materials, Media & Process Compatibility
Materials exposed to vacuum - filament PCG-752 mfr Ni Ni
Materials exposed to vacuum - Filament PCG-752 mfr Ni Ni
Filament mfr 160 °C
Operating Conditions, Environment & Protection
Operating temperature 10 °C
Protection rating 40
Storage temperature 20 °C
Relative humidity year's mean mfr 65
Relative humidity during 60 days mfr 85
Relative humidity - Year's mean mfr 65
Relative humidity - During 60 days mfr 85

Description

Construction and series positioning

Measurement behavior and media

Installation, connection, and operation

Integration & Operating Notes

Electrical
Use a fused SELV supply within 15–30 VDC and ≤1 Vpp ripple

The gauge must be powered from a grounded protective extra-low voltage (SELV) source and the circuit must include a 1 AT fuse. Exceeding the 30 VDC upper limit or injecting excessive ripple can damage the electronics or corrupt the analog output; Agilent controllers already meet this requirement.

User Manual, Section 'Electrical Connection' / Technical Specifications p.18
Process Limits
Never exceed 5 bar absolute process pressure at the flange

The permissible absolute pressure is ≤5 bar and bursting pressure is 10 bar. Use overpressure-suitable clamps and O-rings with an outer centering ring for the KF connection. The ceramic diaphragm sensor is fragile; avoid sudden pressure surges or mechanical shock that could crack it.

User Manual, Technical Specifications p.19 / Vacuum Connection p.24
Mounting
Mount free of vibration and keep flange temperature ≤80 °C

Any orientation is allowed, but the gauge must not be subjected to vibration (e.g., directly on a roughing pump line). The vacuum connection temperature must stay ≤80 °C; if baking the system, ensure the flange does not exceed this limit. Indoor use only, with relative humidity ≤65% annual mean and no condensation.

User Manual, Vacuum Connection p.24 / Technical Specifications p.19
Control & I/O
Configure SP1/SP2 before using relay outputs for interlocks

Both setpoints are factory set to the lower measurement range limit (5×10⁻⁵ mbar) and therefore do not switch by default. If the relay contacts are wired into a valve or interlock circuit, set the trip points and hysteresis (10% of threshold) via the RS232C interface or buttons first; otherwise the interlock will never trigger.

User Manual, Switching Functions p.33
Control & I/O
Do not press <SP> while the analog signal controls a process

Pushing the <SP> button suppresses the measurement signal and outputs the threshold value instead. If a controller or PLC uses the analog output for pressure control, this action will cause a sudden false reading and possible malfunction. Use the RS232C diagnostic port for parameter changes instead.

User Manual, Adjusting the Setpoints p.36
Maintenance
Perform ATM and HV adjustments after contamination or drift

Zero drift can occur after long operation or contamination. ATM adjustment (default 1000 mbar) must be done at atmospheric pressure after at least 10 minutes stabilization; HV adjustment (default 5×10⁻⁵ mbar) requires evacuation to p << 10⁻⁵ mbar and a 2-minute wait. Successful HV adjustment is indicated by a 4.99×10⁻⁵ mbar output. Replace the sensor if contamination is severe.

User Manual, Maintenance and Adjusting the Gauge p.43–44

Selection & Integration Q&A

When should I choose the PCG 750 instead of the PCG 752 or a setpoint/display variant?

Choose the PCG 750 for standard vacuum monitoring with a tungsten filament and plain analog output. The PCG 752 offers a nickel-plated filament for slightly different chemical compatibility. If you need local readout, relay setpoints, or Profibus, select the SD or SP variants; those functions are not present on the base PCG 750.

What power supply and cabling are required to integrate the PCG 750 with a controller?

Use a SELV supply of +15…+30 VDC with ≤1 Vpp ripple, fused at 1 AT. The 8-pin FCC-68 connector carries supply, analog output, and relay contacts. Use shielded cable with 0.14 mm² conductors, max 100 m, and ground the shield on one side. The analog output is logarithmic (0.61–10.23 V) and must be interpreted accordingly.

What maintenance is needed and what accessories are not included in the base delivery?

The gauge is factory calibrated; perform ATM and HV adjustments if drift occurs, and replace the sensor if severely contaminated. The base delivery includes the gauge only—the power supply (AGPS45INT), mating FCC-68 connector, RS232C cable, and Profibus GSD file are not included and must be ordered separately. Setpoints are factory set to the lower limit and need configuration.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
PCG-750 / PCG-752 Vacuum Gauge Controller User Manual
Manual ·51 pp.
View PDF