Avantik QS12UV
Avantik QS12UV — Cryostats.
The Avantik QS12UV is a cryostat with 27 stations, including 4 Peltier freezing stations that reach -55 °C, and automated UV disinfection. It is distinguished from the base QS12 by its UV disinfection system, which enables between-session sterilization without manual chemical disinfection. The instrument is intended for sectioning frozen tissue in Mohs surgery and histology workflows, where precise temperature control and reproducible section thickness are critical.
Full specifications
| Chamber temperature | -55 degree Celsius |
| Weight | 315 lbs |
Description
Construction and positioning
- The QS12UV is a floor-standing cryostat weighing 315 lbs (142.9 kg), built around a 27-station workspace that includes 4 Peltier freezing stations capable of -55 °C. - It shares the core sectioning architecture with the QS12 sibling but adds automated UV disinfection, making it suitable for laboratories requiring enhanced contamination control between specimens. - The microtome uses stepper motor technology for section advance, with a dovetail clamping knifeholder and an extra-wide, fully adjustable anti-roll glass plate. ### Behavior under load and in operation
- Section thickness is set in microns via the EZ-touch touchscreen interface; the stepper motor provides consistent advance, while automatic specimen retraction on the return stroke minimizes knife damage and section compression. - The Peltier freezing stations operate at -55 °C, allowing rapid freezing of specimens; the chamber temperature is maintained at -55 °C for sectioning. - A memory function enables predefined advance to the knife edge, reducing setup time and improving reproducibility when switching between specimens. ### Installation, control and operation
- The QS12UV requires a stable, level floor location due to its weight; the touchscreen interface centralizes control of sectioning parameters, trim function, and disinfection cycles. - The smooth-rotation handwheel provides manual sectioning control, while the stepper motor handles fine advance; the trim function is controlled electronically for efficient block trimming. - Automated disinfection is initiated from the interface, eliminating manual chemical handling and reducing downtime between sessions.
Integration & Operating Notes
The 315 lb (142.9 kg) weight of the QS12UV requires a structurally adequate, level floor. Uneven surfaces can induce vibration during sectioning, degrading section quality. Verify floor loading capacity and use vibration-isolation pads if necessary.
The Peltier freezing stations and UV disinfection system draw significant current. Although exact electrical ratings are not specified, a dedicated circuit is advisable to prevent voltage fluctuations that could affect temperature stability and motor performance.
The automated UV disinfection cycle should be interlocked with the sectioning operation to prevent UV exposure during tissue handling. Ensure the control system enforces this interlock and that the chamber is clear before initiating disinfection.
The chamber should be defrosted according to the manual's schedule to maintain cooling efficiency. The UV disinfection system's lamps have a limited life and should be inspected periodically; replacement intervals are specified in the service manual.
Selection & Integration Q&A
Choose the QS12UV when automated UV disinfection between specimens is required for contamination control, such as in Mohs surgery with infectious cases. The base QS12 lacks UV disinfection and is suitable for standard histology where chemical disinfection is acceptable.
The QS12UV uses a dovetail clamping knifeholder, which accepts standard cryostat blades. Ensure blades are compatible with the dovetail clamp and that the anti-roll glass is adjusted to the blade width for proper section flattening.
The base delivery includes the cryostat and standard accessories, but consumables such as blades, OCT compound, and specimen holders must be ordered separately. UV lamps for disinfection are also not included as spares; their replacement interval is specified in the service manual.