BINDER E 28
BINDER E 28 — Drying ovens with Temperature 60…230 °C.
The BINDER E 28 is a naturally ventilated, electrically heated drying and heating oven with a 28 L stainless steel (V2A 1.4301) interior chamber and a 0–120 minute mechanical timer. It operates from 60 °C to 230 °C, distinguishing it from the B 28 incubator sibling (30–70 °C) within the same 28 L chassis. It is intended for drying, heat treatment and hot-air sterilization of laboratory loads up to 25 kg total.
Full specifications
| Temperature | 60...230 °C |
| Heating time to 150°C | 36 min |
| Interior volume | 28 L |
| Load per rack | 10 kg |
| Nominal power | 0.8 kW |
| Permitted load | 25 kg |
| Rear wall clearance | 160 mm |
| Side wall clearance | 100 mm |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 19 min |
| Unit doors mfr | 1 |
| Depth | 425 mm |
| Height | 405 mm |
| Interior depth | 250 mm |
| Interior height | 280 mm |
| Interior width | 400 mm |
| Width | 580 mm |
| Net weight of the unit (empty) mfr | 22 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The E 28 shares its 28 L chamber architecture (interior 400 × 280 × 250 mm W×H×D) with the B 28 incubator, but is configured as a drying/heating oven: mechanical thermostat, 0–120 min timer, fresh-air supply and hot-air sterilization provisions.
- Interior, pre-heating chamber and interior door sides are stainless steel V2A (1.4301 / AISI 304); the powder-coated housing is RAL 7035 or RAL 9003. Single door, net weight 22 kg.
- Unlike the B 28, which uses a main power switch and thermostat brake for overtemperature setting, the E 28 is started and stopped via its timer; both models carry a Class 1 temperature safety device per DIN 12880:2007.
Thermal behavior and load
- Working range 60–230 °C at ambient +18 to +40 °C; heating from ambient to 150 °C takes 36 min, and recovery after a 30 s door opening at 150 °C is 19 min.
- Nominal power 0.8 kW at 230 V 1~ 50/60 Hz; permitted load 25 kg total, 10 kg per rack.
- Operation above 150 °C may cause harmless yellowish-brown or blue oxidation discoloration of metallic surfaces; this is not a malfunction.
Installation and operation
- Natural ventilation requires rear clearance 160 mm, side clearance 100 mm, and at least 200 mm above the chamber; units must not be stacked.
- Supplied with a fixed power cable of at least 1800 mm and a grounded CEE 7/7 plug; IP 20, so splash-proof siting is mandatory.
- Temperature is set on the mechanical thermostat; the timer governs run time. Cleaning while energized is prohibited and water must not enter the chamber.
Integration & Operating Notes
The fixed power cable is at least 1800 mm long and terminates in a CEE 7/7 grounded plug. The outlet must remain accessible for emergency disconnection, and supply tolerance is ±10 % at 50 or 60 Hz. The 230 V variant carries two chamber fuses.
The E 28 control range begins at 60 °C; lower set-points are outside the specified range. Total charge must not exceed 25 kg, with no more than 10 kg per rack, to preserve airflow and thermal uniformity. Avoid corrosive charges such as acids and halides that attack the stainless steel.
Performance data are referenced to +22 °C ± 3 °C ambient. Operation outside +18 to +40 °C, above 70 % relative humidity, above 2000 m altitude, or in damp areas voids the specified behavior and risks condensation damage given IP 20 protection.
Selection & Integration Q&A
Choose the E 28 for drying, heat treatment or hot-air sterilization from 60 °C to 230 °C. Choose the B 28 when the process requires incubation between 30 °C and 70 °C with a main power switch and thermostat-brake overtemperature setting.
It needs a grounded 230 V, 50/60 Hz supply within ±10 %, nominal current 3.5 A and 0.8 kW. The 230 V variant has two chamber fuses. Size the upstream circuit for at least the nominal current and keep the plug accessible for emergency disconnection.
Plan annual maintenance and electrical repeat inspection by BINDER-authorized personnel, plus door gasket replacement when cold. The base delivery includes only the fixed cable and plug; racks, spare fuses and gaskets are accessories to be ordered separately.