BINDER ED-S 115
BINDER ED-S 115 — Drying ovens with Temperature 7…250 °C.
The ED-S 115 is a 118 L natural-convection drying and heating oven (Solid.Line) with a stainless steel interior and PID controller, operating from +7 to 250 °C. It is the larger 115-size variant of the ED-S series, distinguished from the ED-S 56 by its 118 L volume and 75 kg total load capacity. It serves gentle drying, heating, and thermal treatment of non-combustible materials where forced air is not required.
BINDER ED-S Series — Model Comparison & Selection
| Model | Temperature, °C | Depth, mm | Heating up time to 150 c, min | Height, cm | Interior depth, mm | Interior volume, L |
|---|---|---|---|---|---|---|
| ED-S 115 Current | 7–250 | 616 | 55 | 73.5 | 390 | 118 |
| ED-S 56 | 7–250 | 577 | 60 | 62.5 | 350 | 62 |
Full specifications
| Temperature | 7...250 °C |
| Heating time to 150°C | 55 min |
| Interior volume | 118 L |
| Load per rack | 20 kg |
| Nominal power | 1.25 kW |
| Permitted load | 75 kg (EDS115-230V) |
| Temperature fluctuation | 0.9 K |
| Temperature uniformity | 2.8 K |
| Rear wall clearance | 160 mm (EDS115-230V) |
| Side wall clearance | 100 mm (EDS115-230V) |
| Article Number mfr | 9090 |
| Recovery time after door was opened for 30 s at 150°C mfr | 40 min |
| Depth | 616 mm (EDS115-230V) |
| Height | 735 mm (EDS115-230V) |
| Interior depth | 390 mm (EDS115-230V) |
| Interior height | 550 mm (EDS115-230V) |
| Interior width | 550 mm (EDS115-230V) |
| Width | 764 mm (EDS115-230V) |
| Net weight of the unit (empty) mfr | 50 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 6.3 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and positioning
- The ED-S 115 uses natural convection (no fan) for gentle, uniform heat distribution, unlike the FD-S series which employs forced convection. Within the ED-S series, the 115 size offers 118 L interior volume (550 × 550 × 390 mm) versus the 56 size, and supports a total permissible load of 75 kg (20 kg per rack).
- The interior and door are stainless steel 1.4016 (AISI 430); the housing is polyester-coated RAL 9003. The chamber is equipped with an R-S microprocessor PID controller with digital display (1 °C resolution) and a class 2 overtemperature safety controller per DIN 12880.
- Electrical variants: 230 V 1~ 50/60 Hz (CEE 7/7 plug, 6.3 A fuse) or 120 V 1~ 60 Hz (NEMA 5-15 plug, 12.5 A fuse). The 230 V version draws 1.25 kW.
Behavior under load and in different modes
- Temperature range is +7 °C above ambient to 250 °C. At 150 °C, heating-up time is 55 min and recovery after a 30 s door opening is 40 min. Temperature fluctuation is 0.9 K and uniformity 2.8 K, suitable for reproducible drying but not for sub-degree incubation (display resolution 1 °C).
- Natural convection relies on the adjustable exhaust air flap to control air change; no fan means lower turbulence and reduced particle entrainment, but slower recovery after door openings compared to forced-convection models.
- Above 150 °C, metallic surfaces may discolour by oxidation; this is harmless and does not affect function.
Installation, connection, and operation
- Minimum clearances: 160 mm at the rear and 100 mm at the sides for heat dispersal. Do not install in unventilated recesses. The chamber must be lifted at its four lower corners by two people; never use the door handle for lifting.
- Electrical supply must be within ±10 % of nominal voltage and protected by the specified fuse. The 230 V model requires a grounded CEE 7/7 outlet; the 120 V model requires a NEMA 5-15 receptacle.
- The R-S controller supports set-point entry, temperature unit selection (°C/°F), and a delayed-off timer. The safety controller (class 2) must be set according to the overtemperature protection section of the manual.
- After cold storage, allow at least one hour at ambient temperature before start-up to avoid condensation. Maintenance is required at least annually, including electrical safety tests per EN 50678/VDE 0701 and EN 50699/VDE 0702.
Integration & Operating Notes
Insufficient clearance prevents heat dispersal and can cause overheating or damage. Do not install the chamber in unventilated recesses; ensure the room has adequate ventilation to remove dissipated heat.
The 230 V model requires a 1N~ 230 V 50/60 Hz supply with protective earth and a 6.3 A fuse. Verify mains quality before installation; out-of-tolerance voltage can affect controller operation and heating performance.
The chamber has no explosion protection. Operating temperature must remain below the flash point or sublimation point of the charging material. Only harmless materials may be processed; corrosive substances (acids, halides) can damage the stainless steel interior.
The ED-S 115 is fitted with a class 2 overtemperature safety controller per DIN 12880. Configure its set-point according to the manual to ensure independent protection against overheating; an unconfigured safety controller may not trip at the intended limit.
Perform maintenance at least once a year and repeat electrical inspections per EN 50678/VDE 0701 and EN 50699/VDE 0702 before initial start-up and after any repair. Disconnect the power plug before maintenance; do not remove the rear panel. Replace the door gasket only when cold.
Never lift or transport the chamber using the door handle or door. After cold storage, wait at least one hour at ambient temperature before start-up to prevent condensation-related electrical hazards.
Selection & Integration Q&A
Choose the ED-S 115 for larger loads up to 75 kg and 118 L where gentle natural convection is acceptable. The ED-S 56 offers only 56 L. Select an FD-S model if faster recovery after door openings or forced air circulation is needed.
The 230 V variant requires a 1N~ 230 V 50/60 Hz supply with protective earth, a CEE 7/7 plug, and a 6.3 A fuse. Voltage must remain within ±10 % of nominal. The 120 V variant uses a NEMA 5-15 plug and 12.5 A fuse.
Annual maintenance and electrical safety tests per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required. The base delivery includes the chamber only; racks, exhaust connection parts, and spare door gaskets must be ordered separately. Replace the door gasket only when cold.