BINDER FD-S 115
BINDER ·Made in Germany · In Active Production

BINDER FD-S 115

BINDER FD-S 115 — Drying ovens with Temperature 10…250 °C.

The FD-S 115 is a forced-convection drying and heating oven with a 106 L stainless-steel chamber (550 × 550 × 350 mm) and an R-S PID controller, operating from +10 to 250 °C. Within the FD-S series it is the larger 115-size chamber, distinguished from the 56-size sibling by volume and load capacity (75 kg total, 20 kg per rack). It is intended for reproducible drying, heating and thermal treatment of harmless materials where forced air circulation and defined air-change rate are required.

Temperature
10…250 °C

BINDER FD-S Series — Model Comparison & Selection

Model Temperature, °CAir change rate, 1/hourDepth, mmHeating up time to 150 c, minHeight, cmInterior depth, mm
FD-S 115 Current 10–25026162073.5350
FD-S 56 10–25045771562.5310

Full specifications

Core Performance & Functional Specifications
Temperature 10...250 °C
Air change rate 2 x/h
Heating time to 150°C 20 min
Interior volume 106 L
Load per rack 20 kg (FDS115-230V)
Nominal power 1.3 kW
Permitted load 75 kg (FDS115-230V)
Temperature fluctuation 0.4 K
Temperature uniformity 2.0 K
Rear wall clearance 160 mm (FDS115-230V)
Side wall clearance 100 mm (FDS115-230V)
Article Number mfr 9090
Recovery time after door was opened for 30 s at 150°C mfr 8 min
Dimensions, Weight & Mechanical Design
Depth 616 mm (FDS115-230V)
Height 735 mm (FDS115-230V)
Interior depth 350 mm (FDS115-230V)
Interior height 550 mm (FDS115-230V)
Interior width 550 mm (FDS115-230V)
Width 764 mm (FDS115-230V)
Net weight of the unit (empty) mfr 48 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

The FD-S 115 (Art. No. 9090-0024, designation 115-230) is the 230 V, 50 Hz variant of the 115-size forced-convection oven. The interior and door interior are stainless steel 1.4016 (AISI 430); the housing is polyester-coated RAL 9003. Compared with the FD-S 56 sibling, the 115 offers a 106 L interior instead of the smaller 56-size chamber, with net dimensions of 764 × 735 × 616 mm and a net weight of 48 kg. Unlike the natural-convection ED-S line, the FD-S uses forced-air circulation, which supports faster heat-up (20 min to 150 °C) and defined air exchange.

Thermal behaviour and control

Installation and operation

Integration & Operating Notes

Electrical
Supply must stay within ±10 % of 230 V at 50 Hz

The chamber is rated for 230 V 1~ 50 Hz with a 6.3 A fuse and 1.3 kW nominal power. Operating outside the ±10 % voltage tolerance stated on the type plate can cause controller faults and invalid thermal results; verify mains quality and provide a protective-earth connection before commissioning.

Instruction Manual BD-ED-FD-S (E1) 06-2026, chap. 4.3 and type plate
Mounting
Keep 160 mm rear and 100 mm side clearance

The 115-230 V variant requires 160 mm wall clearance at the back and 100 mm at the sides. Installing the chamber in an unventilated recess or closer than these distances prevents heat dispersal and can lead to overheating damage; the manual explicitly prohibits unventilated recess installation.

Instruction Manual BD-ED-FD-S (E1) 06-2026, chap. 3.4 Installation
Process Limits
Charging material must stay below its flash point

The chamber has no explosion-protection measures. Do not introduce combustible or explosive substances, and ensure the operating temperature remains below the flash point or sublimation point of the load. Corrosive materials (acids, halides) can attack the 1.4016 stainless steel interior and are excluded from liability.

Instruction Manual BD-ED-FD-S (E1) 06-2026, chap. 1.9 Intended Use
Process Limits
Respect 75 kg total and 20 kg per rack load

The permitted load for the 115-230 V variant is 75 kg overall, with a maximum of 20 kg per rack. Exceeding these limits can deform racks or the chamber floor and compromise temperature uniformity; distribute loads evenly across racks.

Factory Specification Sheet, FD-S 115
Control & I/O
Configure safety controller class 2 before unattended runs

The FD-S line uses an overtemperature safety controller of class 2 per DIN 12880, in addition to the standard class 1 protective device. Set the safety-controller cut-out above the working set-point but below the load's critical temperature, and verify alarm response before leaving the chamber unattended.

Instruction Manual BD-ED-FD-S (E1) 06-2026, chap. 9 Overtemperature Protection
Maintenance
Annual electrical inspection and service are mandatory

Plan maintenance at least once a year, including repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 before initial start-up and after any repair. Disconnect the power plug before service; do not remove the rear panel; replace the door gasket only when cold.

Instruction Manual BD-ED-FD-S (E1) 06-2026, chap. 12 Maintenance and Service
Cooling & Media
Air-change rate is set via the exhaust air flap

The 2 ×/h air-change rate is adjusted at the exhaust air flap during start-up. If an external exhaust connection is used, account for the additional ventilation path when setting the flap, since it affects both the air-change rate and the thermal uniformity inside the 106 L chamber.

Instruction Manual BD-ED-FD-S (E1) 06-2026, chap. 5 Start Up and Adjusting Air Change

Selection & Integration Q&A

When should I choose the FD-S 115 over the smaller FD-S 56, and when is the 56 the better fit?

Choose the FD-S 115 when the 106 L interior and 75 kg total load (20 kg per rack) are needed for larger batches or bulky parts. The FD-S 56 is preferable for smaller loads where its reduced chamber volume heats faster and uses less bench space; both share the +10 to 250 °C range and forced convection.

What supply and fusing does the 230 V FD-S 115 require, and can it be connected to a standard wall outlet?

It requires 230 V 1~ 50 Hz with protective earth, a CEE 7/7 grounded plug and a 6.3 A chamber fuse; nominal power is 1.3 kW. A standard grounded outlet is acceptable provided the supply stays within ±10 % of nominal voltage and the circuit is fused at 6.3 A.

What ambient and installation conditions must be met, and what happens after cold storage?

Install with 160 mm rear and 100 mm side clearance, never in an unventilated recess. Transport/storage limits are -10 to +60 °C and max. 70 % r.h. non-condensing; after cold storage, wait at least one hour at ambient temperature and ensure the chamber is dry before start-up.

What maintenance and accessories are required beyond the base delivery?

Plan annual maintenance and repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702. Racks, exhaust/ventilation connections and spare door gaskets are accessories/spare parts not included in base delivery; order them separately. Replace the door gasket only when cold.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model FD S 115 en
Datasheet ·394,8 КБ ·5 pp.
View PDF
PDF
BD ED FD S (E1) 06 2026 en
Manual ·10,8 МБ ·65 pp.
View PDF