BINDER FP 720
BINDER FP 720 — Drying ovens with Temperature 12…300 °C.
The FP 720 is a forced-convection drying and heating oven with a 741 L interior, configured for 400 V 3/N/PE 50 Hz supply (FP720-400V, Art. 9010-0366). It operates from 12 °C to 300 °C with 4.5 kW nominal power, distinguishing it from smaller FP 56/115/260 siblings by chamber volume and three-phase connection. It serves bulk thermal processing—drying, heating, and conditioning—where uniform temperature and high load capacity (315 kg total, 45 kg per rack) are required.
BINDER FP Series — Model Comparison & Selection
Full specifications
| Temperature | 12...300 °C |
| Air change rate | 5 x/h |
| Heating time to 150°C | 25 min |
| Interior volume | 741 L |
| Load per rack | 45 kg (FP720-400V / FP720UL-208V) |
| Nominal power | 4.5 kW |
| Permitted load | 315 kg (FP720-400V / FP720UL-208V) |
| Temperature fluctuation | 0.6 K |
| Temperature uniformity | 2.4 K |
| Rear wall clearance | 160 mm (FP720-400V / FP720UL-208V) |
| Side wall clearance | 100 mm (FP720-400V / FP720UL-208V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 6 min |
| Depth | 816 mm (FP720-400V / FP720UL-208V) |
| Height | 1590 mm (FP720-400V / FP720UL-208V) |
| Width | 1160 mm (FP720-400V / FP720UL-208V) |
| Net weight of the unit (empty) mfr | 166 kg |
| Power frequency | 50 Hz |
| Rated voltage | 400 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 3 N/PE |
Description
Construction and series positioning
- The FP 720 is the largest standard model in the BINDER FP Avantgarde.Line series, with 741 L net interior volume and net dimensions 1,160 mm W × 1,590 mm H × 816 mm D. Sibling models FP 56, FP 115, and FP 260 share the 10...300 °C range but offer smaller chambers; the FP 720 is the only size in this series requiring a 400 V three-phase supply (3 N/PE, 50 Hz, 16 A fuse).
- Construction includes forced convection, program functions, and the RD4 microprocessor temperature controller. A main power switch is located on the rear panel for this size. The chamber is rated IP 20 per EN 60529 and is intended for indoor use only.
Thermal behavior and load capacity
- The working temperature range is 12...300 °C, with temperature fluctuation of 0.6 K and uniformity of 2.4 K. Heating-up time to 150 °C is 25 min; recovery time after a 30 s door opening at 150 °C is 6 min. Air change rate is 5 x/h.
- The chamber supports a total permitted load of 315 kg, with a maximum of 45 kg per rack. These limits must be respected to avoid mechanical stress or impaired airflow.
Installation, control, and operation
- Installation requires minimum wall clearances: 160 mm at the rear and 100 mm at the sides. The chamber must not be installed in unventilated recesses; ambient conditions and ventilation per the operating manual must be observed.
- The RD4 controller provides set-point entry for temperature, fan speed, and air flap position. It supports program start/stop functions, tolerance range alarms, and a failsafe overtemperature protection device (class 1 standard; class 2/3.1 selectable).
- Electrical connection is via a CEE 16/6 H 5-pin plug for the 400 V variant. A zero-voltage alarm/switch contact is available via a SUB-D socket; an optional analog output requires a separate SUB-D socket. Ethernet and USB interfaces are provided for data logging and transfer.
Integration & Operating Notes
Insufficient clearance restricts heat dissipation and can cause overheating, affecting temperature performance and safety. The chamber must not be installed in unventilated recesses; ensure the installation site meets the ambient conditions specified in the manual.
The FP 720 requires a 400 V 3/N/PE 50 Hz supply with a 16 A fuse. Voltage deviations beyond ±10% may affect temperature control accuracy and safety. Verify supply stability before connection.
The SUB-D socket 'ALARM / SWITCH CONTACT' provides zero-voltage switching contacts for external monitoring. Ensure any connected device complies with EN 61010-1:2010 or EN 60950-1:2006 mod. The optional analog output requires a separate SUB-D socket.
Operating temperature must remain below the flash point or sublimation point of the load, and no explosive mixture with air may form. The chamber has no explosion protection. Also avoid corrosive substances (acids, halides) that may damage stainless steel components.
Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts; otherwise warranty is void. An annual electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is also required. The controller displays a 'Service due!' reminder after one year.
Selection & Integration Q&A
Choose the FP 720 when you need a 741 L chamber and a total load capacity of 315 kg (45 kg per rack). The FP 260 and FP 115 offer smaller volumes and single-phase 230 V connection, while the FP 720 requires a 400 V three-phase supply. Select the FP 720 for bulk processing where its larger capacity justifies the three-phase installation.
The FP 720-400V requires a 400 V 3/N/PE 50 Hz supply with a 16 A fuse and a CEE 16/6 H 5-pin plug. Voltage tolerance is ±10%. Ensure the supply is stable and meets the nominal power of 4.5 kW. For 60 Hz or 208 V applications, the FP720UL-208V variant with NEMA L21-20 plug is available.
Annual maintenance by authorized personnel and annual electrical safety inspection per EN 50678/VDE 0701 are required. The base delivery does not include options such as APT-COM 4 software, Pt100 object temperature sensor, analog output, HEPA fresh air filter, or inert gas connection. These must be ordered separately.