BINDER FP 720
BINDER ·Made in Germany · In Active Production

BINDER FP 720

BINDER FP 720 — Drying ovens with Temperature 12…300 °C.

The FP 720 is a forced-convection drying and heating oven with a 741 L interior, configured for 400 V 3/N/PE 50 Hz supply (FP720-400V, Art. 9010-0366). It operates from 12 °C to 300 °C with 4.5 kW nominal power, distinguishing it from smaller FP 56/115/260 siblings by chamber volume and three-phase connection. It serves bulk thermal processing—drying, heating, and conditioning—where uniform temperature and high load capacity (315 kg total, 45 kg per rack) are required.

Temperature
12…300 °C

BINDER FP Series — Model Comparison & Selection

Model Temperature, °CAir change rate, 1/hourDepth, mmHeating up time to 150 c, minHeight, cmInterior volume, L
FP 720 Current 12–300581625159741
FP 115 10–300326361973.2116
FP 260 10–30097861996.2259
FP 56 10–300805961562.260

Full specifications

Core Performance & Functional Specifications
Temperature 12...300 °C
Air change rate 5 x/h
Heating time to 150°C 25 min
Interior volume 741 L
Load per rack 45 kg (FP720-400V / FP720UL-208V)
Nominal power 4.5 kW
Permitted load 315 kg (FP720-400V / FP720UL-208V)
Temperature fluctuation 0.6 K
Temperature uniformity 2.4 K
Rear wall clearance 160 mm (FP720-400V / FP720UL-208V)
Side wall clearance 100 mm (FP720-400V / FP720UL-208V)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 150°C mfr 6 min
Dimensions, Weight & Mechanical Design
Depth 816 mm (FP720-400V / FP720UL-208V)
Height 1590 mm (FP720-400V / FP720UL-208V)
Width 1160 mm (FP720-400V / FP720UL-208V)
Net weight of the unit (empty) mfr 166 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 400 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 3 N/PE

Description

Construction and series positioning

Thermal behavior and load capacity

Installation, control, and operation

Integration & Operating Notes

Mounting
Maintain 160 mm rear and 100 mm side clearance

Insufficient clearance restricts heat dissipation and can cause overheating, affecting temperature performance and safety. The chamber must not be installed in unventilated recesses; ensure the installation site meets the ambient conditions specified in the manual.

Operating Manual FP (E3.1), chap. 3.4, p. 22
Electrical
Supply voltage must stay within ±10% of nominal

The FP 720 requires a 400 V 3/N/PE 50 Hz supply with a 16 A fuse. Voltage deviations beyond ±10% may affect temperature control accuracy and safety. Verify supply stability before connection.

Operating Manual FP (E3.1), chap. 1.6, p. 9
Control & I/O
Use zero-voltage contact for external alarm integration

The SUB-D socket 'ALARM / SWITCH CONTACT' provides zero-voltage switching contacts for external monitoring. Ensure any connected device complies with EN 61010-1:2010 or EN 60950-1:2006 mod. The optional analog output requires a separate SUB-D socket.

Operating Manual FP (E3.1), chap. 2.2 and 2.5, pp. 19-20
Process Limits
Do not exceed flash point of charging materials

Operating temperature must remain below the flash point or sublimation point of the load, and no explosive mixture with air may form. The chamber has no explosion protection. Also avoid corrosive substances (acids, halides) that may damage stainless steel components.

Operating Manual FP (E3.1), chap. 1.8 and 1.9, pp. 11-12
Maintenance
Annual maintenance by authorized personnel required

Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts; otherwise warranty is void. An annual electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is also required. The controller displays a 'Service due!' reminder after one year.

Operating Manual FP (E3.1), chap. 21.1 and 21.2, pp. 65-66

Selection & Integration Q&A

When should I choose the FP 720 over the smaller FP 260 or FP 115?

Choose the FP 720 when you need a 741 L chamber and a total load capacity of 315 kg (45 kg per rack). The FP 260 and FP 115 offer smaller volumes and single-phase 230 V connection, while the FP 720 requires a 400 V three-phase supply. Select the FP 720 for bulk processing where its larger capacity justifies the three-phase installation.

What are the power supply and connection requirements for the FP 720?

The FP 720-400V requires a 400 V 3/N/PE 50 Hz supply with a 16 A fuse and a CEE 16/6 H 5-pin plug. Voltage tolerance is ±10%. Ensure the supply is stable and meets the nominal power of 4.5 kW. For 60 Hz or 208 V applications, the FP720UL-208V variant with NEMA L21-20 plug is available.

What maintenance is required and what accessories are not included?

Annual maintenance by authorized personnel and annual electrical safety inspection per EN 50678/VDE 0701 are required. The base delivery does not include options such as APT-COM 4 software, Pt100 object temperature sensor, analog output, HEPA fresh air filter, or inert gas connection. These must be ordered separately.