BINDER KBF 1600
BINDER ·Made in Germany · In Active Production

BINDER KBF 1600

BINDER KBF 1600 — Climate chambers with Temperature 0…70 °C.

The BINDER KBF 1600 is a 1,610 L constant climate chamber with a 0...70 °C temperature range and 10...80 % RH humidity control, housed in a 1,795 mm wide cabinet. As the largest standard KBF model, it is distinguished from smaller siblings by its 40 kg per-rack load capacity and 640 W maximum heat compensation. It serves as a high-throughput test platform for stability and material testing under defined climatic conditions.

Temperature
0…70 °C

BINDER KBF Series — Model Comparison & Selection

Model Temperature, °CInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kWTemperature uniformity, K
KBF 1600 Current 0–701610406401.70.4
KBF 1060 0–701070406401.60.3
KBF 130 0–70127302701.60.2
KBF 260 0–70262404101.40.3
KBF 470 0–70472405301.40.2
KBF 720 0–70727406401.60.3

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 1610 L (KBF1600-230V / KBF1600UL-120V)
Load per rack 40 kg (KBF1600-230V / KBF1600UL-120V)
Maximum heat compensation 640 W
Nominal power 1.7 kW (KBF1600-230V / KBF1600UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.4 K
Article Number mfr 9020
Recovery time after door was opened for 30 s at 25°C and 60% RH mfr 3 min
Recovery time after door was opened for 30 s at 40°C and 75% RH mfr 7 min
Dimensions, Weight & Mechanical Design
Height 1951 mm (KBF1600-230V / KBF1600UL-120V)
Width 1795 mm (KBF1600-230V / KBF1600UL-120V)
Electrical, Power & Drive Specifications
Power frequency 50/60 Hz (KBF1600-230V / KBF1600UL-120V)
Rated voltage 220...240 V
Unit fuse 16.0 A (KBF1600-230V / KBF1600UL-120V)
Operating Conditions, Environment & Protection
Humidity fluctuation 1.5 % RH
Humidity range 10...80 % RH

Description

Construction and Series Positioning

The KBF 1600 is the largest chamber in the KBF series, offering 1,610 L of interior volume in a 1,795 mm wide by 1,951 mm high cabinet. Unlike the smaller KBF 130 and 260, which use a triangular instrument panel, the KBF 1600 features a signal triangle and an ergonomically adjustable control terminal for operator access. The chamber is equipped with a temperature safety device according to DIN 12880:2007 class 2/3.3, configurable as a temperature limiter or temperature protection controller. The refrigeration system uses R600A hydrocarbon refrigerant, and the humidification system design depends on the serial number (up to or from 20250000010225).

Behavior Under Load and in Different Media

Installation, Connection, and Control

Integration & Operating Notes

Electrical
Supply must match 220–240 V, 1~, 50/60 Hz with 16 A fuse

The KBF 1600 draws 1.7 kW nominal and requires a 16 A unit fuse. Verify site voltage and frequency before connection; use overcurrent protection rated for the chamber. The 120 V UL variant is a separate model and cannot be connected to a 230 V supply.

Data Sheet Model KBF 1600 en.pdf; Operating Manual KBF+KBF PRO (E7) 09-2025, Section 1.6 Type plate
Process Limits
Humidity sensor must not exceed listed chemical concentrations

Permanent exposure to ammonia above 5500 ppm, acetone above 3300 ppm, ethanol above 3500 ppm, formaldehyde above 2400 ppm, ozone above 0.5 ppm, hydrochloric acid above 300 ppm, hydrogen sulphide above 350 ppm, nitrogen oxides above 5 ppm, or sulphur dioxide above 5 ppm will damage the humidity sensor. Account for these limits when testing materials that off-gas such substances.

Operating Manual KBF+KBF PRO (E7) 09-2025, Section 1.14, p.20
Mounting
Wall spacer required for proper airflow and operation

From chamber size 260 upward, including the KBF 1600, a wall spacer must be installed to maintain the specified distance from the wall. Failure to do so can impair heat dissipation and humidity control, leading to inaccurate results or chamber malfunction.

Operating Manual KBF+KBF PRO (E7) 09-2025, Section 4.1, p.33
Maintenance
Annual maintenance and twice-yearly condenser cleaning required

Regular maintenance must be performed at least once a year, including steam humidifier service. Condensers should be cleaned at least twice a year by a qualified technician; in dusty environments, clean more frequently by suction. The maintenance reminder appears after 8760 operating hours or two years and must be reset by BINDER Service.

Operating Manual KBF+KBF PRO (E7) 09-2025, Section 23.2, p.133-134
Process Limits
Water connections must withstand 11 bar maximum pressure

The chamber's maximum operating pressure is 11 bar. Ensure that freshwater and wastewater connections, hoses, and fittings are rated for at least this pressure to prevent leaks or damage. The humidifying system design varies by serial number, so verify the correct connection method before installation.

Operating Manual KBF+KBF PRO (E7) 09-2025, Section 1.6 Type plate; Section 18.1
Control & I/O
Temperature safety device must be configured for application

The chamber is equipped with a temperature safety device class 2/3.3 per DIN 12880. The user must select between temperature limiter and temperature protection modes and set limit and offset values. Incorrect configuration can lead to either nuisance alarms or inadequate protection for sensitive samples.

Operating Manual KBF+KBF PRO (E7) 09-2025, Section 13 Temperature Safety Devices

Selection & Integration Q&A

When should I choose the KBF 1600 over the smaller KBF 1060 or KBF 470?

Choose the KBF 1600 when you need the largest interior volume (1,610 L) and can accommodate its 1,795 mm width. It supports 40 kg per rack and 640 W heat compensation, making it suitable for high-throughput testing or large batches. Smaller siblings like the KBF 1060 or 470 are preferable when floor space is limited or lower capacity suffices.

What are the requirements for the water supply and drainage connections?

Freshwater and wastewater connections must be rated for at least 11 bar maximum operating pressure. The humidifying system design differs by serial number (up to or from 20250000010225), so verify the correct connection method and spare parts for your chamber. Use hoses and fittings that meet or exceed this pressure rating.

What are the mounting and environmental limits for the KBF 1600?

The chamber requires a wall spacer to maintain proper distance from walls, and it is IP 20 rated so it must not be sprayed with water. Ambient dust increases condenser cleaning frequency. The chamber must be installed on a level surface and connected to a 220–240 V, 1~, 50/60 Hz supply with a 16 A fuse.

What maintenance is required and what accessories are not included in the base delivery?

Annual maintenance by authorized personnel is mandatory, including steam humidifier service and twice-yearly condenser cleaning. The base delivery does not include APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, flexible Pt 100 sensor, external freshwater/wastewater cans, or BINDER Pure Aqua Service; these must be ordered separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KBF 1600 en
Datasheet ·779,8 КБ ·9 pp.
View PDF
PDF
KBF+KBF PRO (E7) 09 2025 en
Manual ·16,5 МБ ·171 pp.
View PDF