BINDER KBF PRO 720
BINDER ·Made in Germany · In Active Production

BINDER KBF PRO 720

BINDER KBF PRO 720 — Climate chambers with Temperature -20…100 °C.

The KBF PRO 720 is a 727 L constant climate chamber with a -20…100 °C temperature range and 10…98 % RH humidity control, using a steam humidifier and R600A refrigeration circuit. It is the mid-large standard model in the KBF PRO series, distinguished from smaller siblings by its 40 kg-per-rack load capacity and 1,000 W max heat compensation. It serves as a programmable test environment for steady-state climatic conditioning of bulk samples and assemblies.

Temperature
-20…100 °C

BINDER KBF PRO Series — Model Comparison & Selection

Model Temperature, °CInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kWTemperature uniformity, K
KBF PRO 720 Current -20–1007274010001.80.3
KBF PRO 1060 -20–10010704010001.90.3
KBF PRO 130 -20–100127304501.90.2
KBF PRO 1600 -20–10016104010001.90.3
KBF PRO 260 -20–100262408001.60.2
KBF PRO 470 -20–100472408001.60.3

Full specifications

Core Performance & Functional Specifications
Temperature -20...100 °C
Interior volume 727 L
Load per rack 40 kg
Maximum heat compensation 1000 W
Nominal power 1.8 kW
Temperature fluctuation 0.1 K
Temperature uniformity 0.3 K
Article Number mfr 9020
Recovery time after door was opened for 30 s at 25°C and 60% RH mfr 2 min
Recovery time after door was opened for 30 s at 40°C and 75% RH mfr 5 min
Dimensions, Weight & Mechanical Design
Net weight of the unit (empty) mfr 312 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 208...240 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1
Operating Conditions, Environment & Protection
Humidity fluctuation 1 % RH
Humidity range 10...98 % RH

Description

Construction and series positioning

Behavior under load and in different media

Installation, connection and control

Integration & Operating Notes

Process Limits
Humidity control only between 10 and 90 °C

Although the chamber reaches -20…100 °C dry, humidity is limited to 10…98 % RH and the temperature range with humidity active is 10…90 °C. Programs that combine humidity with temperatures outside this band will not be controllable and may trigger humidity-system alarms.

Data Sheet Model KBF PRO 720 en.pdf; Operating Manual KBF+KBF PRO (E7) 09-2025, Section 18 Humidification / Dehumidification System
Process Limits
Humidity sensor tolerates only listed trace gases

The humidity sensor has a ±2 % RH tolerance and must not be exposed to substances above the manual's tolerated concentrations — for example ammonia 5,500 ppm, formaldehyde 2,400 ppm, hydrochloric acid 300 ppm, ozone 0.5 ppm under permanent load. Samples outgassing solvents or acids can degrade the sensor and invalidate humidity readings.

Operating Manual KBF+KBF PRO (E7) 09-2025, Section 1.14, p.20
Mounting
Rear wall spacer mandatory for this size

From chamber size 260 onward a wall spacer is required to hold the specified distance from the wall. The 720 needs this rear clearance for its refrigeration circuit and rear water connections; blocking it compromises heat rejection and access to the connection panel.

Operating Manual KBF+KBF PRO (E7) 09-2025, Section 4.1, p.33
Process Media
Water connections must withstand 11 bar

The chamber's maximum operating pressure is 11 bar, and the rear panel carries freshwater and wastewater connections for the steam humidifier. Supply and drain fittings, hoses and valves must be rated for at least this pressure, and the humidifier requires an annual service independent of chamber maintenance.

Operating Manual KBF+KBF PRO (E7) 09-2025, Section 1.6 Type plate; Section 23.2, p.134
Maintenance
Maintenance reminder at 8,760 h or two years

After 8,760 operating hours or two years the controller displays a maintenance message; confirming it only postpones it, and the window reappears every two weeks until BINDER Service resets it. Plan the service visit before the threshold to avoid recurring operator prompts.

Operating Manual KBF+KBF PRO (E7) 09-2025, Section 23.2, p.134
Maintenance
Condensers need cleaning at least twice a year

A qualified technician must clean the condensers at least twice yearly; in dusty ambient air, suction cleaning several times a year is advised. Refrigeration work is restricted to personnel trained per EN 13313:2010 because the circuit contains flammable R600A refrigerant.

Operating Manual KBF+KBF PRO (E7) 09-2025, Section 23.2, p.133-134
Control & I/O
Safety controller class must be set for the application

The overtemperature safety controller per DIN 12880 class 2/3.3 can be configured as temperature limiter or temperature protection, with limit and offset values and an undertemperature protection class 3.2. The chosen mode determines alarm behavior and whether the chamber shuts down or limits heating on deviation, so it must be agreed before validation runs.

Operating Manual KBF+KBF PRO (E7) 09-2025, Section 13 Temperature Safety Devices

Selection & Integration Q&A

When should I choose the KBF PRO 720 rather than the 470 or 1060 siblings?

Pick the 720 when 727 L interior volume and 40 kg per rack with 1,000 W heat compensation match your batch size. The 470 is smaller; the 1060 offers more volume but the same -20…100 °C range, so the decision is sample capacity and floor space, not temperature performance.

What water supply and drain provisions does the humidifier require?

The rear panel has freshwater and wastewater connections for the steam humidifier; fittings must be rated for at least 11 bar maximum operating pressure. Connection method and humidifying-system layout depend on serial number (up to or from 20250000010225), so confirm the unit's serial before ordering plumbing parts.

What ambient and duty-cycle limits apply when installing the 720?

The chamber is IP 20 and must not be sprayed with water. As a size-260-or-larger unit it requires the wall spacer to hold rear distance for refrigeration and water connections. Defrosting occurs during refrigerating operation, and anti-condensation protection limits continuous low-temperature/high-humidity duty.

What maintenance and accessories are not included in the base delivery?

Annual maintenance (including steam humidifier service), twice-yearly condenser cleaning and the 8,760 h/two-year service reset are not included. Optional items such as APT-COM 4 software, analog outputs, zero-voltage alarm contacts, Pt 100 object sensor, door lock, LED light and external water cans must be ordered separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KBF PRO 720 en
Datasheet ·706,0 КБ ·9 pp.
View PDF
PDF
KBF+KBF PRO (E7) 09 2025 en
Manual ·16,5 МБ ·171 pp.
View PDF