No photo available yet

Eppendorf ·Made in Germany · In Active Production

Eppendorf BioFlo ® 720 - Large-Scale Bioreactor

Eppendorf BioFlo ® 720 - Large-Scale Bioreactor — Bioreactors & fermenters.

The BioFlo 720 is a large-scale bioreactor control system for Thermo Scientific HyPerforma 5:1 single-use bioreactors from 50 L to 2000 L. This configuration is the 50 L total-volume single-use vessel variant, with interchangeable vessels and two gas supply/mixing lines. It serves as the controller and vessel platform for scale-up cell culture in mammalian, stem cell and insect applications.

Full specifications

Core Performance & Functional Specifications
Bacteria/yeast/fungi no
Dimensions (w × D × H) 80,6 x 87 x 160,3 cm
Exhaust Condensation Electric heater band
Gas Supply 2
Insect yes
Interchangeable Vessels yes
Mammalian/animal yes
Number Of Parallel Units Single unit
Stem Cells yes
Total Volume 50
Validatable yes
Vessels Single-use vessels
Dimensions, Weight & Mechanical Design
Product Weight 177 kg
Electrical, Power & Drive Specifications
Voltage 200 – 240 V
Standards, Safety & General Data
Standard Gas Flow Rates 0.01
Standard Gas Mixing 2

Description

Construction and positioning

Gas and process behavior

Control, installation and operation

Integration & Operating Notes

Process Limits
Vessel must be a HyPerforma 5:1 SUB within 50–2000 L

The controller is qualified against Thermo Scientific HyPerforma 5:1 single-use bioreactors across the 50 L to 2000 L range; this configuration is the 50 L vessel. Fitting a non-5:1 or third-party vessel leaves mixing and gas-transfer behavior outside the documented scope and voids the scale-up basis.

Fermentors Bioreactors Brochure BioFlo 720 Scale Up
Process Limits
Cell-type scope excludes microbial fermentation

Matched specs list mammalian/animal, stem cell and insect support as yes and bacteria-yeast-fungi as no. Microbial or yeast processes requiring high oxygen transfer and heat removal should not be scheduled on this configuration; use a fermentor-rated platform instead.

Factory Specification Sheet
Electrical
Supply must stay within 200–240 V

The matched electrical spec gives a 200–240 V range (SI nominal 220 V). Feeding the cabinet outside this window risks control and heater-band malfunction; confirm site voltage and breaker sizing against the 177 kg floor-standing cabinet before installation.

Factory Specification Sheet
Maintenance
Single-use vessel is a consumable, not part of the cabinet

The vessel is a single-use HyPerforma SUB and is replaced per batch; the 50 L vessel, gas filters and tubing are consumables outside the durable cabinet delivery. Stock the correct 5:1 SUB size and exhaust heater-band spares before campaign start.

Fermentors Bioreactors Brochure BioFlo 720 Scale Up
Control & I/O
Two gas supply and two gas mixing lines to be plumbed

The matched specs call for two gas supply and two gas mixing lines. Site gas panels must provide both feeds at the required purity and pressure; the standard gas flow field (0.01) carries no unit in the source, so size mass-flow controllers from the operating manual, not from this card.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose this 50 L BioFlo 720 configuration rather than a larger HyPerforma SUB size in the same controller family?

Pick the 50 L vessel for process development and pilot runs where the 0.3 L–2000 L tech-transfer path starts small; move to 100–2000 L SUBs on the same BioFlo 720 controller when bench data are locked and you need production-scale batches.

What adjacent system parts must be specified alongside the cabinet for a complete installation?

A HyPerforma 5:1 single-use vessel in the chosen size, two gas supply and two gas mixing feeds, and a 200–240 V supply. The exhaust heater band is part of the unit; vessel, filters and tubing are consumables ordered separately.

What maintenance and consumable schedule should be planned, and what is not in the base delivery?

The single-use vessel is replaced each batch and is not part of the durable cabinet; gas filters and tubing are likewise consumables. No service-interval data appear in the supplied brochure digest, so set PM intervals from the BioFlo 720 operating manual.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Fermentors Bioreactors Brochure BioFlo 720 Scale Up
Brochure ·2,6 МБ ·8 pp.
View PDF