Eppendorf CellXpert ® C170i - Cell Culture Incubator
Eppendorf ·Made in Germany · In Active Production

Eppendorf CellXpert ® C170i - Cell Culture Incubator

Eppendorf CellXpert ® C170i - Cell Culture Incubator — Incubators with Capacity 170 L.

The CellXpert C170i is a 170 L benchtop CO2 incubator with 4–50 °C control, 0.1 °C stability at 37 °C, and 180 °C high-temperature disinfection. It is the VisioNize touch-enabled variant of the C170 series, adding on-board data logging and BMS relay output for continuous culture monitoring.

Capacity
170 L

Full specifications

Core Performance & Functional Specifications
Capacity 170 L
Benchtop (b), Under Bench (u), Floor (f) Or Stackable (s) B, U, F, S
Bms Relay yes
Co 2 Control Increment 0.1 %
Co 2 Range 0.1...20 %
Co 2 Stability At 5 % Co 2 0.1 %
Co 2 Uniformity 0.1 %
Copper Package no
Dimensions (w × D × H) 71,8 × 71,5 × 90,0 cm
High Temperature Disinfection (htd) 180 °C
Load per rack 40 kg
Max. Load/shelf 10 kg
On-board Data Logging yes
Remote Device Monitoring And Notifications (visionize ® ) VisioNize ® touch enabled
Sealed Inner Glass Doors yes
Segmented Door Options 4
Temperature Accuracy 0.4 °C
Temperature Control Increment 0.1 °C
Temperature range 4...50 °C
Temperature Stability At 37 °c 0.1 °C
Temperature uniformity 0.3 °C
Dimensions, Weight & Mechanical Design
Product Weight 113 kg
Electrical, Power & Drive Specifications
Rated voltage 220 – 240 V
Interfaces, Connections & Control
Data Export Interface USB, Ethernet, VisioNize ®
Power Plug power plug type F
Access port mfr 2
Operating Conditions, Environment & Protection
Water/humidity Monitoring no
Standards, Safety & General Data
Number Of Shelves (standard/max) 4

Description

Construction and series positioning

The C170i is a 170 L benchtop incubator (71.8 × 71.5 × 90.0 cm, 113 kg) with sealed inner glass doors and a segmented outer door available in 1, 4, or 8 segments with right- or left-hand handle. Within the CellXpert C170 family it is the instrumented modification: VisioNize touch interface, on-board data logging, USB and Ethernet data export, and a BMS relay. The sibling C170 uses a standard LED display, 140 °C HTD, and no O2 control; the C170i supports O2/tri-gas control (1–20 %) and 180 °C HTD. Two 2-port access ports pass sensor or probe cabling into the chamber.

Control and atmosphere behavior

Temperature is controlled from 4 to 50 °C in 0.1 °C increments, with 0.1 °C stability at 37 °C, 0.3 °C uniformity, and 0.4 °C accuracy. CO2 is controlled from 0.1 to 20 % in 0.1 % steps, with 0.1 % stability at 5 % CO2 and 0.1 % uniformity. Water-humidity monitoring is not part of the base unit; a humidity monitor and water-level sensor are listed as accessories. No digitized performance curves are supplied, so load-dependent recovery behavior cannot be quantified from the available facts.

Installation, connection, and operation

Mains is 220–240 V with a type F power plug; the brochure also lists 100–120 V 50/60 Hz variants for USA/Japan, so the supply must be confirmed against the nameplate before installation. Data interfaces are USB, Ethernet, and VisioNize touch for remote monitoring and notifications. The BMS relay provides an external alarm contact. Shelving is rated 10 kg per shelf and 40 kg per rack, with 4 shelves standard. The 180 °C HTD cycle is the built-in disinfection routine; the copper package is not included.

Integration & Operating Notes

Electrical
Confirm nameplate voltage before connecting to mains

The unit is specified at 220–240 V with a type F plug, but the series brochure lists separate 100–120 V 50/60 Hz builds for USA/Japan. Connecting a 100–120 V build to a 230 V supply, or the reverse, will damage the controller; verify the nameplate and plug type against the site supply before commissioning.

Factory Specification Sheet; CO2 Incubators Brochure C170 C170i
Control & I/O
Route BMS alarm contacts to the facility monitoring system

The BMS relay must be wired to the building management or alarm system so that CO2 or temperature deviations trigger an external notification. Because the unit also supports VisioNize remote notifications, decide whether the BMS relay or the network path is the primary alarm route and configure both so they do not contradict each other.

Factory Specification Sheet; CO2 Incubators Brochure C170 C170i
Process Limits
Humidity monitoring requires an optional sensor

Relative humidity is not measured in the base configuration; a humidity monitor and water-level sensor are accessory items. If the culture protocol depends on documented humidity or on low-water alarm, these must be ordered and installed before the unit is qualified for that process.

CO2 Incubators Brochure C170 C170i, Accessory/Ordering Tables
Maintenance
Plan the 180 °C HTD cycle around cell work

The 180 °C high-temperature disinfection cycle is the built-in decontamination routine and requires the chamber to be emptied of cultures and heat-sensitive items. Schedule HTD between experimental runs, not during active culture, and allow the chamber to return to 37 °C and CO2 setpoint before reseeding.

CO2 Incubators Brochure C170 C170i
Control & I/O
Two access ports simplify probe and sensor routing

Two access ports allow temperature probes, CO2 sensors, or sampling lines to be introduced without disturbing the door seal. Use them for validation probes during mapping studies, and reseal unused ports to preserve chamber uniformity and gas consumption.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the C170i over the standard C170?

Pick the C170i when you need O2/tri-gas control (1–20 %), 180 °C HTD, VisioNize touch remote monitoring, on-board data logging, and a BMS relay. The C170 is the economical choice with a standard LED display, 140 °C HTD, and no O2 control.

What supply and data connections must the facility provide?

A 220–240 V supply with a type F receptacle (confirm the nameplate, since 100–120 V builds exist), plus an Ethernet drop for VisioNize and USB for local export. Wire the BMS relay to the alarm system if facility-level notification is required.

What accessories and service items are not in the base delivery?

The humidity monitor, water-level sensor, O2 control option, copper package, and segmented door configurations (1/4/8 segments, right/left handle) are order options. The 180 °C HTD cycle is built in; plan it as scheduled downtime between culture runs.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CO2 Incubators Brochure C170 C170i Culture Tomorrow
Brochure ·2,1 МБ ·20 pp.
View PDF