Eppendorf CellXpert ® CS220 - CO 2 Incubator Shaker
Eppendorf ·Made in Germany · In Active Production

Eppendorf CellXpert ® CS220 - CO 2 Incubator Shaker

Eppendorf CellXpert ® CS220 - CO 2 Incubator Shaker — Incubators with Capacity 224 L.

The CellXpert CS220 is a floor-standing, stackable CO2 incubator shaker with a 224 L chamber, 8–60 °C temperature range, 0.1–20 % CO2 control, and Eppendorf X-Drive providing 20–300 rpm at 2.54 cm orbit. It is the large-capacity, non-refrigerated modification for suspension culture and cell expansion where orbital agitation and controlled atmosphere are required in one unit.

Capacity
224 L

Full specifications

Core Performance & Functional Specifications
Capacity 224 L
Act Label yes
Audible And Visual Alarms yes
Benchtop (b), Under Bench (u), Floor (f) Or Stackable (s) F, S
Bms Relay yes
Co 2 Control Increment 0.1 %
Co 2 Range 0.1...20 %
Co 2 Stability At 5 % Co 2 0.1 %
Co 2 Uniformity 0.1 %
Control Accuracy 1 rpm
Door Opening glide up
Drive Mechanism Eppendorf X-Drive
High Temperature Disinfection (htd) 180 °C
On-board Data Logging yes
Orbit 2.5 cm
Platform Size (cm/in) 51.4 cm
Refrigerated no
Remote Device Monitoring And Notifications (visionize ® ) VisioNize ® touch enabled
Sealed Inner Glass Doors no
Segmented Door Options –
Slide-out Platform yes
Temperature Accuracy 0.4 °C
Temperature Control Increment 0.1 °C
Temperature range 8...60 °C
Temperature Stability At 37 °c 0.1 °C
Temperature uniformity 0.3 °C
Dimensions (W × D × H) mfr 131,8 × 80,5 × 66,1 cm / 51,9 × 31,7 × 26,0 in
Speed range mfr 20 – 300 rpm with 2.54 cm (1 in) orbit, 20 – 250
Speed mfr 20 – 300 rpm
Dimensions, Weight & Mechanical Design
Product Weight 240 kg
Electrical, Power & Drive Specifications
Power supply mfr 220 – 240 V ±10 %, 50 – 60 Hz
Interfaces, Connections & Control
Data Export Interface USB, Ethernet, VisioNize ®
Access port mfr 1
Operating Conditions, Environment & Protection
Relative Humidity Sensor Type bidirectional control
Relative humidity range mfr Up to 85 %
Relative humidity accuracy mfr ±5 %
Relative humidity control increment mfr 1 %
Relative humidity stability mfr ±2 %
Standards, Safety & General Data
Number Of Shelves (standard/max) –

Description

Construction and positioning

Behavior under load and in different modes

Installation, connection, control and operation

Integration & Operating Notes

Electrical
Supply must be 220–240 V ±10 %, 50–60 Hz

The unit is specified only for 220–240 V ±10 % at 50–60 Hz. Verify site voltage and frequency before connection; operation outside this window is not covered by the published specification. Provide a dedicated circuit sized for the unit and allow for the 240 kg floor-standing installation.

Factory Specification Sheet
Process Limits
Speed ceiling depends on orbit: 300 rpm at 1 in, 250 rpm at 2 in

The published speed range is 20–300 rpm with 2.54 cm (1 in) orbit and 20–250 rpm with 5.1 cm (2 in) orbit. Set the protocol speed against the actual orbit installed; a 300 rpm setpoint is only valid for the 1 in orbit configuration. Control increment is 1 rpm.

Factory Specification Sheet
Process Limits
Temperature control is above-ambient only (8–60 °C)

The unit is non-refrigerated, so it cannot hold setpoints below ambient and has no active cooling. For workflows requiring sub-ambient or tightly held low temperatures, this model is not suitable; plan cultures and assays within the 8–60 °C range and account for ambient conditions in the lab.

Factory Specification Sheet
Control & I/O
Data export and monitoring via USB, Ethernet and VisioNize

On-board data logging, USB and Ethernet interfaces, VisioNize touch-enabled remote monitoring and notifications, and a BMS relay are provided. Use the BMS relay for central alarm annunciation and the access port to route sensor or cable connections into the chamber without disturbing the door seal.

Factory Specification Sheet
Maintenance
High-temperature disinfection cycle runs at 180 °C

The HTD cycle reaches 180 °C for chamber disinfection. Schedule cleaning and disinfection around this cycle, and verify that all materials, platforms and accessories placed in the chamber are rated for 180 °C before running it. The cycle is part of routine contamination control, not a substitute for mechanical cleaning.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the CS220 over a smaller or refrigerated incubator shaker in the same family?

Choose the CS220 for large-volume suspension culture needing 224 L capacity, orbital shaking to 300 rpm (1 in orbit), CO2 control to 20 %, and stackability up to three units. It is non-refrigerated (8–60 °C only), so sub-ambient work requires a refrigerated model instead.

What supply and data connections must be provisioned for this unit?

Provision 220–240 V ±10 %, 50–60 Hz power. For data, provide USB and Ethernet connections; VisioNize touch enables remote monitoring and notifications, and a BMS relay is available for central alarm signaling. One access port is provided for routing cables into the chamber.

What are the installation and stacking limits for the CS220?

The unit weighs 240 kg and is floor-standing. Stacking is limited to three units, with defined arrangements: single on base, single on high feet, or triple-stack on low feet. Confirm floor loading and access clearance for the 131.8 × 80.5 × 66.1 cm footprint before installation.

What maintenance and accessory items are not part of the base delivery?

The base unit includes the chamber, slide-out platform, X-Drive and control interfaces. Consumables and accessories such as additional platforms, flask clamps, and cabling for USB/Ethernet/BMS are not listed as included and must be ordered separately. The 180 °C HTD cycle is the built-in disinfection routine.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Shakers Brochure CellXpert CS220 CO2 Incubator Shaker Accelerate Your
Brochure ·5,8 МБ ·8 pp.
View PDF