Eppendorf CellXpert ® CS220 - CO 2 Incubator Shaker
Eppendorf CellXpert ® CS220 - CO 2 Incubator Shaker — Incubators with Capacity 224 L.
The CellXpert CS220 is a floor-standing, stackable CO2 incubator shaker with a 224 L chamber, 8–60 °C temperature range, 0.1–20 % CO2 control, and Eppendorf X-Drive providing 20–300 rpm at 2.54 cm orbit. It is the large-capacity, non-refrigerated modification for suspension culture and cell expansion where orbital agitation and controlled atmosphere are required in one unit.
Full specifications
| Capacity | 224 L |
| Act Label | yes |
| Audible And Visual Alarms | yes |
| Benchtop (b), Under Bench (u), Floor (f) Or Stackable (s) | F, S |
| Bms Relay | yes |
| Co 2 Control Increment | 0.1 % |
| Co 2 Range | 0.1...20 % |
| Co 2 Stability At 5 % Co 2 | 0.1 % |
| Co 2 Uniformity | 0.1 % |
| Control Accuracy | 1 rpm |
| Door Opening | glide up |
| Drive Mechanism | Eppendorf X-Drive |
| High Temperature Disinfection (htd) | 180 °C |
| On-board Data Logging | yes |
| Orbit | 2.5 cm |
| Platform Size (cm/in) | 51.4 cm |
| Refrigerated | no |
| Remote Device Monitoring And Notifications (visionize ® ) | VisioNize ® touch enabled |
| Sealed Inner Glass Doors | no |
| Segmented Door Options | – |
| Slide-out Platform | yes |
| Temperature Accuracy | 0.4 °C |
| Temperature Control Increment | 0.1 °C |
| Temperature range | 8...60 °C |
| Temperature Stability At 37 °c | 0.1 °C |
| Temperature uniformity | 0.3 °C |
| Dimensions (W × D × H) mfr | 131,8 × 80,5 × 66,1 cm / 51,9 × 31,7 × 26,0 in |
| Speed range mfr | 20 – 300 rpm with 2.54 cm (1 in) orbit, 20 – 250 |
| Speed mfr | 20 – 300 rpm |
| Product Weight | 240 kg |
| Power supply mfr | 220 – 240 V ±10 %, 50 – 60 Hz |
| Data Export Interface | USB, Ethernet, VisioNize ® |
| Access port mfr | 1 |
| Relative Humidity Sensor Type | bidirectional control |
| Relative humidity range mfr | Up to 85 % |
| Relative humidity accuracy mfr | ±5 % |
| Relative humidity control increment mfr | 1 % |
| Relative humidity stability mfr | ±2 % |
| Number Of Shelves (standard/max) | – |
Description
Construction and positioning
- Floor-standing (F) and stackable (S) configuration; up to three units can be stacked, with single-unit-on-base, single-unit-on-high-feet, and triple-stack-on-low-feet arrangements.
- 224 L chamber with a 51.4 cm slide-out platform and a door that glides up rather than swinging out.
- No sealed inner glass doors and no segmented door options; access is through the single glide-up door.
- Non-refrigerated: temperature control is above-ambient only (8–60 °C), so it is not a cooled incubator.
Behavior under load and in different modes
- Orbital shaking via the Eppendorf X-Drive at 2.5 cm orbit; speed range 20–300 rpm at 2.54 cm (1 in) orbit and 20–250 rpm at 5.1 cm (2 in) orbit, with 1 rpm control increment and 1 rpm control accuracy.
- Temperature: 8–60 °C range, 0.1 °C control increment, 0.4 °C accuracy, 0.1 °C stability at 37 °C, 0.3 °C uniformity.
- CO2: 0.1–20 % range, 0.1 % control increment, 0.1 % stability at 5 % CO2, 0.1 % uniformity, ±0.3 % accuracy at 5 % CO2; NDIR sensor.
- Relative humidity: up to 85 %, ±5 % accuracy, 1 % control increment, ±2 % stability, bidirectional control.
- High-temperature disinfection (HTD) cycle at 180 °C.
Installation, connection, control and operation
- Power supply 220–240 V ±10 %, 50–60 Hz; unit weight 240 kg; dimensions 131.8 × 80.5 × 66.1 cm.
- One access port for routing cables or sensors into the chamber.
- Data export via USB, Ethernet, and VisioNize; VisioNize touch enabled for remote device monitoring and notifications; on-board data logging; BMS relay; audible and visual alarms; ACT label.
- Slide-out platform supports loading and unloading; door glides up for top access.
Integration & Operating Notes
The unit is specified only for 220–240 V ±10 % at 50–60 Hz. Verify site voltage and frequency before connection; operation outside this window is not covered by the published specification. Provide a dedicated circuit sized for the unit and allow for the 240 kg floor-standing installation.
The published speed range is 20–300 rpm with 2.54 cm (1 in) orbit and 20–250 rpm with 5.1 cm (2 in) orbit. Set the protocol speed against the actual orbit installed; a 300 rpm setpoint is only valid for the 1 in orbit configuration. Control increment is 1 rpm.
The unit is non-refrigerated, so it cannot hold setpoints below ambient and has no active cooling. For workflows requiring sub-ambient or tightly held low temperatures, this model is not suitable; plan cultures and assays within the 8–60 °C range and account for ambient conditions in the lab.
On-board data logging, USB and Ethernet interfaces, VisioNize touch-enabled remote monitoring and notifications, and a BMS relay are provided. Use the BMS relay for central alarm annunciation and the access port to route sensor or cable connections into the chamber without disturbing the door seal.
The HTD cycle reaches 180 °C for chamber disinfection. Schedule cleaning and disinfection around this cycle, and verify that all materials, platforms and accessories placed in the chamber are rated for 180 °C before running it. The cycle is part of routine contamination control, not a substitute for mechanical cleaning.
Selection & Integration Q&A
Choose the CS220 for large-volume suspension culture needing 224 L capacity, orbital shaking to 300 rpm (1 in orbit), CO2 control to 20 %, and stackability up to three units. It is non-refrigerated (8–60 °C only), so sub-ambient work requires a refrigerated model instead.
Provision 220–240 V ±10 %, 50–60 Hz power. For data, provide USB and Ethernet connections; VisioNize touch enables remote monitoring and notifications, and a BMS relay is available for central alarm signaling. One access port is provided for routing cables into the chamber.
The unit weighs 240 kg and is floor-standing. Stacking is limited to three units, with defined arrangements: single on base, single on high feet, or triple-stack on low feet. Confirm floor loading and access clearance for the 131.8 × 80.5 × 66.1 cm footprint before installation.
The base unit includes the chamber, slide-out platform, X-Drive and control interfaces. Consumables and accessories such as additional platforms, flask clamps, and cabling for USB/Ethernet/BMS are not listed as included and must be ordered separately. The 180 °C HTD cycle is the built-in disinfection routine.