Eppendorf Centrifuge 5420 - Centrifuge for Eppendorf tubes — Centrifuges.

The Centrifuge 5420 is a non-refrigerated benchtop centrifuge with a 24 × 1.5/2.0 mL fixed-angle rotor capacity, reaching 21,300 × g at 15,060 rpm. It is distinguished within its series by a compact 24 × 35 × 22 cm footprint and a 10-ramp soft-brake system. It is intended for routine pelleting and spin-down of microvolume samples in molecular biology workflows.

Full specifications

Core Performance & Functional Specifications
Acceleration Time 1) 15 s
Deceleration Time 1) 15 s
Dimensions (w × D × H) 24 x 35 x 22 cm
Max. Rcf 21300 g
Max. Rcf With Fixed-angle Rotor 21300 g
Rotational speed 100...15060 rpm
Soft Brake Function 10 ramps
Timer 10 s to 9:59 h
Rotors available mfr 2
Max. capacity mfr 24 × 1.5/2.0 mL
Dimensions, Weight & Mechanical Design
Height (with Open Lid) 44 cm
Product Weight 13 kg
Electrical, Power & Drive Specifications
Max. Power Consumption 290 W
Voltage 230 V
Interfaces, Connections & Control
Power Plug power plug type F
Operating Conditions, Environment & Protection
Cooling non-refrigerated
Noise level 56 dB

Description

Construction and positioning

Performance and control

Electrical and installation

Integration & Operating Notes

Electrical
Requires 230 V supply with type F outlet

The 5420 is wired for 230 V/50–60 Hz and ships with a type F (Schuko) plug. Installations in regions with type A/B/G/I outlets need a compatible adapter or receptacle change; verify circuit capacity for the 290 W motor.

Factory Specification Sheet
Mounting
Allow 44 cm height clearance for open lid

With the lid fully open the unit reaches 44 cm, so bench or shelf spacing must accommodate this to avoid obstructing lid travel. Blocking the lid can prevent proper locking and centrifugation.

Factory Specification Sheet
Process Limits
Non-refrigerated operation limits heat-sensitive samples

Because the 5420 has no cooling, rotor chamber temperature rises during long runs. For temperature-sensitive assays, limit run duration or pre-cool rotors; otherwise use a refrigerated model from the series.

Centrifugation Flyer Centrifuge 5420 Just Spin
Maintenance
Aerosol-tight rotor requires regular seal inspection

The FA-24x2 rotor is supplied aerosol-tight; its lid seal and rotor cavities should be checked for cracks or residue per the manual to maintain containment. Replace seals as recommended.

Centrifugation Flyer Centrifuge 5420 Just Spin

Selection & Integration Q&A

When should I choose the 5420 over a refrigerated sibling model?

Pick the 5420 for ambient-temperature spins of 1.5/2.0 mL or PCR tubes where cooling is unnecessary. Choose a refrigerated sibling when working with heat-labile samples or long runs that would warm the chamber.

What rotor and adapter setup do I need for 0.2 mL PCR strips?

Use the F-32x0.2-PCR rotor, which holds 32 × 0.2 mL PCR tubes or 4 × PCR strips. For 0.4 mL or 0.5 mL tubes, add the corresponding adapters in the FA-24x2 rotor.

What is included in the base delivery and what must be ordered separately?

Variant 5420 000 318 ships without a rotor; order the FA-24x2 or F-32x0.2-PCR rotor separately. Variant 5420 000 113 includes the aerosol-tight FA-24x2. Adapters for 0.4/0.5/0.2 mL tubes are separate accessories.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Centrifugation Flyer Centrifuge 5420 Just Spin
Brochure ·338,6 КБ ·2 pp.
View PDF